FPGA Electronics for OPET: A Dual-Modality Optical and Positron Emission Tomograph

被引:15
作者
Douraghy, Ali [1 ]
Rannou, Fernando R. [2 ]
Silverman, Robert W. [1 ]
Chatziioannou, Arion F. [1 ]
机构
[1] Univ Calif Los Angeles, David Geffen Sch Med, Crump Inst Mol Imaging, Los Angeles, CA 90095 USA
[2] Univ Santiago Chile, Dept Ingn Informat, Santiago 7251657, Chile
基金
美国国家卫生研究院;
关键词
Bioluminescence; optical imaging; positron emission tomography; small animal imaging;
D O I
10.1109/TNS.2008.2002257
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The development of a prototype dual-modality optical and PET (OPET) small animal imaging tomograph is underway in the Crump Institute for Molecular Imaging at the University of California Los Angeles. OPET consists of a single ring of six detector modules with a diameter of 3.5 cm. Each detector has an. 8 x 8 array of optically isolated 1360 scintillators which are coupled to multichannel photomultiplier tubes and open on the front end. The system operates in either PET or optical mode and reconstructs the data sets as 3D tomograms. The detectors are capable of detecting both annihilation events (511 keV) from PET tracers as well as Single Photon Events (SPEs) (2-3 eV) from bioluminescence. Detector channels are readout using a custom multiplex readout scheme and then filtered in analog circuitry using either a gamma-ray or SPE specific filter. Shaped pulses are sent to a Digital Signal Processing (DSP) unit. for event processing. The DSP unit has 100 MHz Analog-to-Digital Converters on the front-end which send digitized samples to Field Programmable Gate Arrays which are programmed via user configurable algorithms to process PET - coincidence events or bioluminescence SPEs. Information determined using DSP includes: event timing, energy determination-discrimination, position determination-lookup, and coincidence processing. Coincidence or SPE events are recorded to an external disk and minimal post processing is required prior to image reconstruction. Initial imaging results from a phantom filled with (18)FDG solution and an optical pattern placed on the front end of a detector module in the vicinity of a SPE source are shown.
引用
收藏
页码:2541 / 2545
页数:5
相关论文
共 13 条
[1]   Tomographic bioluminescence imaging by use of a combined optical-PET (OPET) system: a computer simulation feasibility study [J].
Alexandrakis, G ;
Rannou, FR ;
Chatziioannou, AF .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (17) :4225-4241
[2]   Effect of optical property estimation accuracy on tomographic bioluminescence imaging: simulation of a combined optical-PET (OPET) system [J].
Alexandrakis, George ;
Rannou, Fernando R. ;
Chatziioannou, Arion F. .
PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (08) :2045-2053
[3]   Evaluation of scintillator afterglow for use in a combined optical and PET imaging tomograph [J].
Douraghy, Ali ;
Prout, David L. ;
Silverman, Robert W. ;
Chatziioannou, Arion F. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2006, 569 (02) :557-562
[4]  
JUNNARKAR SS, 2008, P IEEE NUCL SCI S, V55, P919
[5]  
LEROUX JD, 2003, P IEEE NSS C OCT, V3, P1723
[6]   Readout of the optical PET (OPET) detector [J].
Prout, DL ;
Silverman, RW ;
Chatziioannou, A .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2005, 52 (01) :28-32
[7]   Detector concept for OPET - A combined PET and optical Imaging system [J].
Prout, DL ;
Silverman, RW ;
Chatziioannou, A .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2004, 51 (03) :752-756
[8]   Investigation of OPET performance using GATE, a Geant4-based simulation software [J].
Rannou, FR ;
Kohli, V ;
Prout, DL ;
Chatziioannou, AF .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2004, 51 (05) :2713-2717
[9]  
Ray P, 2003, CANCER RES, V63, P1160
[10]   Simple charge division readouts for imaging scintillator arrays using a multi-channel PMT [J].
Siegel, S ;
Silverman, RW ;
Shao, YP ;
Cherry, SR .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1996, 43 (03) :1634-1641