Probe-tone S-parameter measurements

被引:6
|
作者
Martens, J [1 ]
Kapetanic, P [1 ]
机构
[1] Anritsu Co, Morgan Hill, CA 95037 USA
关键词
modulated measurements; power amplifier measurements; S-parameters;
D O I
10.1109/TMTT.2002.802319
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The measurement of device behavior under complex actual operating conditions is an increasingly important measurement problem. In particular, it can be difficult to accurately measure gain and some reflection coefficients of a power amplifier operating under a realistic modulated signal drive. A small-signal measurement alone of a power amplifier is generally incorrect since the device-under-test will not be biased correctly. A fully modulated measurement, however, may require very dedicated equipment, long measurement times for adequate stability, and special calibration techniques. The methodology discussed here, i.e., the use of S-parameter probe signals in addition to power signals, will allow self-consistent S-parameter measurement under these conditions with full (traceable) vector calibrations and reduced uncertainties. In some sense, the small probe signal is used to quantify nonlinearities introduced by the modulated signal. In addition, the measurement has the flexibility to perform frequency-domain profiling to elucidate the behavior that may be experienced by interfering signals.
引用
收藏
页码:2076 / 2082
页数:7
相关论文
共 50 条
  • [1] IMPEDANCE MEASUREMENTS - PROBE-TONE INTENSITY AND MIDDLE-EAR REFLEXES
    TERKILDSEN, K
    OSTERHAM.P
    NIELSEN, SS
    ACTA OTO-LARYNGOLOGICA, 1970, : 205 - +
  • [2] 3-PROBE MICROSTRIP MEASURING SYSTEM FOR S-PARAMETER MEASUREMENTS
    LI, MY
    CHANG, K
    ELECTRONICS LETTERS, 1991, 27 (10) : 836 - 837
  • [3] On peculiarities of S-parameter measurements
    Rolain, Yves
    Van Moer, Wendy
    Jargon, Jeffrey A.
    DeGroot, Donald C.
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2007, 56 (05) : 1967 - 1972
  • [4] Performing S-parameter measurements
    Sundberg, G
    MICROWAVES & RF, 2001, 40 (06) : 99 - 100
  • [5] PROBE-TONE FREQUENCY AND DIAGNOSTIC VALUE OF TYMPANOMETRY
    ALBERTI, PW
    JERGER, JF
    ARCHIVES OF OTOLARYNGOLOGY-HEAD & NECK SURGERY, 1974, 99 (03) : 206 - 210
  • [6] True multiport S-parameter measurements
    不详
    MICROWAVE JOURNAL, 1997, 40 (10) : 140 - &
  • [7] Verification concepts in S-parameter measurements
    Mubarak, F.
    Zeier, M.
    Hoffmann, J.
    Ridler, N. M.
    Salter, M. J.
    Kuhlmann, K.
    2016 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS (CPEM 2016), 2016,
  • [8] S-Parameter Measurements in the Time Domain
    Tkadlec, Roman
    RADIOENGINEERING, 2004, 13 (02) : 12 - 16
  • [9] Characterizing Uncertainty in S-Parameter Measurements
    Buber, Tekamul
    Narang, Pragti
    Esposito, Giampiero
    Padmanabhan, Sathya
    Zeier, Markus
    MICROWAVE JOURNAL, 2019, 62 (10) : 88 - +
  • [10] 100-GHZ ACTIVE ELECTRONIC PROBE FOR ON-WAFER S-PARAMETER MEASUREMENTS
    MAJIDIAHY, R
    SHAKOURI, M
    BLOOM, DM
    ELECTRONICS LETTERS, 1989, 25 (13) : 828 - 830