Automation and validation of a MALDI-TOF MS (Mass-Fix) replacement of immunofixation electrophoresis in the clinical lab

被引:37
作者
Kohlhagen, Mindy [1 ]
Dasari, Surendra [1 ]
Willrich, Maria [1 ]
Hetrick, MeLea [1 ]
Netzel, Brian [1 ]
Dispenzieri, Angela [1 ]
Murray, David L. [1 ]
机构
[1] Mayo Clin, Dept Lab Med & Pathol, Rochester, MN 55906 USA
关键词
immunofixation; MALDI TOF; mass spectrometry; M-proteins; plasma cell disorders; MINIMAL RESIDUAL DISEASE; M-PROTEINS; MULTIPLE-MYELOMA; PEPTIDES;
D O I
10.1515/cclm-2020-0581
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Objectives: A matrix assisted laser desorption ionization time of flight mass spectrometry (MALDI-TOF MS) method (Mass-Fix) as a replacement for gel-based immunofixation (IFE) has been recently described. To utilize Mass-Fix clinically, a validated automated method was required. Our aim was to automate the pre-analytical processing, improve positive specimen identification and ergonomics, reduce paper data storage and increase resource utilization without increasing turnaround time. Methods: Serum samples were batched and loaded onto a liquid handler along with reagents and a barcoded sample plate. The pre-analytical steps included: (1) Plating immunopurification beads. (2) Adding 10 mu l of serum. (3) Bead washing. (4) Eluting the immunoglobulins (Igs), and reducing to separate the heavy and light Ig chains. The resulting plate was transferred to a second low-volume liquid handler for MALDI plate spotting. MALDI-TOF mass spectra were collected. Integrated in-house developed software was utilized for sample tracking, driving data acquisition, data analysis, history tracking, and result reporting. A total of 1,029 residual serum samples were run using the automated system and results were compared to prior electrophoretic results. Results: The automated Mass-Fix method was capable of meeting the validation requirements of concordance with IFE, limit of detection (LOD), sample stability and reproducibility with a low repeat rate. Automation and integrated software allowed a single user to process 320 samples in an 8 h shift. Software display facilitated identification of monoclonal proteins. Additionally, the process maintains positive specimen identification, reduces manual pipetting, allows for paper free tracking, and does not significantly impact turnaround time (TAT). Conclusions: Mass-Fix is ready for implementation in a high-throughput clinical laboratory.
引用
收藏
页码:155 / 163
页数:9
相关论文
共 16 条
[1]   Monitoring free light chains in serum using mass spectrometry [J].
Barnidge, David R. ;
Dispenzieri, Angela ;
Merlini, Giampaolo ;
Katzmann, Jerry A. ;
Murray, David L. .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2016, 54 (06) :1073-1083
[2]   Using matrix-assisted laser desorption/ionization time-of-flight mass spectrometry to detect monoclonal immunoglobulin light chains in serum and urine [J].
Barnidge, David R. ;
Krick, Thomas P. ;
Griffin, Timothy J. ;
Murray, David L. .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2015, 29 (21) :2057-2060
[3]   Monitoring M-Proteins in Patients with Multiple Myeloma Using Heavy-Chain Variable Region Clonotypic Peptides and LC-MS/MS [J].
Barnidge, David R. ;
Tschumper, Renee C. ;
Theis, Jason D. ;
Snyder, Melissa R. ;
Jelinek, Diane F. ;
Katzmann, Jerry A. ;
Dispenzieri, Angela ;
Murray, David L. .
JOURNAL OF PROTEOME RESEARCH, 2014, 13 (04) :1905-1910
[4]   Clonotypic Light Chain Peptides Identified for Monitoring Minimal Residual Disease in Multiple Myeloma without Bone Marrow Aspiration [J].
Bergen, H. Robert, III ;
Dasari, Surendra ;
Dispenzieri, Angela ;
Mills, John R. ;
Ramirez-Alvarado, Marina ;
Tschumper, Renee C. ;
Jelinek, Diane F. ;
Barnidge, David R. ;
Murray, David L. .
CLINICAL CHEMISTRY, 2016, 62 (01) :243-251
[5]   An international multi-center serum protein electrophoresis accuracy and M-protein isotyping study. Part II: limit of detection and follow-up of patients with small M-proteins [J].
Jacobs, Joannes F. M. ;
Turner, Katherine A. ;
Graziani, Maria Stella ;
Frinack, Jody L. ;
Ettore, Michael W. ;
Tate, Jillian R. ;
Booth, Ronald A. ;
McCudden, Christopher R. ;
Keren, David F. ;
Delgado, Julio C. ;
Zemtsovskaja, Galina ;
Fullinfaw, Robert O. ;
Caldini, Anna ;
de Malmanche, Theo ;
Katakouzinos, Katina ;
Burke, Matthew ;
Palladini, Giovanni ;
Altinier, Sara ;
Zaninotto, Martina ;
Righetti, Gabriella ;
Melki, Marie Therese ;
Bell, Stephen ;
Willrich, Maria Alice Vieira .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2020, 58 (04) :547-559
[6]   Assay to rapidly screen for immunoglobulin light chain glycosylation: a potential path to earlier AL diagnosis for a subset of patients [J].
Kumar, Sanjay ;
Murray, David ;
Dasari, Surendra ;
Milani, Paolo ;
Barnidge, David ;
Madden, Benjamin ;
Kourelis, Taxiarchis ;
Arendt, Bonnie ;
Merlini, Giampaolo ;
Ramirez-Alvarado, Marina ;
Dispenzieri, Angela .
LEUKEMIA, 2019, 33 (01) :254-257
[7]   Prevalence of monoclonal gammopathy of undetermined significance [J].
Kyle, RA ;
Therneau, TM ;
Rajkumar, SV ;
Larson, DR ;
Plevak, MF ;
Offord, JR ;
Dispenzieri, A ;
Katzmann, JA ;
Melton, LJ .
NEW ENGLAND JOURNAL OF MEDICINE, 2006, 354 (13) :1362-1369
[8]  
Litwin CM, 1999, AM J CLIN PATHOL, V112, P411
[9]   Prognostic value of deep sequencing method for minimal residual disease detection in multiple myeloma [J].
Martinez-Lopez, Joaquin ;
Lahuerta, Juan J. ;
Pepin, Francois ;
Gonzalez, Marcos ;
Barrio, Santiago ;
Ayala, Rosa ;
Puig, Noemi ;
Montalban, Maria A. ;
Paiva, Bruno ;
Weng, Li ;
Jimenez, Cristina ;
Sopena, Maria ;
Moorhead, Martin ;
Cedena, Teresa ;
Rapado, Immaculada ;
Victoria Mateos, Maria ;
Rosinol, Laura ;
Oriol, Albert ;
Blanchard, Maria J. ;
Martinez, Rafael ;
Blade, Joan ;
San Miguel, Jesus ;
Faham, Malek ;
Gracia-sanz, Ramon .
BLOOD, 2014, 123 (20) :3073-3079
[10]   The utility of MASS-FIX to detect and monitor monoclonal proteins in the clinic [J].
Milani, Paolo ;
Murray, David L. ;
Barnidge, David R. ;
Kohlhagen, Mindy C. ;
Mills, John R. ;
Merlini, Giampaolo ;
Dasari, Surendra ;
Dispenzieri, Angela .
AMERICAN JOURNAL OF HEMATOLOGY, 2017, 92 (08) :772-779