Determination of arsenic and selenium in river water by hydride generation inductively coupled plasma-mass spectrometry with high resolution

被引:24
作者
Narasaki, H
Cao, JY
机构
[1] Department of Chemistry, Faculty of Science, Saitama University, Shimo-Okubo
关键词
arsenic; selenium; river water; hydride generation; inductively coupled plasma-mass spectrometry;
D O I
10.2116/analsci.12.623
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Hydride generated in a mixing coil was separated in a gas-liquid separator which was closed with an electromagnetic pinch valve and then swept continuously into the torch of a high-resolution inductively coupled plasma-mass spectrometer. The system was operated with variable speed-peristaltic pumps by manipulating electromagnetic relays and a timer. Water vapor in the hydride gas was removed by inserting a glass trap immersed in an ice bath. Interference with transition matals was eliminated by using Chelex 100 resin. The detection limits (3 times the signal-to-noise ratio) of arsenic and selenium are 0.03 and 0.06 ng/ml, respectively, at a resolution of 10000. The accuracies checked with biological standard reference materials obtained from the National Institute of Science and Technology (NIST), USA, and the National Research Council(NRC), Canada, were within the certified values. The arsenic and selenium contents in river water were determined by the present method.
引用
收藏
页码:623 / 627
页数:5
相关论文
共 11 条
[1]   DETERMINATION OF TOTAL AND LEACHABLE ARSENIC AND SELENIUM IN SOILS BY CONTINUOUS HYDRIDE GENERATION INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETRY [J].
ANDERSON, STG ;
ROBERT, RVD ;
FARRER, HN .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1994, 9 (10) :1107-1110
[2]   DETERMINATION OF ARSENIC BY HYDRIDE GENERATION INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY USING A TUBULAR MEMBRANE GAS-LIQUID SEPARATOR [J].
BRANCH, S ;
CORNS, WT ;
EBDON, L ;
HILL, S ;
ONEILL, P .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1991, 6 (02) :155-158
[3]   DETERMINATION OF SELENIUM BY INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY UTILIZING A NEW HYDRIDE GENERATION SAMPLE INTRODUCTION SYSTEM [J].
BUCKLEY, WT ;
BUDAC, JJ ;
GODFREY, DV ;
KOENIG, KM .
ANALYTICAL CHEMISTRY, 1992, 64 (07) :724-729
[4]   DETERMINATION OF MULTIPLE SELENIUM STABLE ISOTOPE TRACERS BY INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
BUCKLEY, WT ;
BUDAC, JJ ;
GODFREY, DV ;
KOENIG, KM .
BIOLOGICAL MASS SPECTROMETRY, 1992, 21 (10) :473-478
[5]   DETERMINATION OF ARSENIC, CHROMIUM, SELENIUM AND VANADIUM IN BIOLOGICAL SAMPLES BY INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETRY USING ONLINE ELIMINATION OF INTERFERENCE AND PRECONCENTRATION BY FLOW-INJECTION [J].
EBDON, L ;
FISHER, AS ;
WORSFOLD, PJ .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1994, 9 (05) :611-614
[6]   CONTINUOUS HYDRIDE GENERATION FOR SIMULTANEOUS MULTIELEMENT DETECTION WITH INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
HEITKEMPER, DT ;
CARUSO, JA .
APPLIED SPECTROSCOPY, 1990, 44 (02) :228-234
[7]   COMPARISON OF PNEUMATIC NEBULIZATION AND HYDRIDE GENERATION INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY FOR ISOTOPIC ANALYSIS OF SELENIUM [J].
JANGHORBANI, M ;
TING, BTG .
ANALYTICAL CHEMISTRY, 1989, 61 (07) :701-708
[8]  
Narasaki H, 1996, ATOM SPECTROSC, V17, P77
[9]   INTRODUCTION OF GASEOUS HYDRIDES INTO AN INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETER [J].
POWELL, MJ ;
BOOMER, DW ;
MCVICARS, RJ .
ANALYTICAL CHEMISTRY, 1986, 58 (13) :2864-2867
[10]   ELEMENTAL SPECIATION BY ANION-EXCHANGE AND SIZE-EXCLUSION CHROMATOGRAPHY WITH DETECTION BY INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETRY WITH DIRECT-INJECTION NEBULIZATION [J].
SHUM, SCK ;
HOUK, RS .
ANALYTICAL CHEMISTRY, 1993, 65 (21) :2972-2976