NMP-Modified PMMA Bone Cement with Adapted Mechanical and Hardening Properties for the Use in Cancellous Bone Augmentation

被引:28
|
作者
Boger, Andreas [1 ]
Wheeler, Kurtis [1 ]
Montali, Andrea [1 ]
Gruskin, Elliot [2 ]
机构
[1] Synthes GmbH, R&D Biomat, CH-4436 Oberdorf, Switzerland
[2] Synthes Inc, R&D Biomat, W Chester, PA 19380 USA
关键词
PMMA bone cement; stiffness adaptation; NMP; polymerization temperature; viscosity; VERTEBROPLASTY; FRACTURES;
D O I
10.1002/jbm.b.31345
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The use of poly methyl methacrylate (PMMA) to reinforce vertebral bodies (Vertebroplasty) leads to an increase in the Young's modulus of the augmented vertebral body. Fractures in the adjacent vertebrae may be the consequence thereof. Hence, PMMA with a reduced Young's modulus may be suitable for vertebroplasty. The goal of this study was to produce and characterize stiffness-adapted PMMA cements. Modified PMMA bone cements were produced by adding N-methyl-pyrrolidone (NMP). Young's modulus, yield strength, polymerization temperature, setting time, and hardening behavior of different cements were analyzed. Focus was on the mechanical properties of the material after different storage conditions (in air at room temperature and in PBS at 37 degrees C). The Young's modulus decreased from 2670 MPa (air)/2384 MPa (PBS) for the regular cement to 76 MPa (air)/320 MPa (PBS) for a material composition with 60% of the MMA substituted by NMIP. Yield strength decreased from 85 MPa (air)/78 MPa (PBS) to 2 MPa (air)/24 MPa (PBS) between the regular cement and the 60% composition. Polymerization temperature decreased from 70 degrees C (regular cement) to 48 degrees C for the 30% composition. The hardening behavior exhibited an extension in handling time up to 200% by the modification presented. Modification of PMMA cement using NMP seems to be a promising method to make the PMMA cement more compliant for the use in cancellous bone augmentation in osteoporotic patients: adjustment of its mechanical properties close to those of cancellous bone, lower polymerization temperature, and extended handling time. (C) 2009 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater 90B: 760-766, 2009
引用
收藏
页码:760 / 766
页数:7
相关论文
共 44 条
  • [41] Evaluation of the Effect of Selected Physiological Fluid Contaminants on the Mechanical Properties of Selected Medium-Viscosity PMMA Bone Cements
    Karpinski, Robert
    Szabelski, Jakub
    Krakowski, Przemyslaw
    Jojczuk, Mariusz
    Jonak, Jozef
    Nogalski, Adam
    MATERIALS, 2022, 15 (06)
  • [42] Mechanical properties and antibiotic release characteristics of poly(methyl methacrylate)-based bone cement formulated with mesoporous silica nanoparticles
    Letchmanan, Kumaran
    Shen, Shou-Cang
    Ng, Wai Kiong
    Kingshuk, Poddar
    Shi, Zhilong
    Wang, Wilson
    Tan, Reginald B. H.
    JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2017, 72 : 163 - 170
  • [43] Influence of N-acetyl cysteine (NAC) and 2-methylene-1,3-dioxepane (MDO) on the properties of polymethyl methacrylate (PMMA) bone cement
    Zhao, Kangquan
    Pi, Bin
    Zhao, Liping
    Tian, Shoujin
    Ge, Jianfei
    Yang, Huilin
    Sha, Weiping
    Wang, Liming
    RSC ADVANCES, 2019, 9 (21) : 11833 - 11841
  • [44] Long-term mechanical properties of a novel low-modulus bone cement for the treatment of osteoporotic vertebral compression fractures
    Robo, Celine
    Ohman-Magi, Caroline
    Persson, Cecilia
    JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2021, 118