摘要
The use of ankle external fixator to treat pilon fracture Type III is popular amongst surgeons as it can reduce complications such as non-union and mal-union. Even though configurations of the connecting bars are important, the material also plays a major factor for a successful outcome. In this paper, the Delta external fixator with simulated ankle pilon fractures Type III were modelled and analysed under two different materials; titanium alloy and stainless steel. The finite element model includes tibia, fibula, talus, calcaneus, cuboid, navicular, three cuneiforms and five metatarsals bone. To simulate the pilon fractures Type III, a cutting segment was utilised. The ligaments were assigned with linear spring properties and cartilages were modelled using Mooney-Rivlin hyper-elastic behaviour. The Delta external fixator was designed using a three-dimensional software with two different material properties - titanium alloy and stainless steel. High von Mises stress concentrated at the pin-bone interface with the highest value observed for the titanium fixation. The results also showed less deformation for the stainless steel compared to titanium.
引用
收藏
页码:183 / 188
页数:6
相关论文
共 22 条
[11]   Bio-mechanical analysis of a fractured tibia with composite bone plates according to the diaphyseal oblique fracture angle [J].
Kim, Hyun-Jun ;
Kim, Suk-Hun ;
Chang, Seung-Hwan .
COMPOSITES PART B-ENGINEERING, 2011, 42 (04) :666-674
[12]   Biomechanical response of the pubic symphysis in lateral pelvic impacts: A finite element study [J].
Li, Zuoping ;
Kim, Jong-Eun ;
Davidson, James S. ;
Etheridge, Brandon S. ;
Alonso, Jorge E. ;
Eberhardt, Alan W. .
JOURNAL OF BIOMECHANICS, 2007, 40 (12) :2758-2766
[13]   Computational modeling to predict mechanical function of joints: Application to the lower leg with simulation of two cadaver studies [J].
Liacouras, Peter C. ;
Waynel, Jennifer S. .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2007, 129 (06) :811-817
[14]  
Nakamura S, 1981, Bull Prosthet Res, V10-35, P27
[15]  
Nash RA, 2001, AM J VET RES, V62, P955, DOI 10.2460/ajvr.2001.62.955
[16]  
Netter F.H., 2003, Atlas of Human Anatomy, V3rd
[17]   Tensile properties of the interosseous membrane of the human forearm [J].
Pfaeffle, HJ ;
Tomaino, MM ;
Grewal, R ;
Xu, JM ;
Boardman, ND ;
Woo, SLY ;
Herndon, JH .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1996, 14 (05) :842-845
[18]   THE MECHANICAL CHARACTERISTICS OF THE COLLATERAL LIGAMENTS OF THE HUMAN ANKLE JOINT [J].
SIEGLER, S ;
BLOCK, J ;
SCHNECK, CD .
FOOT & ANKLE, 1988, 8 (05) :234-242
[19]  
Teeny S, 1990, ORTHOP T, V14, P265
[20]   Finite element analysis of the initial stability of ankle arthrodesis with internal fixation:: flat cut versus intact joint contours [J].
Vázquez, AA ;
Lauge-Pedersen, H ;
Lidgren, L ;
Taylor, M .
CLINICAL BIOMECHANICS, 2003, 18 (03) :244-253