Accuracy and preference of measuring resting energy expenditure using a handheld calorimeter in healthy adults

被引:6
作者
Madden, A. M. [1 ]
Parker, L. J. F. [1 ]
Amirabdollahian, F. [2 ]
机构
[1] Univ Hertfordshire, Dept Human & Environm Sci, Hatfield AL10 9AB, Herts, England
[2] Liverpool Hope Univ, Dept Hlth Sci, Liverpool, Merseyside, England
关键词
energy requirements; nutritional assessment; nutrition support; prediction equations; resting energy expenditure; METABOLIC-RATE MEASUREMENTS; PREDICTION EQUATIONS; OXYGEN-CONSUMPTION; VENTILATED-HOOD; CANCER-PATIENTS; FACE-MASK; VALIDATION; CHILDREN; VALIDITY; DEVICE;
D O I
10.1111/jhn.12045
中图分类号
R15 [营养卫生、食品卫生]; TS201 [基础科学];
学科分类号
100403 ;
摘要
BackgroundAccurate estimates of energy expenditure are required in clinical nutrition in order to determine the requirements of individuals and to inform feeding regimes. Calorimetry can provide accurate measurements but is often impractical in clinical or community settings; prediction equations are widely used to estimate resting energy expenditure (REE) but have limited accuracy. A portable, self-calibrating, handheld calorimeter (HHC) may offer an alternative way of determining REE. The aim of the study was to evaluate whether estimates of REE derived using an HHC are closer to accurate measurements than values calculated using selected prediction equations. MethodsREE was measured in 36 healthy adults aged 21-58years using a flow-through indirect calorimeter (FIC) and HHC. Estimated REE was calculated using three predictive equations (Harris & Benedict; Schofield; Henry). Differences in REE between the gold standard' values derived using the FIC and those derived using the HHC and equations were examined using paired t-tests and Bland Altman plots. ResultsMean REEHHC was significantly lower than mean REEFIC [45561042kJ (1089 +/- 249kcal) versus 6230 +/- 895kJ (1489 +/- 214 kcal), P=0.000] and also significantly lower than mean values calculated using all three equations. The mean difference between REEHHC and REEFIC [1674 +/- 908kJ (400 +/- 217kcal)] was significantly greater (P=0.000) than the mean differences between the values calculated using the three prediction equations [272 +/- 490kJ (65 +/- 117kcal) (Harris-Benedict), 264 +/- 510 kJ (63 +/- 122kcal) (Schofield), 84 +/- 502kJ (20 +/- 120kcal) (Henry)]. ConclusionsThe HHC provides estimates of REE in healthy people that are less accurate than those calculated using the prediction equations and so does not provide a useful alternative.
引用
收藏
页码:587 / 595
页数:9
相关论文
共 45 条
[1]   Validity and reproducibility of resting metabolic rate measurements in rural Bangladeshi women:: comparison of measurements obtained by Medgem™ and by Deltatrac™ device [J].
Alam, DS ;
Hulshof, PJM ;
Roordink, D ;
Meltzer, M ;
Yunus, M ;
Salam, MA ;
van Raaij, JMA .
EUROPEAN JOURNAL OF CLINICAL NUTRITION, 2005, 59 (05) :651-657
[2]  
Almajwal AM, 2011, SAUDI MED J, V32, P718
[3]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]   COMPARISON OF A FACE-MASK AND A MOUTHPIECE FOR MEASURING RESTING ENERGY-EXPENDITURE WITH THE OXYLOG [J].
BUKKENS, SGF ;
MCNEILL, G .
ANNALS OF NUTRITION AND METABOLISM, 1990, 34 (02) :112-118
[5]   Comparison between Medgem and Deltatrac resting metabolic rate measurements [J].
Compher, C ;
Hise, M ;
Sternberg, A ;
Kinosian, BP .
EUROPEAN JOURNAL OF CLINICAL NUTRITION, 2005, 59 (10) :1136-1141
[6]   Assessing Validity and Reliability of Resting Metabolic Rate in Six Gas Analysis Systems [J].
Cooper, Jamie A. ;
Watras, Abigail C. ;
O'Brien, Matthew J. ;
Luke, Amy ;
Dobratz, Jennifer R. ;
Earthman, Carrie P. ;
Schoeller, Dale A. .
JOURNAL OF THE AMERICAN DIETETIC ASSOCIATION, 2009, 109 (01) :128-132
[7]  
Dulloo A. G., 2011, HUMAN NUTR, P87
[8]   Measuring energy expenditure in community-dwelling older adults: Are portable methods valid and acceptable? [J].
Fares, Samira ;
Miller, Michelle D. ;
Masters, Stacey ;
Crotty, Maria .
JOURNAL OF THE AMERICAN DIETETIC ASSOCIATION, 2008, 108 (03) :544-548
[9]   MedGem hand-held indirect calorimeter is valid for resting energy expenditure measurement in healthy children [J].
Fields, David A. ;
Kearney, Jay T. ;
Copeland, Kenneth C. .
OBESITY, 2006, 14 (10) :1755-1761
[10]   Handheld Calorimeter Is a Valid Instrument to Quantify Resting Energy Expenditure in Hospitalized Cirrhotic Patients: A Prospective Study [J].
Glass, Cathy ;
Hipskind, Peggy ;
Cole, Denise ;
Lopez, Rocio ;
Dasarathy, Srinivasan .
NUTRITION IN CLINICAL PRACTICE, 2012, 27 (05) :677-688