Two reconstruction plates provide superior stability of displaced midshaft clavicle fractures in comparison to single plating - A biomechanical study

被引:8
作者
Yanev, Parvan [1 ,2 ]
Zderic, Ivan [1 ]
Pukalski, Yavor [1 ,2 ]
Enchev, Dian [2 ]
Rashkov, Mihail [2 ]
Varga, Peter [1 ]
Gehweiler, Dominic [1 ]
Richards, Geoff [1 ]
Gueorguiev, Boyko [1 ]
Baltov, Asen [2 ]
机构
[1] AO Res Inst Davos, Davos, Switzerland
[2] Univ Multiprofile Hosp Act Treatment & Emergency, Sofia, Bulgaria
关键词
Clavicle; Midshaft fracture; Dynamic compression plate; Locking compression plate; Reconstruction plate; Biomechanical testing; FINITE-ELEMENT; LOCKING PLATE; FIXATION; OSTEOSYNTHESIS; MIGRATION;
D O I
10.1016/j.clinbiomech.2020.105199
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background: Displaced midshaft fractures are the most common surgically treated clavicle fractures. However, they are associated with high complication rates following plating due to fixation failure in terms of plate breakage, screw breakage and/or screw loosening. The aim of this study was to compare the biomechanical competence of three different plating techniques for fixation of displaced midshaft clavicle fractures. Methods: Displaced midshaft fractures type 2B according to the Robinson classification were simulated by standardized osteotomy gap in 18 synthetic clavicles, assigned to three groups (n = 6) for plating with either superiorly placed Dynamic Compression Plate (width/thickness 11.0/4.0 mm), locked Superior Anterior Clavicle Locking Compression Plate (width/thickness 10.2/2.0 mm), or two non-locked Reconstruction Plates placed superiorly and anteriorly (width/thickness 10.0/2.8 mm). Each specimen was cyclically tested at 3 Hz under craniocaudal cantilever bending, superimposed with torsion around the shaft axis over 720'000 cycles or until failure occurred. The latter was defined by plate breakage, screw breakage or screw loosening. Findings: Initial construct stiffness (N/mm) and cycles to failure in group Reconstruction Plates (22.30 +/- 4.07; 712'778 +/- 17'691) were significantly higher compared with both groups Compression Plate (12.53 +/- 2.09; 348'541 +/- 212'941) and Locking Plate (4.19 +/- 0.46; 19'536 +/- 3'586), p <= 0.019. In addition, these two outcomes were significantly higher in group Compression Plate versus Locking Plate, p <= 0.029. Interpretation: Double plating of unstable midshaft clavicle fractures with reconstruction plates seems to provide superior fixation stability under dynamic loading, when compared to single compression or locked plating, whereas the latter is associated with inferior performance.
引用
收藏
页数:6
相关论文
共 35 条
  • [1] Are clinical outcomes affected by type of plate used for management of mid-shaft clavicle fractures?
    Alzahrani, Mohammad M.
    Cota, Adam
    Alkhelaifi, Khalid
    Aleidan, Aljarrah
    Berry, Gregory
    Reindl, Rudy
    Harvey, Edward
    [J]. JOURNAL OF ORTHOPAEDICS AND TRAUMATOLOGY, 2018, 19 (01)
  • [2] Complications of plate fixation of fresh displaced midclavicular fractures
    Bostman, O
    Manninen, M
    Pihlajamaki, H
    [J]. JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 1997, 43 (05) : 778 - 783
  • [3] Unicortical Versus Bicortical Locked Plate Fixation in Midshaft Clavicle Fractures
    Bravman, Jonathan T.
    Taylor, Michal L.
    Baldini, Todd
    Vidal, Armando F.
    [J]. ORTHOPEDICS, 2015, 38 (05) : E411 - E416
  • [4] Biomechanical evaluation of clavicle fracture plating techniques: Does a locking plate provide improved stability?
    Celestre, Paul
    Roberston, Claire
    Mahar, Andrew
    Oka, Richard
    Meunier, Matthew
    Schwartz, Alexandra
    [J]. JOURNAL OF ORTHOPAEDIC TRAUMA, 2008, 22 (04) : 241 - 247
  • [5] A biomechanical comparison of four different fixation methods for midshaft clavicle fractures
    Chen, Yang
    Yang, Yang
    Ma, Xinlong
    Xu, Weiguo
    Ma, Jianxiong
    Zhu, Shaowen
    Ma, Baoyi
    Xing, Dan
    [J]. PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2016, 230 (01) : 13 - 19
  • [6] Fracture toughness and fatigue crack propagation rate of short fiber reinforced epoxy composites for analogue cortical bone
    Chong, Alexander C. M.
    Miller, Forrest
    Buxton, Mckee
    Friis, Elizabeth A.
    [J]. JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2007, 129 (04): : 487 - 493
  • [7] Combined finite element and multibody musculoskeletal investigation of a fractured clavicle with reconstruction plate
    Cronskar, Marie
    Rasmussen, John
    Tinnsten, Mats
    [J]. COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING, 2015, 18 (07) : 740 - 748
  • [8] Biomechanical comparison of the Locking Compression superior anterior clavicle plate with seven and ten hole reconstruction plates in midshaft clavicle fracture stabilisation
    Eden, Lars
    Doht, Stefanie
    Frey, Soenke P.
    Ziegler, Dirk
    Stoyhe, Jan
    Fehske, Kai
    Blunk, Torsten
    Meffert, Rainer H.
    [J]. INTERNATIONAL ORTHOPAEDICS, 2012, 36 (12) : 2537 - 2543
  • [9] Composite Bone Models in Orthopaedic Surgery Research and Education
    Elfar, John J.
    Menorca, Ron Martin Garcia
    Reed, Jeffrey Douglas
    Stanbury, Spencer
    [J]. JOURNAL OF THE AMERICAN ACADEMY OF ORTHOPAEDIC SURGEONS, 2014, 22 (02) : 111 - 120
  • [10] A biomechanical study of plate versus intramedullary devices for midshaft clavicle fixation
    Golish S.R.
    Oliviero J.A.
    Francke E.I.
    Miller M.D.
    [J]. Journal of Orthopaedic Surgery and Research, 3 (1)