Integral-based filtering of continuous glucose sensor measurements for glycaemic control in critical care

被引:30
作者
Chase, J. Geoffrey
Hann, Christopher E.
Jackson, Monique
Lin, Jessica
Lotz, Thomas
Wong, Xing-Wei
Shaw, Geoffrey M.
机构
[1] Univ Canterbury, Ctr Bioengn, Dept Mech Engn, Christchurch, New Zealand
[2] Univ Otago, Dept Intens Care, Christchurch Hosp, Christchurch Sch Med & Hlth Sci, Christchurch, New Zealand
关键词
critical care; glucose; hyperglycemia; CGMS; parameter identification;
D O I
10.1016/j.cmpb.2006.03.004
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
Hyperglycaemia is prevalent in critical illness and increases the risk of further complications and mortality, while tight control can reduce mortality up to 43%. Adaptive control methods are capable of highly accurate, targeted blood glucose regulation using limited numbers of manual measurements due to patient discomfort and labour intensity. Therefore, the option to obtain greater data density using emerging continuous glucose sensing devices is attractive. However, the few such systems currently available can have errors in excess of 20-30%. In contrast, typical bedside testing kits have errors of approximately 7-10%. Despite greater measurement frequency larger errors significantly impact the resulting glucose and patient specific parameter estimates, and thus the control actions determined creating an important safety and performance issue. This paper models the impact of the continuous glucose monitoring system (CGMS, Medtronic, Northridge, CA) on model-based parameter identification and glucose prediction. An integral-based fitting and filtering method is developed to reduce the effect of these errors. A noise model is developed based on CGMS data reported in the literature, and is slightly conservative with a mean Clarke Error Grid (CEG) correlation of R = 0.81 (range: 0.68-0.88) as compared to a reported value of R = 0.82 in a critical care study. Using 17 virtual patient profiles developed from retrospective clinical data, this noise model was used to test the methods developed. Monte-Carlo simulation for each patient resulted in an average absolute 1-h glucose prediction error of 6.20% (range: 4.97-8.06%) with an average standard deviation per patient of 5.22% (range: 3.26-8.55%). Note that all the methods and results are generalisable to similar applications outside of critical care, such as less acute wards and eventually ambulatory individuals. Clinically, the results show one possible computational method for managing the larger errors encountered in emerging continuous blood glucose sensors, thus enabling their more effective use in clinical glucose regulation studies. (c) 2006 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:238 / 247
页数:10
相关论文
共 30 条
  • [1] CLINICAL CONTROL OF DIABETES BY ARTIFICIAL PANCREAS
    ALBISSER, AM
    LEIBEL, BS
    EWART, TG
    DAVIDOVAC, Z
    BOTZ, CK
    ZINGG, W
    SCHIPPER, H
    GANDER, R
    [J]. DIABETES, 1974, 23 (05) : 397 - 404
  • [2] Hyperglycemia and infection: Which is the chicken and which is the egg?
    Bistrian, BR
    [J]. JOURNAL OF PARENTERAL AND ENTERAL NUTRITION, 2001, 25 (04) : 180 - 181
  • [3] BUCKINGHAM B, 2005, DIABETES TECHNOL THE, V7, P347
  • [4] Stress hyperglycaemia and increased risk of death after myocardial infarction in patients with and without diabetes: a systematic overview
    Capes, SE
    Hunt, D
    Malmberg, K
    Gerstein, HC
    [J]. LANCET, 2000, 355 (9206) : 773 - 778
  • [5] Chase J. G., 2005, International Journal of Intelligent Systems Technologies and Applications, V1, P79, DOI 10.1504/IJISTA.2005.007308
  • [6] Adaptive bolus-based targeted glucose regulation of hyperglycaemia in critical care
    Chase, JG
    Shaw, GM
    Lin, J
    Doran, CV
    Hann, C
    Robertson, MB
    Browne, PM
    Lotz, T
    Wake, GC
    Broughton, B
    [J]. MEDICAL ENGINEERING & PHYSICS, 2005, 27 (01) : 1 - 11
  • [7] CHASE JG, 2005, P 12 INT C BIOM ENG
  • [8] CHASE JG, P 26 INT C IEEE ENG, P3463
  • [9] Closed-loop glucose control in critically ill patients using continuous glucose monitoring system (CGMS) in real time
    Chee, F
    Fernando, T
    van Heerden, PV
    [J]. IEEE TRANSACTIONS ON INFORMATION TECHNOLOGY IN BIOMEDICINE, 2003, 7 (01): : 43 - 53
  • [10] EVALUATION OF A NEW SYSTEM FOR SELF BLOOD-GLUCOSE MONITORING
    CLARKE, WL
    BECKER, DJ
    COX, D
    SANTIAGO, JV
    WHITE, NH
    BETSCHART, J
    ECKENRODE, K
    LEVANDOSKI, LA
    PRUSINSKI, EA
    SIMINEIRO, LM
    SNYDER, AL
    TIDEMAN, AM
    YAEGER, T
    [J]. DIABETES RESEARCH AND CLINICAL PRACTICE, 1988, 4 (03) : 209 - 214