Effects of transducer frequency and output power on the ultrasonographic contrast produced by Optison using fundamental and harmonic imaging techniques

被引:9
作者
Wible, JH [1 ]
Wojdyla, JK [1 ]
Hughes, MS [1 ]
Brandenburger, GH [1 ]
机构
[1] Mallinckrodt Inc, St Louis, MO 63134 USA
关键词
ultrasonography; contrast agents; output power; transducer; frequency; perfusion; chamber opacification; Optison;
D O I
10.7863/jum.1999.18.11.753
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
This in vivo study demonstrated the effects of transducer frequency and output power on the sonographic contrast produced by Optison during continuous ultrasonographic imaging in both fundamental and harmonic modes. During fundamental imaging, Optison produced greater contrast intensity while imaging at higher transducer frequencies. However, in the harmonic mode, lower imaging frequencies caused Optison to produce more intense myocardial enhancement. Regardless of imaging mode, increases in transducer output power caused decreases in both contrast intensity and contrast duration produced by Optison. Proper selection of transducer frequency and output power will help optimize the quality and duration of contrast produced by Optison during continuous ultrasonographic imaging.
引用
收藏
页码:753 / 762
页数:10
相关论文
共 18 条
[1]  
Burns PN, 1996, CLIN RADIOL, V51, P50
[2]  
CHEIRIF J, 1997, J AM SOC ECHOCARDIOG, V10, P390
[3]   Improved left ventricular endocardial border delineation and opacification with OPTISON (FS069), a new echocardiographic contrast agent - Results of a phase III multicenter trial [J].
Cohen, JL ;
Cheirif, J ;
Segar, DS ;
Gillam, LD ;
Gottdiener, JS ;
Hausnerova, E ;
Bruns, DE .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1998, 32 (03) :746-752
[4]  
Dittrich H C, 1995, J Am Soc Echocardiogr, V8, P465, DOI 10.1016/S0894-7317(05)80333-5
[5]  
Kaul S, 1997, CIRCULATION, V96, P785
[6]   Direct video microscopic observation of the dynamic effects of medical ultrasound on ultrasound contrast microspheres [J].
Klibanov, AL ;
Ferrara, KW ;
Hughes, MS ;
Wible, JH ;
Wojdyla, JK ;
Dayton, PA ;
Morgan, KE ;
Brandenburger, GH .
INVESTIGATIVE RADIOLOGY, 1998, 33 (12) :863-870
[7]   Frequency and concentration dependence of the backscatter coefficient of the ultrasound contrast agent Albunex® [J].
Marsh, JN ;
Hughes, MS ;
Hall, CS ;
Lewis, SH ;
Trousil, RL ;
Brandenburger, GH ;
Levene, H ;
Miller, JG .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1998, 104 (03) :1654-1666
[8]   Myocardial contrast echocardiography: Reliable, safe, and efficacious myocardial perfusion assessment after intravenous injections of a new echocardiographic contrast agent [J].
Meza, M ;
Greener, Y ;
Hunt, R ;
Perry, B ;
Revall, S ;
Barbee, W ;
Murgo, JP ;
Cheirif, J .
AMERICAN HEART JOURNAL, 1996, 132 (04) :871-881
[9]   Relation of contrast echo intensity and flow velocity to the amplification of contrast opacification produced by intermittent ultrasound transmission [J].
Ohmori, K ;
Cotter, B ;
Kwan, OL ;
Mizushige, K ;
DeMaria, AN .
AMERICAN HEART JOURNAL, 1997, 134 (06) :1066-1074
[10]   Detection of myocardial perfusion in multiple echocardiographic windows with one intravenous injection of microbubbles using transient response second harmonic imaging [J].
Porter, TR ;
Li, SP ;
Kricsfeld, D ;
Armbruster, RW .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1997, 29 (04) :791-799