An evaporation-based disposable micropump concept for continuous monitoring applications

被引:52
作者
Effenhauser, CS [1 ]
Harttig, H [1 ]
Krämer, P [1 ]
机构
[1] Roche Diagnost GmbH, Roche Diabet Care, New Technol DD T, D-68305 Mannheim, Germany
关键词
microfluidics; micropumps; continuous monitoring;
D O I
10.1023/A:1014215728074
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
An inexpensive and simple pumping principle is described that is capable of delivering both small and constant flow rates (10-1,000 nl/min) over a longer period of time (days to weeks). The concept is based on controlled evaporation of a liquid through a membrane into a gas space containing a sorption agent. As long as the sorption agent keeps the vapor pressure in the gas phase below saturation, fluid evaporated from the membrane is replaced by capillary forces inducing flow from a reservoir. In a feasibility study, a total volume of 300 mul of Ringer's solution has been continuously pumped over a period of six days, resulting in an constant average flow rate of 35 nl/min (590 pl/s). The maximum liquid volume transported is limited by sorption capacity and amount of the sorption agent. Low fabrication costs, high reliability (no moving parts), the suitability for integration into planar system architectures and the lack of a special external energy source besides an environment of regulated temperature are important features of the concept, in particular with regard to its potential application in continuous patient monitoring. Truly continuous flow can be achieved in contrast to many other pump mechanisms leading to discontinuous, pulse-type flow. A challenge for a broader range of applications is the inherent temperature dependence of the flow rate. In its current version, the pump can only be used in a suction-mode.
引用
收藏
页码:27 / 32
页数:6
相关论文
共 20 条
  • [1] A valve-less diffuser micropump for microfluidic analytical systems
    Andersson, H
    van der Wijngaart, W
    Nilsson, P
    Enoksson, P
    Stemme, G
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 2001, 72 (03) : 259 - 265
  • [2] Bedside Clinical Chemistry: From Catheter Tip Sensor Chips Towards Micro Total Analysis Systems
    Bergveld, P.
    [J]. BIOMEDICAL MICRODEVICES, 2000, 2 (03) : 185 - 195
  • [3] An Integrated Micromachined Electrochemical Pump and Dosing System
    Bohm, S.
    Olthuis, W.
    Bergveld, P.
    [J]. BIOMEDICAL MICRODEVICES, 1999, 1 (02) : 121 - 130
  • [4] Nanopore Technology for Biomedical Applications
    Desai, Tejal A.
    Hansford, Derek J.
    Kulinsky, Lawrence
    Nashat, Amir H.
    Rasi, Guido
    Tu, Jay
    Wang, Yuchun
    Zhang, Miqin
    Ferrari, Mauro
    [J]. BIOMEDICAL MICRODEVICES, 1999, 2 (01) : 11 - 40
  • [5] Effenhauser CS, 1998, TOP CURR CHEM, V194, P51
  • [6] TOWARDS INTEGRATED MICROLIQUID HANDLING SYSTEMS
    ELWENSPOEK, M
    LAMMERINK, TSJ
    MIYAKE, R
    FLUITMAN, JHJ
    [J]. JOURNAL OF MICROMECHANICS AND MICROENGINEERING, 1994, 4 (04) : 227 - 245
  • [7] Effects of glucose and insulin levels on adipose tissue glucose measurement by microdialysis probes retained for three weeks in Type 1 diabetic patients
    Hullegie, LM
    Lutgers, HL
    Dullaart, RPF
    Sluiter, WJ
    Wientjes, KJ
    Schoonen, AJM
    Hoogenberg, K
    [J]. NETHERLANDS JOURNAL OF MEDICINE, 2000, 57 (01) : 13 - 19
  • [8] Continuous ultraslow microdialysis and ultrafiltration for subcutaneous sampling as demonstrated by glucose and lactate measurements in rats
    Kaptein, WA
    Zwaagstra, JJ
    Venema, K
    Korf, J
    [J]. ANALYTICAL CHEMISTRY, 1998, 70 (22) : 4696 - 4700
  • [9] Current developments in electrophoretic and chromatographic separation methods on microfabricated devices
    Kutter, JP
    [J]. TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 2000, 19 (06) : 352 - 363
  • [10] Microdialysis measurement of glucose in subcutaneous adipose tissue up to three weeks in Type 1 diabetic patients
    Lutgers, HL
    Hullegie, LM
    Hoogenberg, K
    Sluiter, WJ
    Dullaart, RPF
    Wientjes, KJ
    Schoonen, AJM
    [J]. NETHERLANDS JOURNAL OF MEDICINE, 2000, 57 (01) : 7 - 12