Fractionated stereotactic radiotherapy: A method to evaluate geometric and dosimetric uncertainties using radiochromic films

被引:17
作者
Coscia, Gianluca [1 ]
Vaccara, Elena [1 ]
Corvisiero, Roberta [1 ]
Cavazzani, Paolo [2 ]
Ruggieri, Filippo Grillo [2 ]
Taccini, Gianni [1 ]
机构
[1] EO Osped Galliera Genova, SC Fis Sanit, I-16128 Genoa, Italy
[2] EO Osped Galliera Genova, SC Radioterapia, I-16128 Genoa, Italy
关键词
brain; collimators; computerised tomography; dosimetry; error analysis; image registration; linear accelerators; masks; medical image processing; neurophysiology; phantoms; radiation therapy; REPOSITIONING ACCURACY; RADIATION-THERAPY; EBT FILM; BRAIN; HEAD; SYSTEM; FRAME; BEAM; REPRODUCIBILITY; ALIGNMENT;
D O I
10.1118/1.3134246
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In the authors' hospital, stereotactic radiotherapy treatments are performed with a Varian Clinac 600C equipped with a BrainLAB m3 micro-multileaf-collimator generally using the dynamic conformal arc technique. Patient immobilization during the treatment is achieved with a fixation mask supplied by BrainLAB, made with two reinforced thermoplastic sheets fitting the patient's head. With this work the authors propose a method to evaluate treatment geometric accuracy and, consequently, to determine the amount of the margin to keep in the CTV-PTV expansion during the treatment planning. The reproducibility of the isocenter position was tested by simulating a complete treatment on the anthropomorphic phantom Alderson Rando, inserting in between two phantom slices a high sensitivity Gafchromic EBT film, properly prepared and calibrated, and repeating several treatment sessions, each time removing the fixing mask and replacing the film inside the phantom. The comparison between the dose distributions measured on films and computed by TPS, after a precise image registration procedure performed by a commercial piece of software (FILMQA, 3cognition LLC (Division of ISP), Wayne, NJ), allowed the authors to measure the repositioning errors, obtaining about 0.5 mm in case of central spherical PTV and about 1.5 mm in case of peripheral irregular PTV. Moreover, an evaluation of the errors in the registration procedure was performed, giving negligible values with respect to the quantities to be measured. The above intrinsic two-dimensional estimate of treatment accuracy has to be increased for the error in the third dimension, but the 2 mm margin the authors generally use for the CTV-PTV expansion seems adequate anyway. Using the same EBT films, a dosimetric verification of the treatment planning system was done. Measured dose values are larger or smaller than the nominal ones depending on geometric irradiation conditions, but, in the authors' experimental conditions, always within 4%.
引用
收藏
页码:2870 / 2880
页数:11
相关论文
共 24 条
[1]   Repositioning accuracy of fractionated stereotactic irradiation: assessment of isocentre alignment for different dental fixations by using sequential CT scanning [J].
Baumert, BG ;
Egli, P ;
Studer, S ;
Dehing, C ;
Davis, JB .
RADIOTHERAPY AND ONCOLOGY, 2005, 74 (01) :61-66
[2]   Repositioning accuracy of two different mask systems - 3D revisited: Comparison using true 3D/3D matching with cone-beam CT [J].
Boda-Heggemann, Judit ;
Walter, Cornelia ;
Rahn, Angelika ;
Wertz, Hansjoerg ;
Loeb, Iris ;
Lohr, Frank ;
Wenz, Frederik .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2006, 66 (05) :1568-1575
[3]   Post-irradiation colouration of Gafchromic EBT radiochromic film [J].
Cheung, T ;
Butson, MJ ;
Yu, PKN .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (20) :N281-N285
[4]   Precise radiochromic film dosimetry using a flat-bed document scanner [J].
Devic, S ;
Seuntjens, J ;
Sham, E ;
Podgorsak, EB ;
Schmidtlein, CR ;
Kirov, AS ;
Soares, CG .
MEDICAL PHYSICS, 2005, 32 (07) :2245-2253
[5]   Predicting error in rigid-body point-based registration [J].
Fitzpatrick, JM ;
West, JB ;
Maurer, CR .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1998, 17 (05) :694-702
[6]   Repositioning accuracy of a commercially available thermoplastic mask system [J].
Fuss, M ;
Salter, BJ ;
Cheek, D ;
Sadeghi, A ;
Hevezi, JM ;
Herman, TS .
RADIOTHERAPY AND ONCOLOGY, 2004, 71 (03) :339-345
[7]   Comparison of setup accuracy of three different thermoplastic masks for the treatment of brain and head and neck tumors [J].
Gilbeau, L ;
Octave-Prignot, M ;
Loncol, T ;
Renard, L ;
Scalliet, P ;
Grégoire, V .
RADIOTHERAPY AND ONCOLOGY, 2001, 58 (02) :155-162
[8]   2D/3D image fusion for accurate target localization and evaluation of a mask based stereotactic system in fractionated stereotactic radiotherapy of cranial lesions [J].
Jin, Jian-Yue ;
Ryu, Samuel ;
Faber, Kathleen ;
Mikkelsen, Tom ;
Chen, Qing ;
Li, Shidong ;
Movsas, Benjamin .
MEDICAL PHYSICS, 2006, 33 (12) :4557-4566
[9]   Repositioning accuracy with the Laitinen frame for fractionated stereotactic radiation therapy in adult and pediatric brain tumors: Preliminary report [J].
Kalapurakal, JA ;
Ilahi, Z ;
Kepka, AG ;
Bista, T ;
Goldman, S ;
Tomita, T ;
Marymont, MH .
RADIOLOGY, 2001, 218 (01) :157-161
[10]   Three-dimensional accuracy and interfractional reproducibility of patient fixation and positioning using a stereotactic head mask system [J].
Karger, CP ;
Jäkel, O ;
Debus, J ;
Kuhn, S ;
Hartmann, GH .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2001, 49 (05) :1493-1504