A cryocooled helmet-shaped MEG measurement system

被引:4
作者
Sata, K [1 ]
Yoshida, T [1 ]
Fujimoto, S [1 ]
Miyahara, S [1 ]
Kang, YM [1 ]
机构
[1] Daikin Ind Ltd, MEC Lab, Tsukuba, Ibaraki 3050841, Japan
关键词
D O I
10.1088/0953-2048/12/11/379
中图分类号
O59 [应用物理学];
学科分类号
摘要
We developed a cryocooled helmet-shaped magnetoencephalographic measurement system. We determined the sources of white noise that contribute to the total white noise of this system and determined the source of cyclic noise generated by the Gifford-McMahon-Joule-Thomson cryocooler. Design modifications based on these results will reduce the white noise of our future cryocooled systems to 5 fT Hz(-1/2) for cyclic noise to 1 pT.
引用
收藏
页码:959 / 962
页数:4
相关论文
共 7 条
  • [1] COLLINGS EW, 1994, CRYOGENICS, V25, P713
  • [2] FUJIMOTO S, 1996, IN PRESS P BIOM 9681
  • [3] FUKUI N, 1996, IN PRESS P BIOM 96
  • [4] KOUCHI M, 1994, HUMAN BODY DMENSIONS, V2
  • [5] A 61-channel SQUID system for MEG measurement cooled by a GM/JT cryocooler
    Sata, K
    Fujimoto, S
    Fukui, N
    Haraguchi, E
    Kido, T
    Nishiguchi, K
    Kang, YM
    [J]. IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 1997, 7 (02) : 2526 - 2529
  • [6] SATA K, 1991, 8 INT C BIOM, P131
  • [7] SATA K, 1996, IN PRESS P BIOM 96