Bipolar Radiofrequency Ablation: Development of a New Expandable Device

被引:4
作者
Ito, Nobutake [1 ,2 ]
Pfeffer, Jochen [1 ]
Isfort, Peter [1 ,2 ]
Penzkofer, Tobias [1 ,2 ]
Kuhl, Christiane K. [1 ]
Mahnken, Andreas H. [3 ]
Schmitz-Rode, Thomas [2 ]
Bruners, Philipp [1 ]
机构
[1] Rhein Westfal TH Aachen, Dept Diagnost Radiol, Aachen, Germany
[2] Rhein Westfal TH Aachen, Helmholtz Inst Biomed Engn, Aachen, Germany
[3] Univ Marburg, Dept Radiol, Marburg, Germany
关键词
Liver; Minimally invasive therapy; Radiofrequency ablation; INTERNALLY COOLED ELECTRODES; ENHANCED HEPATIC BIPOLAR; VIVO BOVINE LIVER; EFFICIENCY; MONOPOLAR; TUMORS; COMPLICATIONS; APPLICATORS; COAGULATION; PRINCIPLES;
D O I
10.1007/s00270-013-0703-2
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
To test the performance of an expandable bipolar probe as a simple technical solution for extending the coagulation volume. On the basis of a commercially available monopolar radiofrequency (RF) probe (LeVeen), an expandable bipolar RF probe was developed by integrating a second electrode into the probe shaft. The influence of length on the second electrode, and the distance between both electrodes and generator output was investigated by performing ten ablations for each condition on a freshly excised bovine liver. Macroscopically quantified coagulation volumes, lesion shape characteristics, and procedure durations were recorded. Results of the prototype featuring the optimal configuration were compared to the original LeVeen probe and commonly used bipolar RF probe (CelonLabPower). Extension of the shaft electrode length, increasing distance between the shaft electrode and the tip electrode, and reduction of generator output resulted in increasing coagulation volumes. The coagulation volumes the prototype generated were significantly smaller and more elliptically shaped than the monopolar probe (9.4 +/- A 1.5 cm(3) vs. 12.1 +/- A 1.6 cm(3)), but were larger than the commercially available bipolar RF probe (vs. 7.3 +/- A 0.5). The procedure duration of the prototype was comparable to the monopolar probe (467 +/- A 31 s vs. 464 +/- A 17 s) and shorter than the bipolar probe (vs. 2009 +/- A 444 s). In comparison to the commercially available bipolar system, the developed prototype exhibited favorable results. The first benchmark testing of the developed bipolar prototype had promising results. However, further optimization of the applicator design and ablation protocol is needed to enlarge the achievable coagulation volume.
引用
收藏
页码:770 / 776
页数:7
相关论文
共 50 条
[41]   Radiofrequency ablation of chondroblastoma using a multi-tined expandable electrode system: initial results [J].
Tins, B ;
Cassar-Pullicino, V ;
McCall, I ;
Cool, P ;
Williams, D ;
Mangham, D .
EUROPEAN RADIOLOGY, 2006, 16 (04) :804-810
[42]   Bipolar radiofrequency ablation of tibialchondroblastomas: A report of three cases [J].
Rajalakshmi, Prathiba ;
Srivastava, Deep N. ;
Rastogi, Shishir ;
Julka, Pramod Kumar ;
Bhatnagar, Sushma ;
Gamanagatti, Shivanand .
WORLD JOURNAL OF RADIOLOGY, 2012, 4 (07) :335-340
[43]   Haemostatic partial nephrectomy using bipolar radiofrequency ablation [J].
Pareek, G ;
Wilkinson, ER ;
Schutt, D ;
Will, JA ;
Warner, TF ;
Haemmerich, D ;
Mahvi, D ;
Nakada, SY .
BJU INTERNATIONAL, 2005, 96 (07) :1101-1104
[44]   Percutaneous bipolar radiofrequency ablation for spine metastatic lesions [J].
Dimitrios Filippiadis ;
Alexis Kelekis .
European Journal of Orthopaedic Surgery & Traumatology, 2021, 31 :1603-1610
[45]   Bipolar radiofrequency ablation of tibialchondroblastomas: A report of three cases [J].
Prathiba Rajalakshmi ;
Deep N Srivastava ;
Shivanand Gamanagatti ;
Pramod Kumar Julka ;
Sushma Bhatnagar ;
Shishir Rastogi .
World Journal of Radiology, 2012, (07) :335-340
[46]   Infusion of Hypertonic Saline into the Lung Parenchyma during Radiofrequency Ablation of the Lungs with Multitined Expandable Electrodes: Results Using a Porcine Model [J].
Iishi, Tatsuhiko ;
Hiraki, Takao ;
Mimura, Hidefumi ;
Gobara, Hideo ;
Kurose, Taichi ;
Fujiwara, Hiroyasu ;
Sakurai, Jun ;
Yanai, Hiroyuki ;
Yoshino, Tadashi ;
Kanazawa, Susumu .
ACTA MEDICA OKAYAMA, 2009, 63 (03) :137-144
[47]   Efficiency Analysis of Bipolar and Multipolar Radiofrequency Ablation in an In Vivo Porcine Kidney Model Using Three-Dimensional Reconstruction of Histologic Section Series [J].
Neuhaus, Jochen ;
Blachut, Lisa ;
Rabenalt, Robert ;
Stein, Thomas ;
Koenig, Fritjoff ;
Wehner, Markus ;
Liatsikos, Evangelos ;
Stolzenburg, Jens-Uwe .
JOURNAL OF ENDOUROLOGY, 2011, 25 (05) :859-867
[48]   Liver radiofrequency ablation compromises the biological gut barrier [J].
Ypsilantis, P. ;
Lambropoulou, M. ;
Kourkoutas, I. ;
Pechlivanis, A. ;
Simopoulos, C. .
HUMAN & EXPERIMENTAL TOXICOLOGY, 2014, 33 (01) :64-73
[49]   Internally cooled bipolar radiofrequency ablation:: Is a lower power output more effective? [J].
Clasen, S. ;
Geng, A. ;
Herberts, T. ;
Boss, A. ;
Schmidt, D. ;
Schraml, C. ;
Fritz, J. ;
Kroeber, S. M. ;
Claussen, C. D. ;
Pereira, P. L. .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2007, 179 (03) :282-288
[50]   Radiofrequency ablation of large size liver tumours using novel plan-parallel expandable bipolar electrodes: Initial clinical experience [J].
Meijerink, Martijn R. ;
van den Tol, Petrousjka ;
van Tilborg, Aukje A. J. M. ;
van Waesberghe, Jan Hein T. M. ;
Meijer, Sybren ;
van Kuijk, Cornelis .
EUROPEAN JOURNAL OF RADIOLOGY, 2011, 77 (01) :167-171