Bipolar Radiofrequency Ablation: Development of a New Expandable Device

被引:4
作者
Ito, Nobutake [1 ,2 ]
Pfeffer, Jochen [1 ]
Isfort, Peter [1 ,2 ]
Penzkofer, Tobias [1 ,2 ]
Kuhl, Christiane K. [1 ]
Mahnken, Andreas H. [3 ]
Schmitz-Rode, Thomas [2 ]
Bruners, Philipp [1 ]
机构
[1] Rhein Westfal TH Aachen, Dept Diagnost Radiol, Aachen, Germany
[2] Rhein Westfal TH Aachen, Helmholtz Inst Biomed Engn, Aachen, Germany
[3] Univ Marburg, Dept Radiol, Marburg, Germany
关键词
Liver; Minimally invasive therapy; Radiofrequency ablation; INTERNALLY COOLED ELECTRODES; ENHANCED HEPATIC BIPOLAR; VIVO BOVINE LIVER; EFFICIENCY; MONOPOLAR; TUMORS; COMPLICATIONS; APPLICATORS; COAGULATION; PRINCIPLES;
D O I
10.1007/s00270-013-0703-2
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
To test the performance of an expandable bipolar probe as a simple technical solution for extending the coagulation volume. On the basis of a commercially available monopolar radiofrequency (RF) probe (LeVeen), an expandable bipolar RF probe was developed by integrating a second electrode into the probe shaft. The influence of length on the second electrode, and the distance between both electrodes and generator output was investigated by performing ten ablations for each condition on a freshly excised bovine liver. Macroscopically quantified coagulation volumes, lesion shape characteristics, and procedure durations were recorded. Results of the prototype featuring the optimal configuration were compared to the original LeVeen probe and commonly used bipolar RF probe (CelonLabPower). Extension of the shaft electrode length, increasing distance between the shaft electrode and the tip electrode, and reduction of generator output resulted in increasing coagulation volumes. The coagulation volumes the prototype generated were significantly smaller and more elliptically shaped than the monopolar probe (9.4 +/- A 1.5 cm(3) vs. 12.1 +/- A 1.6 cm(3)), but were larger than the commercially available bipolar RF probe (vs. 7.3 +/- A 0.5). The procedure duration of the prototype was comparable to the monopolar probe (467 +/- A 31 s vs. 464 +/- A 17 s) and shorter than the bipolar probe (vs. 2009 +/- A 444 s). In comparison to the commercially available bipolar system, the developed prototype exhibited favorable results. The first benchmark testing of the developed bipolar prototype had promising results. However, further optimization of the applicator design and ablation protocol is needed to enlarge the achievable coagulation volume.
引用
收藏
页码:770 / 776
页数:7
相关论文
共 50 条
[31]   A trucut biopsy needle for bipolar radiofrequency ablation of needle tract: a proof-of-concept experiment [J].
Bruners, Philipp ;
Penzkofer, Tobias ;
Isfort, Peter ;
Pfeffer, Jochen ;
Schmitz-Rode, Thomas ;
Guenther, Rolf W. ;
Mahnken, Andreas H. .
EUROPEAN RADIOLOGY, 2010, 20 (08) :2000-2004
[32]   Clinical evaluation of a novel bipolar radiofrequency ablation system for renal masses [J].
Okhunov, Zhamshid ;
Roy, Ornob ;
Duty, Brian ;
Waingankar, Nikhil ;
Herati, Amin ;
Morgenstern, Nora ;
Sheikh-Fayyaz, Silvat ;
Kavoussi, Louis R. .
BJU INTERNATIONAL, 2012, 110 (05) :688-691
[33]   “Edgeboost”: A Novel Technique to Extend the Ablation Zone Lateral to a Two-Probe Bipolar Radiofrequency Device [J].
Ya Ruth Huo ;
Krishna Pillai ;
Javed Akhter ;
David L. Morris .
CardioVascular and Interventional Radiology, 2016, 39 :97-105
[34]   Evaluation of a bipolar-cooled radiofrequency device for ablation of bone metastases: preclinical assessment in porcine vertebrae [J].
Pezeshki, Padina S. ;
Woo, Jason ;
Akens, Margarete K. ;
Davies, John E. ;
Gofeld, Michael ;
Whyne, Cari M. ;
Yee, Albert J. M. .
SPINE JOURNAL, 2014, 14 (02) :361-370
[35]   Radiofrequency-Assisted Hemostasis in a Trauma Model: A New Indication for a Bipolar Device [J].
Zacharoulis, Dimitris ;
Lazoura, Olga ;
Sioka, Eleni ;
Tzovaras, George ;
Rountas, Christos ;
Spiropoulos, Spiridon ;
Zahari, Eleni ;
Chatzitheofilou, Costas .
JOURNAL OF LAPAROENDOSCOPIC & ADVANCED SURGICAL TECHNIQUES, 2010, 20 (05) :421-426
[36]   Thermal and thermal damage responses during switching bipolar radiofrequency ablation employing bipolar needles: A computational study on the effects of different electrode configuration, input voltage and ablation duration [J].
Cheong, Jason K. K. ;
Ooi, Ean H. ;
Ooi, Ean T. .
INTERNATIONAL JOURNAL FOR NUMERICAL METHODS IN BIOMEDICAL ENGINEERING, 2020, 36 (09)
[37]   CT-guided Bipolar and Multipolar Radiofrequency Ablation (RF Ablation) of Renal Cell Carcinoma: Specific Technical Aspects and Clinical Results [J].
Sommer, C. M. ;
Lemm, G. ;
Hohenstein, E. ;
Bellemann, N. ;
Stampfl, U. ;
Goezen, A. S. ;
Rassweiler, J. ;
Kauczor, H. U. ;
Radeleff, B. A. ;
Pereira, P. L. .
CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2013, 36 (03) :731-737
[38]   Comparison of expandable electrodes in percutaneous radiofrequency ablation of renal cell carcinoma [J].
Gulesserian, Talin ;
Mahnken, Andreas H. ;
Schernthaner, Ruediger ;
Memarsadeghi, Mazda ;
Weber, Michael ;
Tacke, A. ;
Kettenbach, Joachim .
EUROPEAN JOURNAL OF RADIOLOGY, 2006, 59 (02) :133-139
[39]   Influence of Surgical Staples on Radiofrequency Ablation Using Multitined Expandable Electrodes [J].
Yusuke Sakuhara ;
Tadashi Shimizu ;
Daisuke Abo ;
Yu Hasegawa ;
Fumi Kato ;
Yoshihisa Kodama ;
Hiroki Shirato .
CardioVascular and Interventional Radiology, 2007, 30 :1201-1205
[40]   Radiofrequency ablation of chondroblastoma using a multi-tined expandable electrode system: initial results [J].
Bernhard Tins ;
Victor Cassar-Pullicino ;
Iain McCall ;
Paul Cool ;
David Williams ;
David Mangham .
European Radiology, 2006, 16 :804-810