Breath-holding during the Calibration Scan Improves the Reproducibility of Parallel Transmission at 7T for Human Brain

被引:4
作者
Harada, Taisuke [1 ,2 ]
Kudo, Kohsuke [2 ]
Uwano, Ikuko [3 ]
Yamashita, Fumio [3 ]
Kameda, Hiroyuki [1 ,3 ]
Matsuda, Tsuyoshi [4 ]
Sasaki, Makoto [3 ]
Shirato, Hiroki [1 ]
机构
[1] Hokkaido Univ, Grad Sch Med, Dept Radiat Med, Sapporo, Hokkaido, Japan
[2] Hokkaido Univ Hosp, Dept Diagnost & Intervent Radiol, Kita Ku, N14 W5, Sapporo, Hokkaido 0608648, Japan
[3] Iwate Med Univ, Div Ultrahigh Field MRI, Morioka, Iwate, Japan
[4] GE Healthcare, MR Applicat & Workflow, Tokyo, Japan
基金
日本学术振兴会;
关键词
7T magnetic resonance imaging; parallel transmission; breath-holding; Bloch-Siegert method; PULSE DESIGN; INHOMOGENEITY; TESLA; FMRI; EXCITATION; ARRAYS; MRI;
D O I
10.2463/mrms.mp.2015-0137
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The B0 and B1+ maps required for calculation of the radiofrequency (RF) pulse of parallel transmission (pTx) are obtained in calibration scans; however, they may be affected by respiratory motion. We aimed to compare the reproducibility of B0 and B1+ maps and gradient echo (GRE) images of the brain scanned with pTx at 7T between free-breathing (FB) and breath-holding (BH) conditions during the calibration scan. Methods: Nine healthy volunteers were scanned by 7T MRI using a two-channel quadrature head coil. In the pTx calibration scans performed with FB and BH, the B0 map was obtained from two different TE images and the B1+ map was calculated by the Bloch-Siegert method. A GRE image (gradient-recalled acquisition in steady state) was also obtained with RF shimming and RF design of pTx with spoke method, as well as quadrature transmission (qTx). All the scans were repeated over five sessions. The reproducibility of the B0 and B1+ maps and GRE image was evaluated with region-of-interest measurements using inter-session standard deviation (SD) and coefficient of variation (CV) values. Intensity homogeneity of GRE images was also assessed with in-plane CV. Results: Inter-session SDs of B0 and B1+ maps were significantly smaller in BH (P < 0.01). Inter-session CVs of GRE images were significantly smaller in qTx than BH and FB (P < 0.01, both); however, the CVs of BH were significantly smaller (P < 0.01). In-plane CVs of FB and BH with RF shimming were not significantly different with qTx; however, CVs of FB and BH with RF design were significantly smaller than those of qTx (P < 0.05 and P < 0.01, respectively). Conclusion: BH could improve the reproducibility of B0 and B1+ maps in pTx calibration scans and GRE images. These results might facilitate the development of pTx in human brain at 7T.
引用
收藏
页码:23 / 31
页数:9
相关论文
共 26 条
[1]   Transmit and receive transmission line arrays for 7 tesla parallel imaging [J].
Adriany, G ;
Van de Moortele, PF ;
Wiesinger, F ;
Moeller, S ;
Strupp, JP ;
Andersen, P ;
Snyder, C ;
Zhang, XL ;
Chen, W ;
Pruessmann, KP ;
Boesiger, P ;
Vaughan, T ;
Ugurbil, K .
MAGNETIC RESONANCE IN MEDICINE, 2005, 53 (02) :434-445
[2]   Combination of optimized transmit arrays and some receive array reconstruction methods can yield homogeneous images at very high frequencies [J].
Collins, CM ;
Liu, W ;
Swift, BJ ;
Smith, MB .
MAGNETIC RESONANCE IN MEDICINE, 2005, 54 (06) :1327-1332
[3]   Saturated double-angle method for rapid B1 plus mapping [J].
Cunningham, Charles H. ;
Pauly, John M. ;
Nayak, Krishna S. .
MAGNETIC RESONANCE IN MEDICINE, 2006, 55 (06) :1326-1333
[4]   Slice-by-slice B1+shimming at 7 T [J].
Curtis, Andrew T. ;
Gilbert, Kyle M. ;
Klassen, L. Martyn ;
Gati, Joseph S. ;
Menon, Ravi S. .
MAGNETIC RESONANCE IN MEDICINE, 2012, 68 (04) :1109-1116
[5]   3-POINT DIXON TECHNIQUE FOR TRUE WATER FAT DECOMPOSITION WITH BO INHOMOGENEITY CORRECTION [J].
GLOVER, GH ;
SCHNEIDER, E .
MAGNETIC RESONANCE IN MEDICINE, 1991, 18 (02) :371-383
[6]   Small-tip-angle spokes pulse design using interleaved greedy and local optimization methods [J].
Grissom, William A. ;
Khalighi, Mohammad-Mehdi ;
Sacolick, Laura I. ;
Rutt, Brian K. ;
Vogel, Mika W. .
MAGNETIC RESONANCE IN MEDICINE, 2012, 68 (05) :1553-1562
[7]   RF Shimming for Spectroscopic Localization in the Human Brain at 7 T [J].
Hetherington, Hoby P. ;
Avdievich, Nikolai I. ;
Kuznetsov, Andrey M. ;
Pan, Jullie W. .
MAGNETIC RESONANCE IN MEDICINE, 2010, 63 (01) :9-19
[8]   MAGNETIC-FIELD MAPPING [J].
HORNAK, JP ;
SZUMOWSKI, J ;
BRYANT, RG .
MAGNETIC RESONANCE IN MEDICINE, 1988, 6 (02) :158-163
[9]   Respiration-induced B0 field fluctuation compensation in balanced SSFP:: Real-time approach for transition-band SSFP fMRI [J].
Lee, J ;
Santos, JM ;
Conolly, SM ;
Miller, KL ;
Hargreaves, BA ;
Pauly, JM .
MAGNETIC RESONANCE IN MEDICINE, 2006, 55 (05) :1197-1201
[10]   Spatially resolved extended phase graphs: Modeling and design of multipulse sequences with parallel transmission [J].
Malik, Shaihan J. ;
Padormo, Francesco ;
Price, Anthony N. ;
Hajnal, Joseph V. .
MAGNETIC RESONANCE IN MEDICINE, 2012, 68 (05) :1481-1494