MRI monitoring of heating produced by ultrasound absorption in the skull: In vivo study in pigs

被引:53
作者
McDannold, N
King, RL
Hynynen, K
机构
[1] Brigham & Womens Hosp, Dept Radiol, Boston, MA 02115 USA
[2] Harvard Univ, Sch Med, Dept Radiol, Boston, MA 02115 USA
关键词
D O I
10.1002/mrm.20043
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The purpose of this study was to test the utility of MR thermometry for monitoring the temperature rise on the brain surface and in the scalp induced by skull heating during ultrasound exposures. Eleven locations in three pigs were targeted with unfocused ultrasound exposures (frequency = 690 kHz; acoustic power = 8.2-16.5 W; duration = 20 s). MR thermometry (a chemical shift technique) showed an average temperature rise in vivo of 2.8degreesC +/- 0.6degreesC and 4.4degreesC +/- 1.4degreesC on the brain surface and scalp, respectively, at an acoustic power level of 10 W. The temperature rise on the scalp agreed with that measured with a thermocouple probe inserted adjacent to the skull (average temperature rise = 4.6degreesC +/- 1.0degreesC). Characterization of the transducer showed that the average acoustic intensity was 1.3 W/cm(2) at an acoustic power of 10 W. The ability to monitor the temperature rise next to the skull with MRI-based thermometry, as shown here, will allow for safety monitoring during clinical trials of transcranial focused ultrasound. (C) 2004 Wiley-Liss, Inc.
引用
收藏
页码:1061 / 1065
页数:5
相关论文
共 20 条
[1]   Experimental demonstration of noninvasive transskull adaptive focusing based on prior computed tomography scans [J].
Aubry, JF ;
Tanter, M ;
Pernot, M ;
Thomas, JL ;
Fink, M .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2003, 113 (01) :84-93
[2]   Thermal dosimetry of a focused ultrasound beam in vivo by magnetic resonance imaging [J].
Chung, AH ;
Jolesz, FA ;
Hynynen, K .
MEDICAL PHYSICS, 1999, 26 (09) :2017-2026
[3]   Correlation of ultrasound phase with physical skull properties [J].
Clement, GT ;
Hynynen, K .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2002, 28 (05) :617-624
[4]   A non-invasive method for focusing ultrasound through the human skull [J].
Clement, GT ;
Hynynen, K .
PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (08) :1219-1236
[5]   Micro-receiver guided transcranial beam steering [J].
Clement, GT ;
Hynynen, K .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2002, 49 (04) :447-453
[6]   A hemisphere array for non-invasive ultrasound brain therapy and surgery [J].
Clement, GT ;
Sun, J ;
Giesecke, T ;
Hynynen, K .
PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (12) :3707-3719
[7]  
FRY F J, 1977, Ultrasound in Medicine and Biology, V3, P179, DOI 10.1016/0301-5629(77)90069-2
[8]  
FRY W J, 1958, Adv Biol Med Phys, V6, P281
[9]   TEMPERATURE ELEVATION AT MUSCLE BONE INTERFACE DURING SCANNED, FOCUSED ULTRASOUND HYPERTHERMIA [J].
HYNYNEN, K ;
DEYOUNG, D .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 1988, 4 (03) :267-279
[10]   Trans-skull ultrasound therapy: The feasibility of using image-derived skull thickness information to correct the phase distortion [J].
Hynynen, K ;
Sun, J .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1999, 46 (03) :752-755