Do multifilament alloplastic meshes increase the infection rate?: Analysis of the polymeric surface, the bacteria adherence, and the in vivo consequences in a rat model

被引:114
作者
Klinge, U
Junge, K
Spellerberg, B
Piroth, C
Klosterhalfen, B
Schumpelick, V
机构
[1] Rhein Westfal TH Aachen, Dept Surg, D-52057 Aachen, Germany
[2] Rhein Westfal TH Aachen, Inst Med Microbiol, D-52057 Aachen, Germany
[3] Rhein Westfal TH Aachen, Inst Pathol, D-52057 Aachen, Germany
来源
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH | 2002年 / 63卷 / 06期
关键词
surgical mesh; bacterial infection; adherence; filament surface; monofilament; multifilament;
D O I
10.1002/jbm.10449
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Within the last decade hernia surgery has changed from suture repair to mesh repair. Biomaterials, and multifilaments in particular, are thought to increase the risk of infection. Therefore, the aim of this study was to study the influence of the presence of either a monofilament or a multifilament mesh material on the bacterial infection risk. The filament surface of a monofilament and a multifilament mesh were calculated on the basis of a theoretical model. The adherence of Staphylococcus aureus was measured in vitro by fluorescence analysis. Additionally, the two mesh materials (8-mm platelets) were implanted subcutaneously in Sprague-Dawley rats with daily surveillance for clinical signs of infection. After 7 days the meshes were explanted for histological and microbiological analysis. Calculations of the mesh surface area revealed a higher level for the multifilament mesh. The extent of adherent bacteria corresponded to the estimated filament surface in vitro. In vivo, the implantation of meshes in the presence of 5 x 10(6) S. aureus did not show an increased infection rate in rats with either monofilament or multifilament material, compared to the control groups (mesh implantation without S. aureus contamination). However, after 7 days bacteria were still detectable in the majority of the implantation sites, and a clinically inapparent intensification of local inflammation and fibrosis was induced. The increased surface area of a multifilament meshes promotes the persistence of bacteria in the implant bed, though this alone is not sufficient to create a clinically apparent infection. This might explain the development of mesh-related infections after a delay of several months or even years. In vivo, the adherence of bacteria to the implant material depends on the surface area, which favors the use of monofilament materials. (C) 2002 Wiley Periodicals, Inc.
引用
收藏
页码:765 / 771
页数:7
相关论文
共 33 条
[1]  
Amid P., 1997, HERNIA, V1, P5
[2]   Mesh infections after laparoscopic inguinal hernia repair [J].
Avtan, L ;
Avci, C ;
Bulut, T ;
Fourtanier, G .
SURGICAL LAPAROSCOPY & ENDOSCOPY, 1997, 7 (03) :192-195
[3]  
BLOMSTEDT B, 1977, ACTA CHIR SCAND, V143, P259
[4]   COMPARISON OF PROSTHETIC MATERIALS FOR ABDOMINAL-WALL RECONSTRUCTION IN THE PRESENCE OF CONTAMINATION AND INFECTION [J].
BROWN, GL ;
RICHARDSON, JD ;
MALANGONI, MA ;
TOBIN, GR ;
ACKERMAN, D ;
POLK, HC .
ANNALS OF SURGERY, 1985, 201 (06) :705-711
[5]   THE CHOICE OF A SUTURE TO CLOSE ABDOMINAL INCISIONS [J].
BUCKNALL, TE ;
TEARE, L ;
ELLIS, H .
EUROPEAN SURGICAL RESEARCH, 1983, 15 (02) :59-66
[6]   ABDOMINAL WOUND CLOSURE - CHOICE OF SUTURE [J].
BUCKNALL, TE .
JOURNAL OF THE ROYAL SOCIETY OF MEDICINE, 1981, 74 (08) :580-585
[7]  
Carbajo M A, 2000, JSLS, V4, P141
[8]   EFFECTS OF PHYSICAL CONFIGURATION AND CHEMICAL-STRUCTURE OF SUTURE MATERIALS ON BACTERIAL ADHESION - A POSSIBLE LINK TO WOUND-INFECTION [J].
CHU, CC ;
WILLIAMS, DF .
AMERICAN JOURNAL OF SURGERY, 1984, 147 (02) :197-204
[9]  
CODA A, 1998, HERNIA, V2, P163
[10]   Pathophysiology, prevention, and management of prosthetic infections in hernia surgery [J].
Deysine, M .
SURGICAL CLINICS OF NORTH AMERICA, 1998, 78 (06) :1105-+