Comparison of Carotid Stents: An In-Vitro Experiment Focusing on Stent Design

被引:68
作者
Mueller-Huelsbeck, Stefan [1 ]
Schaefer, Philipp J. [2 ]
Charalambous, Nikolas
Schaffner, Silvio R. [3 ]
Heller, Martin [2 ]
Jahnke, Thomas [2 ]
机构
[1] Acad Hosp Flensburg, Dept Diagnost & Intervent Radiol Neuroradiol, D-24939 Flensburg, Germany
[2] Univ Hosp Schleswig Holstein, Dept Radiol, Kiel, Germany
[3] Invatec Technol Ctr, Frauenfeld, Switzerland
关键词
carotid stent; nitinol; experimental evaluation; comparison; experimental study; radial force; flexibility; torsion; bending; scaffolding measurements; ANGIOPLASTY;
D O I
10.1583/08-2539.1
中图分类号
R61 [外科手术学];
学科分类号
摘要
Purpose: To examine and compare different carotid stent designs with regard to flexibility, adaptability (adjustability), conformability (compliance) to the vessel, and scaffolding to reduce plaque prolapse and embolization. Methods: Six stents of different design were compared (Precise, Acculink, Protege, Xact, Wallstent, and Cristallo Ideale). Optical microscopy was used to determine exact dimensions and scaffolding of each stent. Radial force was tested using a parallel plate setup, and flexibility (torsion and bending) was measured in water at body temperature. Particle penetration simulation was performed using plastic spheres from 1.5- to 6.0-mm outer diameter. Results: Stent dimensions met the manufacturers' data; none of the products showed any failure during the test program. Cell sizes in the middle part of the stents ranged from 1.36 mm(2) (Wallstent) to 15. 10 mm(2) (Acculink). Bending forces at 20 degrees/30 degrees ranged from 0.063 N / 0.074 N (Cristallo Ideale) to 0.890 N / 0.616 N (Xact); forces to achieve torsion at 10 degrees/15 degrees ranged from 0.032 N / 0.043 N (Acculink) to 0.905 N / 1.071 N (Xact). According to the parallel plate method, mean lowest force was measured for Xact (0.765 N), while the Wallstent had the highest force (2.136 N). Mean radial force measurements were lowest for Cristallo Ideale (9.06 N at mid part) and highest for Protege (24.09 N). The Cristallo Ideale stent at mid part resisted penetration by all but the smallest plastic spheres (1.5-mm spheres penetrated only at 0.65 N); the Precise and Protege stent had the highest variation in sphere penetration (1.5- to 4.0-mm spheres). Only the Acculink let 6-mm spheres penetrate. Conclusion: Despite comparable stent sizes, these carotid stents showed differences in behavior due to stent design. The open-cell design displayed the greatest flexibility and adaptability to the vessel but easily allowed particle penetration due to the open structure. Closed-cell designs had low flexibility and thus low adaptability to the vessel but high resistance to particle penetration due to the closed-cell design and high scaffolding. The hybrid stent design (Cristallo Ideale) was able to combine both the flexibility of an open-cell structure and the resistance to particle penetration of closed-cell structures. J Endovasc Ther. 2009;16:168-177
引用
收藏
页码:168 / 177
页数:10
相关论文
共 8 条
  • [1] Does free cell area influence the outcome in carotid artery stenting?
    Bosiers, M.
    de Donato, G.
    Deloose, K.
    Verbist, J.
    Peeters, P.
    Castriota, F.
    Cremonesi, A.
    Setacci, C.
    [J]. EUROPEAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY, 2007, 33 (02) : 135 - 141
  • [2] Multicenter experience with a new "hybrid" carotid stent
    Cremonesi, Alberto
    Rubino, Paolo
    Grattoni, Chiara
    Scheinert, Dierk
    Castriota, Fausto
    Biamino, Giancarlo
    [J]. JOURNAL OF ENDOVASCULAR THERAPY, 2008, 15 (02) : 186 - 192
  • [3] Physical properties of endovascular stents: An experimental comparison
    Duda, SH
    Wiskirchen, J
    Tepe, G
    Bitzer, M
    Kaulich, TW
    Stoeckel, D
    Claussen, CD
    [J]. JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2000, 11 (05) : 645 - 654
  • [4] Incidence, mechanism, predictors, and long-term prognosis of late stent malapposition after bare-metal stent implantation
    Hong, MK
    Mintz, GS
    Lee, CW
    Kim, YH
    Lee, SW
    Song, JM
    Han, KH
    Kang, DH
    Song, JK
    Kim, JJ
    Park, SW
    Park, SJ
    [J]. CIRCULATION, 2004, 109 (07) : 881 - 886
  • [5] Müller-Hülsbeck S, 2003, ROFO-FORTSCHR RONTG, V175, P1634
  • [6] Stent-protected carotid angioplasty using a membrane stent:: A comparative cadaver study
    Muller-Hulsbeck, Stefan
    Guhne, Albrecht
    Tsokos, Michael
    Husler, Erhard J.
    Schaffner, Silvio R.
    Paulsen, Friedrich
    Hedderich, Jurgen
    Heller, Martin
    Jahnke, Thomas
    [J]. CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2006, 29 (04) : 630 - 636
  • [7] RIEDEL CH, 2006, ROFOFORTSCHR RONT S1, V178, pS229
  • [8] Updated review of the global carotid artery stent registry
    Wholey, MH
    Al-Mubarek, N
    Wholey, MH
    [J]. CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2003, 60 (02) : 259 - 266