Development and characterization of an all-in-one gamma probe with auto-peak detection for sentinel lymph node biopsy based on NEMA NU3-2004 standard

被引:4
作者
Radnia, Aram [1 ]
Abdollahzadeh, Hamed [1 ]
Teimourian, Behnoosh [1 ]
Farahani, Mohammad Hossein [1 ]
Akbari, Mohammad Esmaeil [3 ]
Zaidi, Habib [4 ,5 ,6 ,7 ]
Ay, Mohammad Reza [1 ,2 ]
机构
[1] Univ Tehran Med Sci, Res Ctr Mol & Cellular Imaging, Tehran, Iran
[2] Univ Tehran Med Sci, Dept Med Phys & Biomed Engn, Tehran, Iran
[3] Shahid Beheshti Univ Med Sci, Canc Res Ctr, Tehran, Iran
[4] Geneva Univ Hosp, Dept Med Imaging, Div Nucl Med & Mol Imaging, CH-1211 Geneva 4, Switzerland
[5] Univ Geneva, Geneva Univ Neuroctr, CH-1205 Geneva, Switzerland
[6] Univ Groningen, Univ Med Ctr Groningen, Dept Nucl Med & Mol Imaging, NL-9700 RB Groningen, Netherlands
[7] Univ Southern Denmark, Dept Nucl Med, DK-500 Odense, Denmark
关键词
Intra-operative gamma probe; All-in-one; Sentinel lymph node; Performance evaluation; NEMA NU3;
D O I
10.1007/s12149-021-01581-z
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background A gamma probe is a handheld device used for intraoperative interventions following interstitial injection of a radiotracer to locate regional lymph nodes through the external detection of radiation. This work reports on the design and performance evaluation of a novel fully integrated gamma probe (GammaPen), recently developed by our group. Materials and methods GammaPen is an all-in-one pocket gamma probe with low weight and adequate dimensions, consisting of a detector, a control unit and output all together. The detector module consists of a cylindrical Thallium-activated Cesium Iodide [CsI (Tl)] crystal optically coupled to a Silicon photomultiplier (SiPM), shielded using Tungsten housing on side and back faces. The electronics of the probe consists of two small boards to handle signal processing and analog peak detection tasks. A number of parameters, including probe sensitivity in air/water, spatial resolution in air/water, angular resolution in air/water, and side and back shielding effectiveness, were measured to evaluate the performance of the probe based on NEMA NU3-2004 standards. Results The sensitivity of the probe in air at distances of 10, 30, and 50 mm is 18784, 3500, and 1575 cps/MBq. The sensitivity in scattering medium was also measured at distances of 10, 30, and 50 mm as 17,680, 3050, and 1104 cps/MBq. The spatial and angular resolutions in scattering medium were 47 mm and 87 degree at 30 mm distance from the probe, while they were 40 mm and 77 degree in air. The detector shielding effectiveness and leakage sensitivity are 99.91% and 0.09%, respectively. Conclusion The performance characterization showed that GammaPen can be used effectively for sentinel lymph node localization. The probe was successfully used in several surgical interventions by an experienced surgeon confirming its suitability in a clinical setting.
引用
收藏
页码:438 / 446
页数:9
相关论文
共 33 条
[1]   Lymphoscintigraphy, the sentinel node concept, and the intraoperative gamma probe in melanoma, breast cancer, and other potential cancers [J].
Alazraki, NP ;
Eshima, D ;
Eshima, LA ;
Herda, SC ;
Murray, DR ;
Vansant, JP ;
Taylor, AT .
SEMINARS IN NUCLEAR MEDICINE, 1997, 27 (01) :55-67
[2]   GAMMA-PROBE GUIDED LOCALIZATION OF LYMPH-NODES [J].
ALEX, JC ;
KRAG, DN .
SURGICAL ONCOLOGY-OXFORD, 1993, 2 (03) :137-143
[3]  
[Anonymous], PHOTOMULTIPLIER SENS
[4]   Sentinel lymph node biopsy in breast cancer: Guidelines and pitfalls of lymphoscintigraphy and gamma probe detection [J].
Borgstein, P ;
Pijpers, R ;
Comans, EF ;
van Diest, PJ ;
Boom, RP ;
Meijer, S .
JOURNAL OF THE AMERICAN COLLEGE OF SURGEONS, 1998, 186 (03) :275-283
[5]  
Classe JM, 2005, J NUCL MED, V46, P395
[6]  
CROWLEY NJ, 1990, CANCER-AM CANCER SOC, V66, P2522, DOI 10.1002/1097-0142(19901215)66:12<2522::AID-CNCR2820661213>3.0.CO
[7]  
2-Z
[8]   Multi-institutional melanoma lymphatic mapping experience: The prognostic value of sentinel lymph node status in 612 stage I or II melanoma patients [J].
Gershenwald, JE ;
Thompson, W ;
Mansfield, PF ;
Lee, JE ;
Colome, MI ;
Tseng, CH ;
Lee, JJ ;
Balch, CM ;
Reintgen, DS ;
Ross, MI .
JOURNAL OF CLINICAL ONCOLOGY, 1999, 17 (03) :976-983
[9]  
Harris, 1956, CSI TL CRYSTAL SURG, P14
[10]   Detection of sentinel lymph nodes in minimally invasive surgery using indocyanine green and near-infrared fluorescence imaging for uterine and cervical malignancies [J].
Jewell, Elizabeth L. ;
Huang, Juan Juan ;
Abu-Rustum, Nadeem R. ;
Gardner, Ginger J. ;
Brown, Carol L. ;
Sonoda, Yukio ;
Barakat, Richard R. ;
Levine, Douglas A. ;
Leitao, Mario M., Jr. .
GYNECOLOGIC ONCOLOGY, 2014, 133 (02) :274-277