A mode converter to generate a Gaussian-like mode for injection into the VENUS electron cyclotron resonance ion source

被引:25
作者
Lyneis, C. [1 ]
Benitez, J. [1 ]
Hodgkinson, A. [1 ]
Plaum, B. [2 ]
Strohmeier, M. [1 ]
Thuillier, T. [3 ]
Todd, D. [1 ]
机构
[1] Univ Calif Berkeley, Lawrence Berkeley Natl Lab, Div Nucl Sci, Berkeley, CA 94720 USA
[2] Inst Grenzflachenverfahrenstech & Plasmatechnol I, Stuttgart, Germany
[3] Univ Grenoble 1, CNRS, IN2P3, Lab Phys Subat & Cosmol,Inst Polytech Grenoble, F-38026 Grenoble, France
关键词
28; GHZ; WAVE-GUIDES; CONVERSION; ECRIS;
D O I
10.1063/1.4832064
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A number of superconducting electron cyclotron resonance (ECR) ion sources use gyrotrons at either 24 or 28 GHz for ECR heating. In these systems, the microwave power is launched into the plasma using the TE01 circular waveguide mode. This is fundamentally different and may be less efficient than the typical rectangular, linearly polarized TE10 mode used for launching waves at lower frequencies. To improve the 28 GHz microwave coupling in VENUS, a TE01-HE11 mode conversion system has been built to test launching HE11 microwave power into the plasma chamber. The HE11 mode is a quasi-Gaussian, linearly polarized mode, which should couple strongly to the plasma electrons. The mode conversion is done in two steps. First, a 0.66 m long "snake" converts the TE01 mode to the TE11 mode. Second, a corrugated circular waveguide excites the HE11 mode, which is launched directly into the plasma chamber. The design concept draws on the development of similar devices used in tokamaks and stellerators. The first tests of the new coupling system are described below. (C) 2013 AIP Publishing LLC.
引用
收藏
页数:5
相关论文
共 23 条
[1]   28 GHz 10 kW gyrotron system for electron cyclotron resonance ion source [J].
Bykov, Y ;
Denisov, G ;
Eremeev, A ;
Gorbatushkov, V ;
Kurkin, V ;
Kalynova, G ;
Kholoptsev, V ;
Luchinin, A ;
Plotnikov, I .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2004, 75 (05) :1437-1439
[2]   Modeling of the stability of electron cyclotron resonance ion source plasmas [J].
Cluggish, Brian P. ;
Zhao, Liangji ;
Kim, Jin-Soo .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2011, 631 (01) :111-120
[4]   Operation of the SERSE superconducting electron cyclotron resonance ion source at 28 GHz [J].
Gammino, S ;
Ciavola, G ;
Celona, L ;
Hitz, D ;
Girard, A ;
Melin, G .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2001, 72 (11) :4090-4097
[5]  
Geller R., 1996, ELECT CYCLOTRON RESO, P211
[6]  
Geller R., 1996, ELECT CYCLOTRON RESO, P214
[7]   Multiply charged ion production with ECR ion sources: State of the art and prospects [J].
Hitz, D ;
Girard, A ;
Melin, G ;
Cormier, D ;
Mathonnet, JM ;
Lansaque, F ;
Serebrenikov, K ;
Sun, LT .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2003, 205 :168-172
[8]   Recent progress in high frequency electron cyclotron resonance ion sources [J].
Hitz, Denis .
ADVANCES IN IMAGING AND ELECTRON PHYSICS, VOL 144, 2006, 144 :1-164
[9]   Next generation ECR ion sources: First results of the superconducting 28 GHz ECRIS - VENUS [J].
Leitner, D ;
Lyneis, CM ;
Abbott, SR ;
Collins, D ;
Dwinell, RD ;
Galloway, ML ;
Leitner, M ;
Todd, DS .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2005, 235 :486-493
[10]   MODE CONVERSION DUE TO CURVATURE IN CORRUGATED WAVE-GUIDES [J].
LI, HF ;
THUMM, M .
INTERNATIONAL JOURNAL OF ELECTRONICS, 1991, 71 (02) :333-347