Simultaneous Tc-99m/I-123 dual-radionuclide myocardial perfusion/innervation imaging using Siemens IQ-SPECT with SMARTZOOM collimator

被引:16
作者
Du, Yong [1 ]
Bhattacharya, Manojeet [2 ]
Frey, Eric C. [1 ]
机构
[1] Johns Hopkins Univ, Dept Radiol, Baltimore, MD 21287 USA
[2] Siemens Med Solut USA Inc, Mol Imaging Sci & Technol, Hoffman Estates, IL 60192 USA
关键词
simultaneous dual-radionuclide SPECT; myocardial perfusion; myocardial innervation; CARDIAC SYMPATHETIC INNERVATION; CROSSTALK COMPENSATION; NERVOUS FUNCTION; MODEL; PERFUSION; MIBG; I-123; RECONSTRUCTION; VALIDATION; SCATTER;
D O I
10.1088/0031-9155/59/11/2813
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Simultaneous dual-radionuclide myocardial perfusion/innervation SPECT imaging can provide important information about the mismatch between scar tissue and denervated regions. The Siemens IQ-SPECT system developed for cardiac imaging uses a multifocal SMARTZOOM collimator to achieve a four-fold sensitivity for the cardiac region, compared to a typical parallel-hole low-energy high-resolution collimator, but without the data truncation that can result with conventional converging-beam collimators. The increased sensitivity allows shorter image acquisition times or reduced patient dose, making IQ-SPECT ideal for simultaneous dual-radionuclide SPECT, where reduced administrated activity is desirable in order to reduce patient radiation exposure. However, crosstalk is a major factor affecting the image quality in dual-radionuclide imaging. In this work we developed a model-based method that can estimate and compensate for the crosstalk in IQ-SPECT data. The crosstalk model takes into account interactions in the object and collimator-detector system. Scatter in the object was modeled using the effective source scatter estimation technique (ESSE), previously developed to model scatter with parallel-hole collimators. The geometric collimator-detector response was analytically modeled in the IQ-SPECT projector. The estimated crosstalk was then compensated for in an iterative reconstruction process. The new method was validated with data from both Monte Carlo simulations and physical phantom experiments. The results showed that the estimated crosstalk was in good agreement with simulated and measured results. After model-based compensation the images from simultaneous dual-radionuclide acquisitions were similar in quality to those from single-radionuclide acquisitions that did not have crosstalk contamination. The proposed model-based method can be used to improve simultaneous dual-radionuclide images acquired using IQ-SPECT. This work also demonstrates that ESSE scatter modeling can be applied to non-parallel-beam projection geometries.
引用
收藏
页码:2813 / 2828
页数:16
相关论文
共 33 条
[1]  
Agostini D, 2002, J NUCL MED, V43, p172P
[2]   Simultaneous dual-radionuclide myocardial perfusion imaging with a solid-state dedicated cardiac camera [J].
Ben-Haim, Simona ;
Kacperski, Krzysztof ;
Hain, Sharon ;
Van Gramberg, Dean ;
Hutton, Brian F. ;
Erlandsson, Kjell ;
Sharir, Tali ;
Roth, Nathaniel ;
Waddington, Wendy A. ;
Berman, Daniel S. ;
Ell, Peter J. .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2010, 37 (09) :1710-1721
[3]  
Buvat I, 1999, J NUCL MED, V40, p116P
[4]   Simultaneous 99mTc/201Tl dual-isotope SPET with Monte Carlo-based down-scatter correction [J].
de Jong, HWAM ;
Beekman, FJ ;
Viergever, MA ;
van Rijk, PP .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2002, 29 (08) :1063-1071
[5]  
DEVOUS MD, 1992, J NUCL MED, V33, P2030
[6]   Simultaneous 99mTc/201Tl/123I triple-isotope cardiac SPECT imaging [J].
Du, Y. ;
Frey, E. C. .
2007 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-11, 2007, :4213-4216
[7]   Evaluation of simultaneous 201Tl/99mTc dual-isotope cardiac SPECT imaging with model-based crosstalk compensation using canine studies [J].
Du, Y. ;
Links, J. M. ;
Becker, L. ;
DiPaula, A. F. ;
Frank, T. ;
Schuleri, K. H. ;
Lardo, A. C. ;
Frey, E. C. .
JOURNAL OF NUCLEAR CARDIOLOGY, 2014, 21 (02) :329-340
[8]   Model-based crosstalk compensation for simultaneous 99mTc/123I dual-isotope brain SPECT imaging [J].
Du, Yong ;
Tsui, Benjamin M. W. ;
Frey, Eric C. .
MEDICAL PHYSICS, 2007, 34 (09) :3530-3543
[9]   Quantitative evaluation of simultaneous reconstruction with model-based crosstalk compensation for 99mTc/123I dual-isotope simultaneous acquisition brain SPECT [J].
Du, Yong ;
Frey, Eric C. .
MEDICAL PHYSICS, 2009, 36 (06) :2021-2033
[10]   Scatter and cross-talk corrections in simultaneous Tc-99m/I-123 brain SPECT using constrained factor analysis and artificial neural networks [J].
El Fakhri, G ;
Maksud, P ;
Kijewski, MF ;
Habert, MO ;
Todd-Pokropek, A ;
Aurengo, A ;
Moore, SC .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2000, 47 (04) :1573-1580