Effects of Geometrical Uncertainties on Whole Breast Radiotherapy: A Comparison of Four Different Techniques

被引:15
作者
Nakamura, Naoki [1 ]
Takahashi, Osamu [2 ]
Karno, Minobu [3 ]
Hatanaka, Shogo [1 ]
Endo, Hartma [1 ]
Mizuno, Norifumi [1 ]
Shikama, Naoto [4 ]
Ogita, Many [1 ]
Sekiguchi, Kenji [1 ]
机构
[1] St Lukes Int Hosp, Dept Radiat Oncol, Tokyo 1048560, Japan
[2] St Lukes Int Hosp, Dept Med, Div Gen Internal Med, Tokyo 1048560, Japan
[3] St Lukes Int Hosp, Dept Radiol, Tokyo 1048560, Japan
[4] Saitama Med Univ Int, Med Ctr, Dept Radiat Oncol, Hidaka, Japan
关键词
Breast neoplasms; Intensity-modulated radiotherapy; DOSE DISTRIBUTIONS; IMRT; IRRADIATION; CANCER; MOTION;
D O I
10.4048/jbc.2014.17.2.157
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: The purpose of this study was to quantify the target coverage, homogeneity, and robustness of the dose distributions against geometrical uncertainties associated with four whole breast radiotherapy techniques. Methods: The study was based on the planning-computed tomography-datasets of 20 patients who underwent whole breast radiotherapy. A total of four treatment plans (wedge, field-in-field [FIF], hybrid intensity-modulated radiotherapy [IMRT], and full IMRT) were created for each patient. The hybrid IMRT plans comprised two opposed tangential open beams plus two IMRT beams. Setup errors were simulated by moving the beam isocenters by 5 mm in the anterior or posterior direction. Results: With the original plan, the wedge technique yielded a high volume receiving >= 107% of the prescription dose (V-107; 7.5%+/- 4.2%), whereas the other three techniques yielded excellent target coverage and homogeneity. A 5 mm anterior displacement caused a large and significant increase in the V-107 (+5.2 +/-%4.1%, p<0.01) with the FIF plan, but not with the hybrid IMRT (+0.4%+/- 1.2%, p=0.11) or full IMRT (+0.7%+/- 1.8%, p=0.10) plan. A 5-mm posterior displacement caused a large decrease in the V-95 with the hybrid IMRT (-2.5% +/- 3.7%, p<0.01) and full IMRT (-4.3%+/- 5.1%, p<0.01) plans, but not with the FIF plan (+0.1%+/- 0.7%, p=0.74). The decrease in V95 was significantly smaller with the hybrid IMRT plan than with the full IMRT plan (p<0.01). Conclusion: The FIF, hybrid IMRT, and full IMRT plans offered excellent target coverage and homogeneity. Hybrid IMRT provided better robustness against geometrical uncertainties than full IMRT, whereas FIF provided comparable robustness to that of hybrid IMRT.
引用
收藏
页码:157 / 160
页数:4
相关论文
共 11 条
[1]   Dosimetric improvements following 3D planning of tangential breast irradiation [J].
Aref, A ;
Thornton, D ;
Youssef, E ;
He, T ;
Tekyi-Mensah, S ;
Denton, L ;
Ezzell, G .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2000, 48 (05) :1569-1574
[2]   The sensitivity of dose distributions for organ motion and set-up uncertainties in prostate IMRT [J].
Bos, LJ ;
van der Geer, J ;
van Herk, M ;
Mijnheer, BJ ;
Lebesque, JV ;
Damen, EMF .
RADIOTHERAPY AND ONCOLOGY, 2005, 76 (01) :18-26
[3]   Dosimetric analysis of intact breast irradiation in off-axis planes [J].
Buchholz, TA ;
Gurgoze, E ;
Bice, WS ;
Prestidge, BR .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 39 (01) :261-267
[4]  
de la Torre Ninet, 2004, Med Dosim, V29, P109, DOI 10.1016/j.meddos.2004.03.002
[5]   Intensity modulation under geometrical uncertainty: a deconvolution approach to robust fluence [J].
Fan, Yankhua ;
Nath, Ravinder .
PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (14) :4029-4045
[6]   In clinical toxicities compared with conventional wedge-based breast radiotherapy [J].
Harsolia, Asif ;
Kestin, Larry ;
Grills, Inga ;
Wallace, Michelle ;
Jolly, Shruti ;
Jones, Cortney ;
Lala, Mnaktar ;
Martinez, Alvaro ;
Schell, Scott ;
Vicini, Frank A. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 68 (05) :1375-1380
[7]   Inter-fraction motion and dosimetric consequences during breast intensity-modulated radiotherapy (IMRT) [J].
Jain, Popja ;
Marchant, Tom ;
Green, Melanie ;
Watkins, Gillian ;
Davies, Julie ;
McCarthy, Claire ;
Loncaster, Juliette ;
Stewart, Alan ;
Magee, Brian ;
Moore, Christopher ;
Price, Pat .
RADIOTHERAPY AND ONCOLOGY, 2009, 90 (01) :93-98
[8]   Quantification of Cold Spots Caused by Geometrical Uncertainty in Field-in-field Techniques for Whole Breast Radiotherapy [J].
Nakamura, Naoki ;
Hatanaka, Shogo ;
Shikama, Naoto ;
Akahane, Keiko ;
Sekiguchi, Kenji .
JAPANESE JOURNAL OF CLINICAL ONCOLOGY, 2011, 41 (09) :1127-1131
[9]   Impact of geometrical uncertainties on 3D CRT and IMRT dose distributions for lung cancer treatment [J].
Schwarz, Marco ;
Van Der Geer, Joris ;
Van Herk, Marcel ;
Lebesque, Joos V. ;
Mijnheer, Ben J. ;
Damen, Eugene M. F. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2006, 65 (04) :1260-1269
[10]   Adoption of Intensity-Modulated Radiation Therapy for Breast Cancer in the United States [J].
Smith, Benjamin D. ;
Pan, I-Wen ;
Shih, Ya-Chen T. ;
Smith, Grace L. ;
Harris, Jay R. ;
Punglia, Rinaa ;
Pierce, Lori J. ;
Jagsi, Reshma ;
Hayman, James A. ;
Giordano, Sharon H. ;
Buchholz, Thomas A. .
JNCI-JOURNAL OF THE NATIONAL CANCER INSTITUTE, 2011, 103 (10) :798-809