A Model-Based Transit-Time Ultrasonic Gas Flowrate Measurement Method

被引:30
作者
Jiang, Yandan [1 ]
Wang, Baoliang [1 ]
Huang, Zhiyao [1 ]
Ji, Haifeng [1 ]
Li, Haiqing [1 ]
Li, Xia [2 ]
机构
[1] Zhejiang Univ, State Key Lab Ind Control Technol, Coll Control Sci & Engn, Hangzhou 310027, Peoples R China
[2] Chongqing Acad Metrol & Qual Inspect, Chongqing 401123, Peoples R China
关键词
Gas flowrate; independent component analysis with reference (ICA-R); time-of-flight (TOF); ultrasonic flowmeter; zero-phase filtering; SQUARES PARAMETER-ESTIMATION; PROPAGATION; ECHOES; FLOWMETERS; ALGORITHM; DESIGN; FILTER; FLUID;
D O I
10.1109/TIM.2016.2627247
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Based on the transit-time ultrasonic gas flowrate measurement principle, a new model-based time-of-flight (TOF) measurement method is proposed. The proposed TOF measurement method includes two parts, the signal preprocessing and the model-fitting. Zero-phase filtering and independent component analysis with reference (ICA-R) are used to realize the signal preprocessing. The least-squares algorithm is used to implement the model-fitting and hence to obtain the TOF. A new single-acoustic-path ultrasonic gas flowmeter prototype with the inner diameter of 100 mm is developed and experiments are carried out to test the performance of the prototype. Experimental results show that the proposed signal preprocessing can effectively correct the distorted/noise-polluted measurement signal. In addition, the proposed model-based TOF measurement method is effective and can implement TOF measurement with higher repeatability and low standard deviation. For the gas flowrate ranges from 5 to 50 m(3)/h, the repeatability error of gas flowrate measurement is less than 1.99% and the maximum relative error of flowrate measurement is less than 3.27%. For the gas flowrate ranges from 50 to 500 m(3)/h, the repeatability error of gas flowrate measurement is less than 0.51% and the maximum relative error of flowrate measurement is less than 1.43%.
引用
收藏
页码:879 / 887
页数:9
相关论文
共 39 条
[1]  
American Gas Association, 2003, MEAS GAS MULT ULTR, P10
[2]  
Baker R.C., 2000, FLOW MEASUREMENT HDB
[3]  
Bao JC, 2008, ADV ASTRONAUT SCI, V129, P1413
[4]   A new class of incremental gradient methods for least squares problems [J].
Bertsekas, DP .
SIAM JOURNAL ON OPTIMIZATION, 1997, 7 (04) :913-926
[5]   DETECTION OF ECHOES IN NOISY ENVIRONMENTS FOR MULTILAYER STRUCTURE CHARACTERIZATION [J].
BETTA, G ;
DAPONTE, P .
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1993, 42 (04) :834-841
[6]   High-frequency transducers and correlation method to enhance ultrasonic gas flow metering [J].
Brassier, P ;
Hosten, B ;
Vulovic, F .
FLOW MEASUREMENT AND INSTRUMENTATION, 2001, 12 (03) :201-211
[7]   An adaptive Morlet wavelet filter for time-of-flight estimation in ultrasonic damage assessment [J].
Chen, Hanxin ;
Zuo, Ming J. ;
Wang, Xiaodong ;
Hoseini, Mohammad R. .
MEASUREMENT, 2010, 43 (04) :570-585
[8]   Realization of a multipath ultrasonic gas flowmeter based on transit-time technique [J].
Chen, Qiang ;
Li, Weihua ;
Wu, Jiangtao .
ULTRASONICS, 2014, 54 (01) :285-290
[9]   Acoustic propagation in viscous fluid with uniform flow and a novel design methodology for ultrasonic flow meter [J].
Chen, Yong ;
Huang, Yiyong ;
Chen, Xiaoqian .
ULTRASONICS, 2013, 53 (02) :595-606
[10]   Model-based estimation of ultrasonic echoes part I: Analysis and algorithms [J].
Demirli, R ;
Saniie, J .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2001, 48 (03) :787-802