Comparison of GMT presto assay and Roche cobase® 4800 CT/NG assay for detection of Chlamydia trachomatis and Neisseria gonorrhoeae in dry swabs

被引:13
作者
de Waaij, Dewi J. [1 ,2 ]
Dubbink, Jan Henk [1 ,2 ]
Peters, Remco P. H. [3 ,4 ,5 ]
Ouburg, Sander [1 ]
Morre, Servaas A. [1 ,2 ]
机构
[1] Vrije Univ Amsterdam, Med Ctr, Dept Med Microbiol & Infect Control, Immunogenet Lab, Amsterdam, Netherlands
[2] Maastricht Univ, Res Sch GROW, Sch Oncol & Dev Biol, Fac Hlth Med & Life Sci,IPHG,Dept Genet & Cell Bi, NL-6200 MD Maastricht, Netherlands
[3] Anova Hlth Inst, Johannesburg, South Africa
[4] Anova Hlth Inst, Tzaneen, South Africa
[5] Maastricht Univ, Dept Med Microbiol, NL-6200 MD Maastricht, Netherlands
关键词
Chlamydia trachomatis; Neisseria gonorrhoeae; Real-time PCR; GMT presto assay; Roche cobas 4800 CT/NG test; URINE SPECIMENS; PCR; PERFORMANCE; INFECTIONS; AMPLICOR; STANDARD; CT;
D O I
10.1016/j.mimet.2015.08.020
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Urogenital Chlamydia trachomatis (CT) and Neisseria gonorrhoeae (NG) are the most prevalent bacterial STIs worldwide. Molecular tests are the standard for the detection of CT and NG, as these are difficult to culture. The recently introduced CE-IVD marked GMT Presto assay promises to be a valuable addition in CT and NC diagnostics. The advantage of the Presto assay is that it works on many PCR systems and the DNA can be isolated by any system. We compared the Presto assay to the widely used Roche cobas (R) 4800 CT/NC test for the detection of CT and NC in 612 vaginal and rectal dry collected swabs. Discrepant samples were tested by the TIB MOLBIOL Lightmix Kit 480 HT CT/NC assay. The alloyed gold standard was defined as two concurring Presto and cobas (R) 4800 results, or, with discrepant Presto and cobas (R) results, two concurring results of either test together with the Lightmix Kit 480 HT CT/NC assay. For the Presto assay, we observed 77 CT positive (13%) and 22 NC positive (3,6%) vaginal samples, and 41 CT positive (6,7%) and 11 NC positive (1,8%) rectal samples. For the cobas (R) 4800 assay, we observed 77 CT positive (13%) and 21 NC positive (3,4%) vaginal samples, and 39 CT positive (6,4%) and 11 NC positive (1,8%) rectal samples. Ten CT samples were discrepant between Presto and cobas (R) 4800 CT/NC assays, while two NC samples were discrepant. CT sensitivity in both assays was 100% compared to the alloyed gold standard. The sensitivity was 100% for both vaginal and rectal dry swabs, underlining the suitability of these sample types for detection of CT and NC. The Presto assay is therefore valuable for molecular detection of CT and NC in dry vaginal and rectal swabs. (C) 2015 Elsevier B.V. All rights reserved.
引用
收藏
页码:70 / 74
页数:5
相关论文
共 18 条
[1]  
[Anonymous], SEXUALLY TRANSMITTED
[2]   Evaluation of the Abbott RealTime CT/NG Assay in Comparison to the Roche Cobas Amplicor CT/NG Assay [J].
Cheng, Annie ;
Qian, Qinfang ;
Kirby, James E. .
JOURNAL OF CLINICAL MICROBIOLOGY, 2011, 49 (04) :1294-1300
[3]   Comparison of the Abbott m2000 RealTime CT Assay and the Cepheid GeneXpert CT/NG Assay to the Roche Amplicor CT Assay for Detection of Chlamydia trachomatis in Ocular Samples from Tanzania [J].
Dize, Laura ;
West, Sheila ;
Williams, James A. ;
Van der Pol, Barbara ;
Quinn, Thomas C. ;
Gaydos, Charlotte A. .
JOURNAL OF CLINICAL MICROBIOLOGY, 2013, 51 (05) :1611-1613
[4]   Randomized comparison of vaginal self-sampling by standard vs. dry swabs for Human papillomavirus testing [J].
Eperon, Isabelle ;
Vassilakos, Pierre ;
Navarria, Isabelle ;
Menoud, Pierre-Alain ;
Gauthier, Aude ;
Pache, Jean-Claude ;
Boulvain, Michel ;
Untiet, Sarah ;
Petignat, Patrick .
BMC CANCER, 2013, 13
[5]   Performance of the APTIMA Combo 2 assay for detection of Chlamydia trachomatis and Neisseria gonorrhoeae in female urine and endocervical swab specimens [J].
Gaydos, CA ;
Quinn, TC ;
Willis, D ;
Weissfeld, A ;
Hook, EW ;
Martin, DH ;
Ferrero, DV ;
Schachter, J .
JOURNAL OF CLINICAL MICROBIOLOGY, 2003, 41 (01) :304-309
[6]   Evaluation of dry and wet transported intravaginal swabs in detection of Chlamydia trachomatis and Neisseria gonorrhoeae infections in female soldiers by PCR [J].
Gaydos, CA ;
Crotchfelt, KA ;
Shah, N ;
Tennant, M ;
Quinn, TC ;
Gaydos, JC ;
McKee, KT ;
Rompalo, AM .
JOURNAL OF CLINICAL MICROBIOLOGY, 2002, 40 (03) :758-761
[7]   Performance of the Cepheid CT/NG Xpert Rapid PCR Test for Detection of Chlamydia trachomatis and Neisseria gonorrhoeae [J].
Gaydos, Charlotte A. ;
Van der Pol, Barbara ;
Jett-Goheen, Mary ;
Barnes, Mathilda ;
Quinn, Nicole ;
Clark, Carey ;
Daniel, Grace E. ;
Dixon, Paula B. ;
Hook, Edward W., III .
JOURNAL OF CLINICAL MICROBIOLOGY, 2013, 51 (06) :1666-1672
[8]   Comparison of three commercially available amplification assays, AMP CT, LCx, and COBAS AMPLICOR, for detection of Chlamydia trachomatis in first-void urine [J].
Goessens, WHF ;
Mouton, JW ;
vanderMeijden, WI ;
Deelen, S ;
vanRijsoortVos, TH ;
Toom, NLD ;
Verbrugh, HA ;
Verkooyen, RP .
JOURNAL OF CLINICAL MICROBIOLOGY, 1997, 35 (10) :2628-2633
[9]   Clinical validation of a real-time polymerase chain reaction detection of Neisseria gonorrheae porA pseudogene versus culture techniques [J].
Hjelmevoll, Stig Ove ;
Olsen, Merethe Elise ;
Sollid, Johanna U. Ericson ;
Haaheim, Hakon ;
Melby, Kjetil K. ;
Moi, Harald ;
Unemo, Magnus ;
Skogen, Vegard .
SEXUALLY TRANSMITTED DISEASES, 2008, 35 (05) :517-520
[10]   Comparative performance of the Roche COBAS Amplicor assay and an in-house real-time PCR assay for diagnosis of Chlamydia trachomatis infection [J].
Jalal, Hamid ;
Al-Suwaine, Abdulrahman ;
Stephen, Hannah ;
Carne, Christopher ;
Sonnex, Christopher .
JOURNAL OF MEDICAL MICROBIOLOGY, 2007, 56 (03) :320-322