A critical evaluation of the PTW 2D-ARRAY seven29 and OCTAVIUS II phantom for IMRT and VMAT verification

被引:47
作者
Hussein, Mohammad [1 ,2 ]
Adams, Elizabeth J. [1 ]
Jordan, Thomas J. [1 ]
Clark, Catharine H. [1 ,3 ]
Nisbet, Andrew [1 ,2 ]
机构
[1] Royal Surrey Cty Hosp NHS Fdn Trust, Dept Med Phys, Guildford, Surrey, England
[2] Univ Surrey, Dept Phys, Guildford GU2 5XH, Surrey, England
[3] Natl Phys Lab, Teddington TW11 0LW, Middx, England
关键词
IMRT; VMAT; QA; detector arrays; TUMOR-CONTROL PROBABILITY; VOLUMETRIC MODULATED ARC; ION-CHAMBER ARRAY; QUALITY-ASSURANCE; TREATMENT DELIVERY; DOSE VERIFICATION; FILM DOSIMETRY; NORMAL TISSUE; DIODE-ARRAY; RADIOTHERAPY;
D O I
10.1120/jacmp.v14i6.4460
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Quality assurance (QA) for intensity-and volumetric-modulated radiotherapy (IMRT and VMAT) has evolved substantially. In recent years, various commercial 2D and 3D ionization chamber or diode detector arrays have become available, allowing for absolute verification with near real time results, allowing for streamlined QA. However, detector arrays are limited by their resolution, giving rise to concerns about their sensitivity to errors. Understanding the limitations of these devices is therefore critical. In this study, the sensitivity and resolution of the PTW 2D-ARRAY seven29 and OCTAVIUS II phantom combination was comprehensively characterized for use in dynamic sliding window IMRT and RapidArc verification. Measurement comparisons were made between single acquisition and a multiple merged acquisition techniques to improve the effective resolution of the 2D-ARRAY, as well as comparisons against GAFCHROMIC EBT2 film and electronic portal imaging dosimetry (EPID). The sensitivity and resolution of the 2D-ARRAY was tested using two gantry angle 0 modulated test fields. Deliberate multileaf collimator (MLC) errors of 1, 2, and 5 mm and collimator rotation errors were inserted into IMRT and RapidArc plans for pelvis and head & neck sites, to test sensitivity to errors. The radiobiological impact of these errors was assessed to determine the gamma index passing criteria to be used with the 2D-ARRAY to detect clinically relevant errors. For gamma index distributions, it was found that the 2D-ARRAY in single acquisition mode was comparable to multiple acquisition modes, as well as film and EPID. It was found that the commonly used gamma index criteria of 3% dose difference or 3 mm distance to agreement may potentially mask clinically relevant errors. Gamma index criteria of 3%/2 mm with a passing threshold of 98%, or 2%/2 mm with a passing threshold of 95%, were found to be more sensitive. We suggest that the gamma index passing thresholds may be used for guidance, but also should be combined with a visual inspection of the gamma index distribution and calculation of the dose difference to assess whether there may be a clinical impact in failed regions.
引用
收藏
页码:274 / 292
页数:19
相关论文
共 39 条
[1]  
Agazaryan Nzhde, 2003, J Appl Clin Med Phys, V4, P40, DOI 10.1120/1.1525243
[2]  
[Anonymous], 2009, J APPL CLIN MED PHYS
[3]  
[Anonymous], 2010, J ICRU, DOI [10.1093/jicru_ndq002, DOI 10.1093/JICRU_NDQ002]
[4]   IMRT: a review and preview [J].
Bortfeld, Thomas .
PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (13) :R363-R379
[5]   An intercomparison between film dosimetry and diode matrix for IMRT quality assurance [J].
Buonamici, F. Banci ;
Compagnucci, A. ;
Marrazzo, L. ;
Russo, S. ;
Bucciolini, M. .
MEDICAL PHYSICS, 2007, 34 (04) :1372-1379
[6]   FITTING OF NORMAL TISSUE TOLERANCE DATA TO AN ANALYTIC-FUNCTION [J].
BURMAN, C ;
KUTCHER, GJ ;
EMAMI, B ;
GOITEIN, M .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1991, 21 (01) :123-135
[7]   An analytical approach to acceptance criteria for quality assurance of intensity modulated radiotherapy [J].
Carver, Antony ;
Gilmore, Martyn ;
Riley, Stephen ;
Uzan, Julien ;
Mayles, Philip .
RADIOTHERAPY AND ONCOLOGY, 2011, 100 (03) :453-455
[8]   Comparison of four commercial devices for RapidArc and sliding window IMRT QA [J].
Chandraraj, Varatharaj ;
Stathakis, Sotirios ;
Manickam, Ravikumar ;
Esquivel, Carlos ;
Supe, Sanjay S. ;
Papanikolaou, Nikos .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2011, 12 (02) :338-349
[9]   Detection of IMRT delivery errors using a quantitative 2D dosimetric verification system [J].
Childress, NL ;
Bloch, C ;
White, RA ;
Salehpour, M ;
Rosen, II .
MEDICAL PHYSICS, 2005, 32 (01) :153-162
[10]   TOLERANCE OF NORMAL TISSUE TO THERAPEUTIC IRRADIATION [J].
EMAMI, B ;
LYMAN, J ;
BROWN, A ;
COIA, L ;
GOITEIN, M ;
MUNZENRIDER, JE ;
SHANK, B ;
SOLIN, LJ ;
WESSON, M .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1991, 21 (01) :109-122