A simple scatter correction method for dual energy contrast-enhanced digital breast tomosynthesis

被引:2
|
作者
Lu, Yihuan [1 ]
Lau, Beverly [2 ]
Hu, Yue-Houng [2 ]
Zhao, Wei [2 ]
Gindi, Gene [1 ,2 ]
机构
[1] SUNY Stony Brook, Dept Elect Engn, Stony Brook, NY 11790 USA
[2] SUNY Stony Brook, Dept Radiol, Stony Brook, NY 11794 USA
来源
MEDICAL IMAGING 2014: PHYSICS OF MEDICAL IMAGING | 2014年 / 9033卷
关键词
digital breast tomosynthesis; scatter correction; dual energy; contrast enhanced; mammography; MAMMOGRAPHY;
D O I
10.1117/12.2043190
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
Dual-Energy Contrast Enhanced Digital Breast Tomosynthesis (DE-CE-DBT) has the potential to deliver diagnostic information for vascularized breast pathology beyond that available from screening DBT. DE-CE-DBT involves a contrast (iodine) injection followed by a low energy (LE) and a high energy (HE) acquisitions. These undergo weighted subtraction then a reconstruction that ideally shows only the iodinated signal Scatter in the projection data leads to "cupping" artifacts that can reduce the visibility and quantitative accuracy of the iodinated signal The use of filtered backprojection (FBP) reconstruction ameliorates these types of artifacts, but the use of FBP precludes the advantages of iterative reconstructions. This motivates an effective and clinically practical scatter correction (SC) method for the projection data. We propose a simple SC method, applied at each acquisition angle. It uses scatter-only data at the edge of the image to interpolate a scatter estimate within the breast region. The interpolation has an approximately correct spatial profile but is quantitatively inaccurate. We further correct the interpolated scatter data with the aid of easily obtainable knowledge of SPR (scatter-to-primary ratio) at a single reference point. We validated the SC method using a CIRS breast phantom with iodine inserts. We evaluated its efficacy in terms of SDNR and iodine quantitative accuracy. We also applied our SC method to a patient DE-CE-DBT study and showed that the SC allowed detection of a previously confirmed tumor at the edge of the breast. The SC method is quick to use and may be useful in a clinical setting.
引用
收藏
页数:12
相关论文
共 50 条
  • [1] A scatter correction method for contrast-enhanced dual-energy digital breast tomosynthesis
    Lu, Yihuan
    Peng, Boyu
    Lau, Beverly A.
    Hu, Yue-Houng
    Scaduto, David A.
    Zhao, Wei
    Gindi, Gene
    PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (16) : 6323 - 6354
  • [2] Effect of scatter correction on image noise in contrast-enhanced digital breast tomosynthesis
    Huang, Hailiang
    Duan, Xiaoyu
    Sahu, Pranjal
    Zhao, Wei
    15TH INTERNATIONAL WORKSHOP ON BREAST IMAGING (IWBI2020), 2020, 11513
  • [3] Segmentation methods for breast vasculature in dual-energy contrast-enhanced digital breast tomosynthesis
    Lau, Kristen C.
    Lee, Hyo Min
    Singh, Tanushriya
    Maidment, Andrew D. A.
    MEDICAL IMAGING 2015: PHYSICS OF MEDICAL IMAGING, 2015, 9412
  • [4] Experimental quantification of lesion detectability in contrast enhanced dual energy digital breast tomosynthesis
    Hu, Yue-Houng
    Zhao, Wei
    MEDICAL IMAGING 2012: PHYSICS OF MEDICAL IMAGING, 2012, 8313
  • [5] Design of a contrast-enhanced dual-energy tomosynthesis system for breast cancer imaging
    Hoernig, M. D.
    Baetz, L.
    Mertelmeier, T.
    MEDICAL IMAGING 2012: PHYSICS OF MEDICAL IMAGING, 2012, 8313
  • [6] Impact of image acquisition timing on image quality for dual-energy contrast-enhanced breast tomosynthesis
    Hill, Melissa L.
    Mainprize, James G.
    Puong, Sylvie
    Carton, Ann-Katherine
    Iordache, Razvan
    Muller, Serge
    Yaffe, Martin J.
    MEDICAL IMAGING 2012: PHYSICS OF MEDICAL IMAGING, 2012, 8313
  • [7] Quantification of contrast-enhanced digital breast tomosynthesis
    Maidment, A. D. A.
    Carton, A. -K.
    Chen, S.
    Conant, E. F.
    Schnal, M.
    INTERNATIONAL JOURNAL OF COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2006, 1 : 12 - 14
  • [8] Quantification for contrast-enhanced digital breast tomosynthesis
    Carton, Ann-Katherine
    Li, Jingjing
    Albert, Michael
    Chen, Sara
    Maidment, Andrew D. A.
    MEDICAL IMAGING 2006: PHYSICS OF MEDICAL IMAGING, PTS 1-3, 2006, 6142
  • [9] Effects of image lag and scatter for dual-energy contrast-enhanced digital breast tomosynthesis using a CsI flat-panel based system
    Carton, Ann-Katherine
    Puong, Sylvie
    Iordache, Razvan
    Muller, Serge
    MEDICAL IMAGING 2011: PHYSICS OF MEDICAL IMAGING, 2011, 7961
  • [10] Optimization of a dual-energy contrast-enhanced technique for a photon-counting digital breast tomosynthesis system: II. An experimental validation
    Carton, Ann-Katherine
    Ullberg, Christer
    Maidment, Andrew D. A.
    MEDICAL PHYSICS, 2010, 37 (11) : 5908 - 5913