A Novel Study of Mechanical Heart Valve Cavitation in a Pressurized Pulsatile Duplicator

被引:6
作者
Wu, Changfu [1 ]
Retta, Stephen M. [2 ]
Robinson, Ronald A. [2 ]
Herman, Bruce A. [2 ]
Grossman, Laurence W. [2 ]
机构
[1] Florida Atlantic Univ, Ctr Appl Stochast Res, Coll Engn & Comp Sci, Boca Raton, FL 33431 USA
[2] US FDA, Off Sci & Engn Labs, Div Solid & Fluid Mech, Ctr Devices & Radiol Hlth, Silver Spring, MD USA
基金
美国国家科学基金会;
关键词
CLOSING SOUNDS; PROSTHESIS;
D O I
10.1097/MAT.0b013e3181b4c44f
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Submission of data regarding the cavitation potential of a mechanical heart valve is recommended by the U.S. Food and Drug Administration in the device-review process. An acoustic method has long been proposed for cavitation detection. However, the question as to whether such a method can differentiate the cavitation noise from the mechanical closing sound has not been sufficiently addressed. in this study, cavitation near a Medtronic Hall tilting disc valve was investigated in a pressurized pulsatile duplicator. The purpose of pressurizing the testing chambers was to prevent cavitation under a normally cavitating loading condition to isolate the mechanical closing sound. By comparing the sound signals before and after pressurization, some noticeable differences were found between them. In the time domain, the intensity of the sound under a cavitating condition was much higher. In the frequency domain, the energy distribution of a sound signal was distinctively different depending on whether cavitation occurred or not. The valve closing sound had a large amount of energy in the low-frequency range (less than about 25 kHz). When cavitation took place, the sound energy shifted toward the high-frequency range (from 25 to 500 kHz). ASAIO Journal 2009; 55:445-451.
引用
收藏
页码:445 / 451
页数:7
相关论文
共 34 条
[1]   Intraoperative and postoperative evaluation of cavitation in mechanical heart valve patients [J].
Andersen, TS ;
Johansen, P ;
Christensen, BO ;
Paulsen, PK ;
Nygaard, H ;
Hasenkam, JM .
ANNALS OF THORACIC SURGERY, 2006, 81 (01) :34-41
[2]  
Andersen TS, 2003, J HEART VALVE DIS, V12, P790
[3]   A SQUEEZE FLOW PHENOMENON AT THE CLOSING OF A BILEAFLET MECHANICAL HEART-VALVE PROSTHESIS [J].
BLUESTEIN, D ;
EINAV, S ;
HWANG, NHC .
JOURNAL OF BIOMECHANICS, 1994, 27 (11) :1369-1378
[4]  
DEUVAERT FE, 1989, ACTA CHIR BELG, V89, P15
[5]   In vivo demonstration of cavitation potential of a mechanical heart valve [J].
Dexter, EU ;
Aluri, S ;
Radcliffe, RR ;
Zhu, H ;
Carlson, DD ;
Heilman, TE ;
Chandran, KB ;
Richenbacher, WE .
ASAIO JOURNAL, 1999, 45 (05) :436-441
[6]  
*FDA, 1994, REPL HEART VALV GUID
[7]  
GARRISON LA, 1994, J HEART VALVE DIS S1, V3, pS8
[8]   CAVITATION POTENTIAL OF MECHANICAL HEART-VALVE PROSTHESES [J].
GRAF, T ;
FISCHER, H ;
REUL, H ;
RAU, G .
INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 1991, 14 (03) :169-174
[9]   LASER ASSESSMENT OF LEAFLET CLOSING MOTION IN PROSTHETIC HEART-VALVES [J].
GUO, GX ;
XU, CC ;
HWANG, NHC .
JOURNAL OF BIOMEDICAL ENGINEERING, 1990, 12 (06) :477-481
[10]  
HAUBOLT A, 1996, J HEART VALVE DIS S1, V5