Performance evaluation for pinhole collimators of small gamma camera by MTF and NNPS analysis: Monte Carlo simulation study

被引:3
作者
Jeon, Hosang [1 ]
Kim, Hyunduk [1 ]
Cha, Bo Kyung [1 ]
Kim, Jong Yul [1 ]
Cho, Gyuseong [1 ]
Chung, Yong Hyun [2 ]
Yun, Jong-Il [1 ]
机构
[1] Korea Adv Inst Sci & Technol, Dept Nucl & Quantum Engn, Taejon 305701, South Korea
[2] Yonsei Univ, Dept Radiol Sci, Wonju, South Korea
关键词
MTF; NNPS; FWHM; Sensitivity; Uniformity; Gamma camera; MODULATION TRANSFER-FUNCTION;
D O I
10.1016/j.nima.2009.01.037
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Presently, the gamma camera system is widely used in various medical diagnostic, industrial and environmental fields. Hence, the quantitative and effective evaluation of its imaging performance is essential for design and quality assurance. The National Electrical Manufacturers Association (NEMA) standards for gamma camera evaluation are insufficient to perform sensitive evaluation. In this study, modulation transfer function (MTF) and normalized noise power spectrum (NNPS) will be suggested to evaluate the performance of small gamma camera with changeable pinhole collimators using Monte Carlo Simulation. We simulated the system with a cylinder and a disk source, and seven different pinhole collimators from 1- to 4-mm-diameter pinhole with lead. The MTF and NNPS data were obtained from output images and were compared with full-width at half-maximum (FWHM), sensitivity and differential uniformity. In the result, we found that MTF and NNPS are effective and novel standards to evaluate imaging performance of gamma cameras instead of conventional NEMA standards. (C) 2009 Elsevier B.V. All rights reserved.
引用
收藏
页码:93 / 96
页数:4
相关论文
共 13 条
  • [1] SCINTILLATION CAMERA
    ANGER, HO
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1958, 29 (01) : 27 - 33
  • [2] BORASI G, 2003, MED PHYS, V30
  • [3] DQE(F) OF 4 GENERATIONS OF COMPUTED RADIOGRAPHY ACQUISITION DEVICES
    DOBBINS, JT
    ERGUN, DL
    RUTZ, L
    HINSHAW, DA
    BLUME, H
    CLARK, DC
    [J]. MEDICAL PHYSICS, 1995, 22 (10) : 1581 - 1593
  • [4] GROSSMAN L, 1986, PHYS MED BIOL, V31
  • [5] HANDER TA, 1997, MED PHYS, V24
  • [6] IAN A, 2000, HDB MED IMAGING, V1, pCH2
  • [7] *IEC, 2002, 622201 IEC
  • [8] USE OF THE MODULATION TRANSFER-FUNCTION IN COMPARING THE VISUAL OUTCOME UNITS OF GAMMA-CAMERA
    KAIRENTO, AL
    OLKINUORA, P
    KOSKINEN, M
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1980, 5 (01): : 15 - 17
  • [9] LOGAN KW, 1983, MED PHYS, V10
  • [10] *NEMA, NU12001 NEMA