Evaluation of small mass spectrometer systems for permanent gas analysis

被引:12
作者
Arkin, CR
Griffin, TP
Ottens, AK
Diaz, JA
Follistein, DW
Adams, FW
Helms, WR
机构
[1] Dynacs Inc, Kennedy Space Ctr, FL 32815 USA
[2] Univ Florida, Dept Chem, Gainesville, FL 32611 USA
[3] Univ Costa Rica, Escuela Fis, San Jose, Costa Rica
[4] NASA, Kennedy Space Ctr, FL USA
关键词
D O I
10.1016/S1044-0305(02)00422-1
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
This work is aimed at understanding the aspects of designing a miniature mass spectrometer (MS) system. Several types of small MS systems are evaluated and discussed, including linear quadrupole, quadrupole ion trap, time of flight, and sector. Analysis of hydrogen, helium, oxygen, and argon in a nitrogen background with the concentrations of the components of interest ranging from 0 to 5000 parts per million (ppm). The performance of each system in terms of accuracy, precision, limits of detection, response time, recovery time, scan rate, size, and weight is assessed. The relative accuracies of the systems varied from <1% to similar to40% with an average below 10%. Relative precisions varied from 1% to 20%, with an average below 5%. The detection limits had a large distribution, ranging from 0.2 to 170 ppm. The systems had a diverse response time ranging from 4 to 210 s, as did the recovery time with a 6-to-210-s distribution. Most instruments had scan times near 1 s; however, one instrument exceeded 13 s. System weights varied from 9 to 52 kg and sizes ranged from 15 x 10(3) cm(3) to 110 x 10(3) cm(3). A performance scale is set up to rank each system, and an overall performance score is given to each system. (C) 2002 American Society for Mass Spectrometry.
引用
收藏
页码:1004 / 1012
页数:9
相关论文
共 24 条
[1]  
ARKIN CR, 2001, P 49 ASMS C MASS SPE
[2]  
Badman ER, 2000, J MASS SPECTROM, V35, P659, DOI 10.1002/1096-9888(200006)35:6<659::AID-JMS5>3.0.CO
[3]  
2-V
[4]   Trade-offs in miniature quadrupole designs [J].
Boumsellek, S ;
Ferran, RJ .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2001, 12 (06) :633-640
[5]  
DAWSON PH, 1976, QUADRUPOLE MASS SPEC, P117
[6]   Sub-miniature ExB sector-field mass spectrometer [J].
Diaz, JA ;
Giese, CF ;
Gentry, WR .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2001, 12 (06) :619-632
[7]  
DIAZ JA, 1999, THESIS U MINNESOTA
[8]  
DIAZ JA, 1999, Patent No. 96035
[9]  
DIAZ JA, 1998, THESIS U MINNESOTA
[10]   High-pressure effects in miniature arrays of quadrupole analyzers for residual gas analysis from 10(-9) to 10(-2) Torr [J].
Ferran, RJ ;
Boumsellek, S .
JOURNAL OF VACUUM SCIENCE & TECHNOLOGY A, 1996, 14 (03) :1258-1265