Measurement of PCDDs, PCDFs, and non-ortho-PCBs by comprehensive two-dimensional gas chromatography-isotope dilution time-of-flight mass spectrometry (GC x GC-IDTOFMS)

被引:53
作者
Focant, JF
Reiner, EJ
MacPherson, K
Kolic, T
Sjödin, A
Patterson, DG
Reese, SL
Dorman, FL
Cochran, J
机构
[1] Ctr Dis Control & Prevent, Natl Ctr Environm Hlth, Div Sci Lab, OAT Branch, Atlanta, GA 30341 USA
[2] Minist Environm, Toronto, ON M9P 3V6, Canada
[3] Folsom Lake Coll, Folsom, CA 95630 USA
[4] Restek Corp, Bellefonte, PA 16823 USA
[5] LECO Corp, Las Vegas, NV 89119 USA
关键词
polychlorinated dibenzo-p-dioxins (PCDDs); polychlorinated dibenzofurans (PCDFs); coplanar polychlorinated biphenyls (cPCBs); comprehensive two-dimensional gas chromatography (GC x GC); time-of-flight mass spectrometry (TOFMS);
D O I
10.1016/j.talanta.2004.05.043
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Comprehensive two-dimensional gas chromatography with isotope-dilution time-of-flight mass spectrometry (GC x GC-IDTOFMS) was used to measure polychlorinated dibenzo-p-dioxin (PCDD), polychlorinated dibenzofuran (PCDF), and coplanar polychlorinated biphenyl (cPCB) concentrations in ash, sediment, vegetation, and fish samples. The GC x GC capability was achieved by using a quad jet, dual stage, thermal modulator. Zone compression of the GC peaks from modulation resulted in a significant increase of the signal intensity over classical GC-IDTOFMS. The GC x GC column set used an Rtx-Dioxin 2 phase as the first dimension (D-1) and an Rtx-500 as the second dimension (D-2). The chromatographic separation of the 17 PCDD/Fs and the 4 cPCBs was attained in D-1 except for 2,3,7,8-TCDD and CB126 for which deconvoluted ion currents (DIC) were required to be reported separately. The Rtx-500 phase separated the bulk matrix interfering compounds from the target analytes in D-2. The instrumental limit of detection (iLODs) was 0.5 pg for 2,3,7,8-TCDD. The calibration curves showed good correlation coefficients for all the compounds investigated in the concentration range of 0.5-200 pg. GC x GC-IDTOFMS results compared favorably to those from conventional isotope-dilution one-dimensional gas chromatography-high resolution mass spectrometry (GC-IDHRMS). The comprehensive mass analysis of the TOFMS further permitted the identification of other contaminants of concern in the samples. (C) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:1231 / 1240
页数:10
相关论文
共 35 条
[31]   Toxic equivalency factors (TEFs) for PCBs, PCDDs, PCDFs for humans and wildlife [J].
Van den Berg, M ;
Birnbaum, L ;
Bosveld, ATC ;
Brunström, B ;
Cook, P ;
Feeley, M ;
Giesy, JP ;
Hanberg, A ;
Hasegawa, R ;
Kennedy, SW ;
Kubiak, T ;
Larsen, JC ;
van Leeuwen, FXR ;
Liem, AKD ;
Nolt, C ;
Peterson, RE ;
Poellinger, L ;
Safe, S ;
Schrenk, D ;
Tillitt, D ;
Tysklind, M ;
Younes, M ;
Waern, F ;
Zacharewski, T .
ENVIRONMENTAL HEALTH PERSPECTIVES, 1998, 106 (12) :775-792
[32]  
van Deursen M, 2000, HRC-J HIGH RES CHROM, V23, P507
[33]   Evaluation of time-of-flight mass spectrometric detection for fast gas chromatography [J].
van Deursen, MM ;
Beens, J ;
Janssen, HG ;
Leclercq, PA ;
Cramers, CA .
JOURNAL OF CHROMATOGRAPHY A, 2000, 878 (02) :205-213
[34]   A HIGH-SPEED GAS-CHROMATOGRAPH COUPLED TO A TIME-OF-FLIGHT MASS ANALYZER [J].
WOLLNIK, H ;
BECKER, R ;
GOTZ, H ;
KRAFT, A ;
JUNG, H ;
CHEN, CC ;
VANYSACKER, PG ;
JANSEN, HG ;
SNIJDERS, HMJ ;
LECLERCQ, PA ;
CRAMER, CA .
INTERNATIONAL JOURNAL OF MASS SPECTROMETRY AND ION PROCESSES, 1994, 130 (03) :L7-L11
[35]   Evaluation of two-dimensional gas chromatography-time- of-flight mass spectrometry for the determination of multiple pesticide residues in fruit [J].
Zrostlíková, J ;
Hajslová, J ;
Cajka, T .
JOURNAL OF CHROMATOGRAPHY A, 2003, 1019 (1-2) :173-186