Biomechanical Comparison of Fixed- versus Variable-Angle Locking Screws for Distal Humerus Comminuted Fractures

被引:10
|
作者
Nourbakhsh, Ali [1 ]
Hirschfeld, Adam G. [2 ]
Dhulipala, Sravan [1 ]
Hutton, William [3 ]
Ganey, Timothy [1 ]
Lozada, Luis [1 ]
Schlatterer, Daniel [1 ]
Lourie, Gary Mark [4 ]
机构
[1] WellStar Atlanta Med Ctr, Dept Orthoped, Atlanta, GA USA
[2] MetroHlth Med Ctr, Dept Orthoped, Cleveland, OH USA
[3] Emory Univ, Sch Med, Dept Orthoped, Atlanta, GA USA
[4] Hand & Upper Extrem Ctr Georgia, Dept Orthoped, 980 Johnson Ferry Rd NE,Suite 1020, Atlanta, GA 30342 USA
关键词
Humerus; Fracture fixation; Bone screws; POLYAXIAL LOCKING; PLATE FIXATION; ELBOW JOINT; FORCES;
D O I
10.4055/cios.2019.11.3.302
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: To compare the stability of fixed- versus variable-angle locking constructs for the comminuted distal humerus fracture (AO/OTA 13-A3). Methods: Eight pairs of complete humeri harvested from eight fresh frozen cadavers were used for the study. We fixed the intact humeri using 2.7-mm/3.5-mm locking VA-LCP stainless steel distal humerus posterolateral (nine-hole) and medial (seven-hole) plates. An oscillating saw was used to cut a 1-cm gap above the olecranon fosse. The specimens were loaded in axial mode with the rate of 1 mm per 10 seconds to failure, and stress-strain curves were compared in each pair. The mode of failure was recorded as well as the load needed for 2- and 4-mm displacement at the lateral end of the gap. Results: The stiffness of the constructs, based on the slope of the stress-strain curve, did not show any difference between the fixed- versus variable-angle constructs. Likewise, there was no difference between the force needed for 2- or 4-mm displacement at the lateral gap between the fixed- and variable-angle constructs. The mode of failure was bending of both plates in all specimens and screw pull-out in four specimen pairs in addition to the plate bending. Conclusions: Our results did not show any difference in the biomechanical stability of the fixed- versus variable-angle constructs. There was not any screw breakage or failure of the plate-screw interface.
引用
收藏
页码:302 / 308
页数:7
相关论文
共 39 条
  • [31] A Biomechanical Comparison of Locked Unicortical with Non-Locking Bicortical Screws for Metacarpal Shaft Fractures in a Randomized Human Cadaveric Model
    Minogue, Febe
    Reilly, Keryn
    Walker, Cameron
    Malak, Sharif
    Mayo, Andrew
    Heiss-Dunlop, Wolfgang
    JOURNAL OF HAND SURGERY-ASIAN-PACIFIC VOLUME, 2021, 26 (01) : 10 - 16
  • [32] Biomechanical analysis of blade plate versus locking plate fixation for a proximal humerus fracture: Comparison using cadaveric and synthetic humeri
    Siffri, Paul C.
    Peindl, Richard D.
    Coley, Edward R.
    Norton, James
    Connor, Patrick M.
    Kellam, James E.
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2006, 20 (08) : 547 - 554
  • [33] Biomechanical Analysis of Volar and Dorsal Double Locking Plates for Fixation in Comminuted Extra-articular Distal Radius Fractures: A 3D Finite Element Study
    Lin, Yu-Hao
    Lin, Chun-Li
    Kuo, Hsien-Nan
    Sun, Ming-Tsung
    Chen, Alvin Chao-Yu
    JOURNAL OF MEDICAL AND BIOLOGICAL ENGINEERING, 2012, 32 (05) : 349 - 355
  • [34] Biomechanical Comparison of Titanium Locking Fragment-Specific and Volar Locking Plates for AO B1 and B2 Fractures of the Distal Radius
    Martin, Dennis P.
    Park, Andrew G.
    Jamison, David
    Jones, Christopher M.
    Ilyas, Asif M.
    JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2019, 44 (12): : 1093.e1 - 1093.e8
  • [35] Comparison of Treatment of Acute Unstable Distal Clavicle Fractures Using Anatomical Locking Plates with Versus without Additional Suture Anchor Fixation
    Fan, Jixing
    Zhang, Yufu
    Huang, Qiang
    Jiang, Xieyuan
    He, Liang
    MEDICAL SCIENCE MONITOR, 2017, 23 : 5455 - 5461
  • [36] Biomechanical Comparison of Dorsal Nail Plate Versus Screw and K-Wire Construct for Extra-Articular Distal Radius Fractures in a Cadaver Bone Model
    Klitscher, Daniela
    Mehling, Isabella
    Nowak, Lukas
    Nowak, Tobias
    Rornmens, Pol M.
    Mueller, Lars P.
    JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2010, 35A (04): : 611 - 618
  • [37] Biomechanischer Vergleich von zwei winkelstabilen, anatomisch angepassten Plattensystemen am distalen HumerusBiomechanical comparison of two angle-stable, anatomically designed plate systems on distal humerus fractures
    S. Hungerer
    R. Penzkofer
    P. Augat
    V. Bühren
    Obere Extremität, 2010, 5 (3) : 165 - 172
  • [38] An In Vitro Biomechanical Comparison of Equine Proximal Interphalangeal Joint Arthrodesis Techniques: Two Parallel Transarticular Headless Tapered Variable Pitch Screws Versus Two Parallel Transarticular AO Cortical Bone Screws Inserted in Lag Fashion
    Gudehus, Timm
    Sod, Gary A.
    Riggs, Laura M.
    Mitchell, Colin F.
    Martin, George S.
    VETERINARY SURGERY, 2011, 40 (03) : 261 - 265
  • [39] Comparison of palmar fixed-angle plate fixation with K-wire fixation of distal radius fractures (A0 A2, A3, C1) in elderly patients
    Goehre, F.
    Otto, W.
    Schwan, S.
    Mendel, T.
    Vergroesen, P. P.
    Lindemann-Sperfeld, L.
    JOURNAL OF HAND SURGERY-EUROPEAN VOLUME, 2014, 39 (03) : 249 - 257