Planar Differential Mobility Analyzer with a Resolving Power of 110

被引:25
作者
Amo-Gonzalez, Mario [1 ]
Perez, Sergio [1 ]
机构
[1] SEADM, Parque Tecnol Boecillo 205, Valladolid 47151, Spain
关键词
SPECTROMETRY; SEPARATION; IONS;
D O I
10.1021/acs.analchem.8b00579
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Planar differential mobility analyzers (DMAs) have previously achieved resolving powers of 60-80 in air or N-2 at the mobility of the tetraheptylammonium ion (THA(+), similar to 0.97 cm(2)/V/s). For unclear reasons, this performance is considerably below the theoretical limit. In this work, a performance close to this ideal limit is attained in SEADM's P5 DMA via improved flow laminarization, under otherwise the same flow conditions as in prior work. The new laminarizer remains effective at unusually large gas velocities (reached with two blowers in series), yielding a resolving power of 110. The selectivity of the improved DMA combined with a mass spectrometer was assessed by the analysis of a real sample of extra virgin olive oil.
引用
收藏
页码:6735 / 6741
页数:7
相关论文
共 35 条
  • [1] Isomer separation of polybrominated diphenyl ether metabolites using nanoESI-TIMS-MS
    Adams, Kendra J.
    Montero, Dennise
    Aga, Diana
    Fernandez-Lima, Francisco
    [J]. INTERNATIONAL JOURNAL FOR ION MOBILITY SPECTROMETRY, 2016, 19 (2-3) : 69 - 76
  • [2] Amo M., 2016, U. S. patent, Patent No. [9,297,785, 9297785]
  • [3] Mobility Peak Tailing Reduction in a Differential Mobility Analyzer (DMA) Coupled with a Mass Spectrometer and Several Ionization Sources
    Amo-Gonzalez, Mario
    de la Mora, Juan Fernandez
    [J]. JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2017, 28 (08) : 1506 - 1517
  • [4] [Anonymous], 1988, THESIS
  • [5] Asbury GR, 2000, J MICROCOLUMN SEP, V12, P172, DOI 10.1002/(SICI)1520-667X(2000)12:3<172::AID-MCS7>3.3.CO
  • [6] 2-N
  • [7] Mass and Charge Distribution Analysis in Negative Electrosprays of Large Polyethylene Glycol Chains by Ion Mobility Mass Spectrometry
    Criado-Hidalgo, Ernesto
    Fernandez-Garcia, Juan
    de la Mora, Juan Fernandez
    [J]. ANALYTICAL CHEMISTRY, 2013, 85 (05) : 2710 - 2716
  • [8] Fast high-resolution nanoDMA measurements with a 25ms response time electrometer
    de la Mora, Juan Fernandez
    Perez-Lorenzo, Luis Javier
    Arranz, Gonzalo
    Amo-Gonzalez, Mario
    Burtscher, Heinz
    [J]. AEROSOL SCIENCE AND TECHNOLOGY, 2017, 51 (06) : 724 - 734
  • [9] High-resolution ion mobility measurements
    Dugourd, P
    Hudgins, RR
    Clemmer, DE
    Jarrold, MF
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1997, 68 (02) : 1122 - 1129
  • [10] Fernandez de la Mora J., 2012, U. S. Patent, Patent No. [8,278,622, 8278622]