DA-ME: revitalizing a low-background counting chamber and developing a radon-tight sample holder for gamma-ray spectroscopy measurements

被引:9
|
作者
Kis, Zoltan [1 ]
Voelgyesi, Peter [2 ]
Szabo, Zsuzsanna [2 ,3 ]
机构
[1] Hungarian Acad Sci, Nucl Anal & Radiog Dept, Energy Res Ctr, H-1525 Budapest 114, Hungary
[2] Eotvos Lorand Univ, Lithosphere Fluid Res Lab, Dept Petrol & Geochem, H-1117 Budapest, Hungary
[3] Hungarian Acad Sci, Environm Phys Dept, Energy Res Ctr, H-1525 Budapest 114, Hungary
关键词
Gamma-ray spectroscopy; Low-background chamber; Radon-tight HDPE container; Efficiency transfer; True coincidence summing; ABSOLUTE PEAK EFFICIENCY; PERMEABILITY; SPECTROMETRY; GEOMETRIES; DETECTORS; RA-226;
D O I
10.1007/s10967-013-2691-8
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
There was an emerging need at the Nuclear Analysis and Radiography Department of the Centre for Energy Research, Hungarian Academy of Sciences, Budapest to be able to perform low level radioactivity measurements of various samples from in-beam activation and from environmental studies. Important aspects of reusing the low-background chamber called DA-ME, which had been unused for some years, were the development of an easily reusable radon-tight sample container and the setup of a measurement system capable of counting extended samples in close-in geometry. As a result of our efforts a special sample container made of HDPE (High-density Polyethylene) was developed, and it is proved that the probability of a radon loss larger than the 2 % of its radioactive decay constant is < 95 %. Due to the lack of reference samples, containing the same radionuclides as the unknown sample, the absolute method of measuring the activity concentration of nuclides in the sample had to be applied, which implied the reliable determination of the full-energy peak efficiency. A method called efficiency transfer combined with the correction for true coincidence summing effects is proven to be providing appropriate results and applied.
引用
收藏
页码:2029 / 2035
页数:7
相关论文
共 1 条
  • [1] DÖME: revitalizing a low-background counting chamber and developing a radon-tight sample holder for gamma-ray spectroscopy measurements
    Zoltán Kis
    Péter Völgyesi
    Zsuzsanna Szabó
    Journal of Radioanalytical and Nuclear Chemistry, 2013, 298 : 2029 - 2035