A literature review of the association between wear rate and osteolysis in total hip arthroplasty

被引:503
作者
Dumbleton, JH
Manley, MT
Edidin, AA
机构
[1] School of Biomedical Engineering Science and Health Systems, Drexel University, Philadelphia, PA
关键词
total hip arthroplasty (THA); osteolysis; wear; polyethylene;
D O I
10.1054/arth.2002.33664
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
The establishment of a polyethylene wear rate threshold for the development of osteolysis at the hip would allow surgeons to identify patients at risk for osteolysis and to implement selective, more frequent follow-up. We reviewed publications that met certain criteria for wear and osteolysis measurement. Based on this review, the incidence of osteolysis increases as the rate of wear increases. The literature indicates that osteolysis rarely is observed at a wear rate of <0.1 mm/y. We suggest that a practical wear rate threshold of 0.05 mm/y would eliminate osteolysis. This wear threshold suggests that the new cross-linked polyethylenes would reduce osteolysis, provided that in vivo wear rates mirror those observed in vitro. To facilitate future comparison of published data, we suggest that longitudinal wear studies adopt consistent edge detection-based wear measurement techniques and uniform osteolytic lesion classification and measurement schema.
引用
收藏
页码:649 / 661
页数:13
相关论文
共 71 条
[1]  
[Anonymous], CLIN ORTHOPAEDICS
[2]   The porous-coated anatomic total hip prosthesis: Failure of the metal-backed acetabular component [J].
Astion, DJ ;
Saluan, P ;
Stulberg, BN ;
Rimnac, CM ;
Li, S .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1996, 78A (05) :755-766
[3]  
Barrack RL, 1997, CLIN ORTHOP RELAT R, P211
[4]   The pathology of total joint arthroplasty - II. Mechanisms of implant failure [J].
Bauer, TW ;
Schils, J .
SKELETAL RADIOLOGY, 1999, 28 (09) :483-497
[5]   Charnley total hip arthroplasty with cement - Minimum twenty-five-year follow-up [J].
Callaghan, JJ ;
Albright, JC ;
Goetz, DD ;
Olejniczak, JP ;
Johnston, RC .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2000, 82A (04) :487-497
[6]  
Campbell D G, 1999, Instr Course Lect, V48, P43
[7]  
Campbell P, 1996, Proc Inst Mech Eng H, V210, P167, DOI 10.1243/PIME_PROC_1996_210_409_02
[8]   Hydroxyapatite-coated total hip femoral components in patients less than fifty years old - Clinical and radiographic results after five to eight years of follow-up [J].
Capello, WN ;
DAntonio, JA ;
Feinberg, JR ;
Manley, MT .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1997, 79A (07) :1023-1029
[9]  
CHEN JH, 1996, J ORTHOP SURG ROC, V13, P59
[10]  
COLLIER MB, 2001, T ORTHOP RES SOC, V26, P164