Technical description, phantom accuracy, and clinical feasibility for fiducial-free frameless real-time image-guided spinal radiosurgery

被引:64
作者
Muacevic, Alexander
Staehler, Michael
Drexler, Christian
Wowra, Berndt
Reiser, Maximilian
Tonn, Joerg-Christian
机构
[1] European CyberKnife Ctr Munich, D-81377 Munich, Germany
[2] Univ Hosp Munich, Dept Urol, Munich, Germany
[3] Univ Hosp Munich, Dept Clin Radiol, Munich, Germany
[4] Univ Hosp Munich, Dept Neurosurg, Munich, Germany
关键词
spine; CyberKnife; radiosurgery; fiducial-free registration;
D O I
10.3171/spi.2006.5.4.303
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Object. The authors describe the technical application of the Xsight Spine Tracking System, data pertaining to accuracy obtained during phantom testing, and the initial clinical feasibility of using this fiducial-free alignment system with the CyberKnife in spinal radiosurgery. Methods. The Xsight integrates with the CyberKnife radiosurgery system to eliminate the need for implantation of radiographic markers or fiducials prior to spinal radiosurgery. It locates and tracks spinal lesions relative to spinal osseous landmarks. The authors performed 10 end-to-end tests of accuracy using an anthropomorphic head and cervical spine phantom. Xsight was also used in the treatment of 50 spinal lesions in 42 patients. Dose planning was based on 1.5-mm-thick computed tomography slices in which an inverse treatment planning technique was used. All lesions could be treated using the fiducial-free tracking procedure. Phantom tests produced an overall mean targeting error of 0.52 +/- 0.22 mm. The setup time for patient alignment averaged 6 minutes (range 2-45 minutes). The treatment doses varied from 12 to 25 Gy to the median prescription isodose of 65% (40 to 70%). The tumor volume ranged between 1.3 and 152.8 cm(3). The mean spinal cord volume receiving greater than 8 Gy was 0.69 +/- 0.35 cm(3). No short-term adverse events were noted during the 1- to 7-month follow-up period. Axial and radicular pain was relieved in 14 of 15 patients treated for pain. Conclusions. Fiducial-free tracking is a feasible, accurate, and reliable tool for radiosurgery of the entire spine. By eliminating the need for fiducial implantation, the Xsight system offers patients noninvasive radiosurgical intervention for intra- and paraspinal tumors.
引用
收藏
页码:303 / 312
页数:10
相关论文
共 14 条
  • [11] The cyberknife stereotactic radiosurgery system: Description, installation, and an initial evaluation of use and functionality
    Kuo, JS
    Yu, C
    Petrovich, Z
    Apuzzo, MIJ
    [J]. NEUROSURGERY, 2003, 53 (05) : 1235 - 1239
  • [12] Patterns of patient movement during frameless image-guided radiosurgery
    Murphy, MJ
    Chang, SD
    Gibbs, IC
    Le, QT
    Hai, J
    Kim, D
    Martin, DP
    Adler, JR
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2003, 55 (05): : 1400 - 1408
  • [13] Image-guided hypo-fractionated stereotactic radiosurgery to spinal lesions
    Ryu, SI
    Chang, SD
    Kim, DH
    Murphy, MJ
    Le, QT
    Martin, DP
    Adler, JR
    [J]. NEUROSURGERY, 2001, 49 (04) : 838 - 846
  • [14] Accuray CyberKnife® image-guided radiosurgical system
    Welch, WC
    Gerszten, PC
    [J]. EXPERT REVIEW OF MEDICAL DEVICES, 2005, 2 (02) : 141 - 147