New algorithm for extracting motion information from PROPELLER data and head motion correction in T1-weighted MRI

被引:0
作者
Feng, Yanqiu [1 ]
Chen, Wufan [1 ]
机构
[1] So Med Univ, Dept Biomed Engn, Guangzhou 510515, Peoples R China
来源
2005 27TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-7 | 2005年
关键词
motion correction; PROPELLER; mutual information; correlation; T1-weighted; MRI;
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
PROPELLER (Periodically Rotated Overlapping ParallEl Lines with Enhanced Reconstruction) MRI, proposed by J. G. Pipe [1], offers a novel and effective means for compensating motion. For the reconstruction of PROPLLER data, algorithms to retiably and accurately extract inter-strip motion from data in central overlapped area are crucial to motion artifacts suppression. When implemented on T1-weighted MR data, the reconstruction algorithm, with motion estimated by registration based on maximizing correlation energy in frequency domain (CF), produces images with low quality due to the inaccurate estimation of motion. In this paper, a new algorithm is proposed for motion estimation based on the registration by maximizing mutual information in spatial domain (MIS). Furthermore, the optimization process is initialized by CF algorithm, so the algorithm is abbreviated as CF-MIS algorithm in this paper. With phantom and in vivo MR imaging, the CF-AIIS algorithm was shown to be of higher accuracy in rotation estimation than CF algorithm. Consequently, the head motion in T1-weighted PROPELLER MRI was better corrected.
引用
收藏
页码:1378 / 1381
页数:4
相关论文
共 11 条
  • [1] ADAPTIVE TECHNIQUE FOR HIGH-DEFINITION MR IMAGING OF MOVING STRUCTURES
    EHMAN, RL
    FELMLEE, JP
    [J]. RADIOLOGY, 1989, 173 (01) : 255 - 263
  • [2] PROPELLER MRI: Clinical testing of a novel technique for quantification and compensation of head motion
    Forbes, KPN
    Pipe, JG
    Bird, CR
    Heiserman, JE
    [J]. JOURNAL OF MAGNETIC RESONANCE IMAGING, 2001, 14 (03) : 215 - 222
  • [3] SELECTION OF A CONVOLUTION FUNCTION FOR FOURIER INVERSION USING GRIDDING
    JACKSON, JI
    MEYER, CH
    NISHIMURA, DG
    MACOVSKI, A
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 1991, 10 (03) : 473 - 478
  • [4] The shoulder: Adaptive motion correction of MR images
    McGee, KP
    Grimm, RC
    Felmlee, JP
    Rydberg, JR
    Riederer, SJ
    Ehman, RL
    [J]. RADIOLOGY, 1997, 205 (02) : 541 - 545
  • [5] FAST SPIRAL CORONARY-ARTERY IMAGING
    MEYER, CH
    HU, BS
    NISHIMURA, DG
    MACOVSKI, A
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1992, 28 (02) : 202 - 213
  • [6] Pipe JG, 1999, MAGNET RESON MED, V42, P963, DOI 10.1002/(SICI)1522-2594(199911)42:5<963::AID-MRM17>3.0.CO
  • [7] 2-L
  • [8] PIPE JG, 2001, P INT SOC MAG RESON, V9, P743
  • [9] Mutual-information-based registration of medical images: A survey
    Pluim, JPW
    Maintz, JBA
    Viergever, MA
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2003, 22 (08) : 986 - 1004
  • [10] Schäffter T, 1999, MAGNET RESON MED, V41, P954, DOI 10.1002/(SICI)1522-2594(199905)41:5<954::AID-MRM15>3.0.CO