Biomechanical Evaluation of S2 Alar-Iliac Screws Effect of Length and Quad-Cortical Purchase as Compared With Iliac Fixation

被引:100
作者
O'Brien, Joseph R. [1 ]
Yu, Warren [1 ]
Kaufman, Brian E. [1 ]
Bucklen, Brandon [2 ]
Salloum, Kanaan [2 ]
Khalil, Saif [2 ]
Gudipally, Manasa [2 ]
机构
[1] George Washington Univ MFA, Washington, DC USA
[2] Globus Med Inc, Valley Forge Business Ctr, Audubon, PA USA
关键词
sacropelvic; lumbopelvic; iliac screw; fixation; pedicle screw; adult deformity; sacroiliac joint; S2 alar-iliac screw; quad-cortical purchase; S2AI; PROFILE PELVIC FIXATION; SACROPELVIC FIXATION; LUMBOSACRAL FIXATION; SPINAL DEFORMITY; COMPLICATIONS; MODEL; S1;
D O I
10.1097/BRS.0b013e31829e17ff
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Study Design. A biomechanical study conducted on cadaveric specimens. Objective. (1) To compare the biomechanical strength of the S2 alar-iliac (S2AI) screw to traditional iliac fixation and (2) to examine the effect of length and trajectory on the S2AI screw. Summary of Background Data. A recent technique to attain spinal fixation distal to S1 pedicle screws is the S2AI screw using either an open or a percutaneous approach with an altered S2 alar screw trajectory to obtain purchase in the ilium. A novel modification of the S2AI screw is placement with bicortical purchase in the ilium (quad-cortical screw). This may allow for a shorter-length screw with equivalent biomechanics. Methods. Seven human cadaveric spines (L2-Pelvis) were fixed at L2 proximally and the pubis distally. Pedicle screws were placed from L3-S1 with S2AI screw lengths of 65-mm, 80-mm, or 90-mm iliac screws. S2AI screws were tested with and without quad-cortical purchase. Each specimen was tested on the 6 degrees of freedom spine simulator. A load control protocol with an unconstrained pure moment of 10 Nm was used in fl exionextension, lateral bending, and axial rotation for a total of 3 load/ unload cycles. The range of motion was normalized to the intact cadaveric spine (100%). Results. All the instrumented constructs significantly reduced range of motion compared with the intact spine. The L3-S1 construct was statistically significantly less stable than all instrumented constructs in fl exion-extension. There was statistically no significant difference between the S2AI screws of all lengths and the iliac screw constructs with offset connectors. Conclusion. S2AI screws are biomechanically as stable as the test constructs using iliac screws in all loading modes. Sixty-fivemillimeter S2AI screws were biomechanically equivalent to 90-mm iliac screws and 80-mm S2AI screws. Quad-cortical purchase did not statistically significantly improve the biomechanical strength of S2AI screws.
引用
收藏
页码:E1250 / E1255
页数:6
相关论文
共 20 条
[1]   The pros and cons to saving the L5-S1 motion segment in a long scoliosis fusion construct [J].
Bridwell, KH ;
Edwards, CC ;
Lenke, LG .
SPINE, 2003, 28 (20) :S234-S242
[2]  
Bridwell KH, 2004, SACROPELVIC FIXATION
[3]   Low Profile Pelvic Fixation Anatomic Parameters for Sacral Alar-Iliac Fixation Versus Traditional Iliac Fixation [J].
Chang, Tai-Li ;
Sponseller, Paul D. ;
Kebaish, Khaled M. ;
Fishman, Elliot K. .
SPINE, 2009, 34 (05) :436-440
[4]   Outcome and complications of long Fusions to the Sacrum in adult spine deformity - Luque-galveston, combined iliac and sacral screws, and sacral fixation [J].
Emami, A ;
Deviren, V ;
Berven, S ;
Smith, JA ;
Hu, SS ;
Bradford, DS .
SPINE, 2002, 27 (07) :776-786
[5]   Test protocols for evaluation of spinal implants [J].
Goel, VK ;
Panjabi, MM ;
Patwardhan, AG ;
Dooris, AP ;
Serhan, H .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2006, 88A :103-109
[6]   Sacropelvic Fixation Techniques and Complications [J].
Kebaish, Khaled M. .
SPINE, 2010, 35 (25) :2245-2251
[7]  
Kim JH, 2010, J KORENA SOC SPINE S, V17, P7
[8]   Pseudarthrosis in adult spinal deformity following multisegmental instrumentation and arthrodesis [J].
Kim, YJ ;
Bridwell, KH ;
Lenke, LG ;
Cho, KJ ;
Enwards, CC ;
Rinella, AS .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2006, 88A (04) :721-728
[9]   A radiological evaluation of the morphometry and safety of S1, S2 and S2-ilium screws in the Asian population using three dimensional computed tomography scan: an analysis of 180 pelvis [J].
Kwan, Mun Keong ;
Jeffry, Amit ;
Chan, Chris Yin Wei ;
Saw, Lim Beng .
SURGICAL AND RADIOLOGIC ANATOMY, 2012, 34 (03) :217-227
[10]   Biomechanical comparison of lumbosacral fixation techniques in a calf spine model [J].
Lebwohl, NH ;
Cunningham, BW ;
Dmitriev, A ;
Shimamoto, N ;
Gooch, L ;
Devlin, V ;
Boachie-Adjei, O ;
Wagner, TA .
SPINE, 2002, 27 (21) :2312-2320