A high-frequency electron paramagnetic resonance spectrometer for multi-dimensional, multi-frequency, and multi-phase pulsed measurements

被引:25
作者
Cho, F. H. [1 ]
Stepanov, V. [2 ]
Takahashi, S. [1 ,2 ]
机构
[1] Univ So Calif, Dept Phys, Los Angeles, CA 90089 USA
[2] Univ So Calif, Dept Chem, Los Angeles, CA 90089 USA
关键词
SPIN-RESONANCE; SPECTROSCOPY; TIME; ECHO;
D O I
10.1063/1.4889873
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We describe instrumentation for a high-frequency electron paramagnetic resonance (EPR) and pulsed electron-electron double resonance (PELDOR) spectroscopy. The instrumentation is operated in the frequency range of 107-120 GHz and 215-240 GHz and in the magnetic field range of 0-12.1 T. The spectrometer consisting of a high-frequency high-power solid-state source, a quasioptical system, a phase-sensitive detection system, a cryogenic-free superconducting magnet, and a He-4 cryostat enables multi-frequency continuous-wave EPR spectroscopy as well as pulsed EPR measurements with a few hundred nanosecond pulses. Here we discuss the details of the design and the pulsed EPR sensitivity of the instrumentation. We also present performance of the instrumentation in unique experiments including PELDOR spectroscopy to probe correlations in an insulating electronic spin system and application of dynamical decoupling techniques to extend spin coherence of electron spins in an insulating solid-state system. (C) 2014 AIP Publishing LLC.
引用
收藏
页数:7
相关论文
共 40 条
  • [1] Barra AL, 2000, CHEM-EUR J, V6, P1608, DOI 10.1002/(SICI)1521-3765(20000502)6:9<1608::AID-CHEM1608>3.3.CO
  • [2] 2-#
  • [3] Quantum oscillations in a molecular magnet
    Bertaina, S.
    Gambarelli, S.
    Mitra, T.
    Tsukerblat, B.
    Mueller, A.
    Barbara, B.
    [J]. NATURE, 2008, 453 (7192) : 203 - U5
  • [4] Blok H., 2005, BRUKER SPIN REPORT, V156, P3
  • [5] Multiple-quantum ESR and distance measurements
    Borbat, PP
    Freed, JH
    [J]. CHEMICAL PHYSICS LETTERS, 1999, 313 (1-2) : 145 - 154
  • [6] Bylander J, 2011, NAT PHYS, V7, P565, DOI [10.1038/NPHYS1994, 10.1038/nphys1994]
  • [7] EFFECTS OF DIFFUSION ON FREE PRECESSION IN NUCLEAR MAGNETIC RESONANCE EXPERIMENTS
    CARR, HY
    PURCELL, EM
    [J]. PHYSICAL REVIEW, 1954, 94 (03): : 630 - 638
  • [8] A kilowatt pulsed 94 GHz electron paramagnetic resonance spectrometer with high concentration sensitivity, high instantaneous bandwidth, and low dead time
    Cruickshank, Paul A. S.
    Bolton, David R.
    Robertson, Duncan A.
    Hunter, Robert I.
    Wylde, Richard J.
    Smith, Graham M.
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 2009, 80 (10)
  • [9] Distance measurements across randomly distributed nitroxide probes from the temperature dependence of the electron spin phase memory time at 240 GHz
    Edwards, Devin T.
    Takahashi, Susumu
    Sherwin, Mark S.
    Han, Songi
    [J]. JOURNAL OF MAGNETIC RESONANCE, 2012, 223 : 198 - 206
  • [10] New technologies in electron spin resonance
    Freed, JH