Superior reproducibility and repeatability in automated quantitative pupillometry compared to standard manual assessment, and quantitative pupillary response parameters present high reliability in critically ill cardiac patients

被引:16
作者
Nyholm, Benjamin [1 ]
Obling, Laust [1 ]
Hassager, Christian [1 ]
Grand, Johannes [1 ]
Moller, Jacob [1 ]
Othman, Marwan [2 ]
Kondziella, Daniel [2 ,3 ]
Kjaergaard, Jesper [1 ]
机构
[1] Copenhagen Univ Hosp, Rigshosp, Dept Cardiol, Copenhagen, Denmark
[2] Copenhagen Univ Hosp, Rigshosp, Dept Neurol, Copenhagen, Denmark
[3] Univ Copenhagen, Fac Hlth & Med Sci, Dept Clin Med, Copenhagen, Denmark
关键词
RESUSCITATION; AGREEMENT;
D O I
10.1371/journal.pone.0272303
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Background Quantitative pupillometry is part of multimodal neuroprognostication of comatose patients after out-of-hospital cardiac arrest (OHCA). However, the reproducibility, repeatability, and reliability of quantitative pupillometry in this setting have not been investigated. Methods In a prospective blinded validation study, we compared manual and quantitative measurements of pupil size. Observer and device variability for all available parameters are expressed as mean difference (bias), limits of agreement (LoA), and reliability expressed as intraclass correlation coefficients (ICC) with a 95% confidence interval. Results Fifty-six unique quadrupled sets of measurement derived from 14 sedated and comatose patients (mean age 7012 years) were included. For manually measured pupil size, inter-observer bias was -0.140.44 mm, LoA of -1.00 to 0.71 mm, and ICC at 0.92 (0.86-0.95). For quantitative pupillometry, we found bias at 0.030.17 mm, LoA of -0.31 to 0.36 mm and ICCs at 0.99. Quantitative pupillometry also yielded lower bias and LoA and higher ICC for intra-observer and inter-device measurements. Correlation between manual and automated pupillometry was better in larger pupils, and quantitative pupillometry had less variability and higher ICC, when assessing small pupils. Further, observers failed to detect 26% of the quantitatively estimated abnormal reactivity with manual assessment. We found ICC >0.91 for all quantitative pupillary response parameters (except for latency with ICC 0.81-0.91). Conclusion Automated quantitative pupillometry has excellent reliability and twice the reproducibility and repeatability than manual pupillometry. This study further presents novel estimates of variability for all quantitative pupillary response parameters with excellent reliability.
引用
收藏
页数:15
相关论文
共 36 条
[1]  
[Anonymous], 2021, NEUROPTICS NPI 200 P
[2]   Early impairment of intracranial conduction time predicts mortality in deeply sedated critically ill patients: a prospective observational pilot study [J].
Azabou, Eric ;
Rohaut, Benjamin ;
Heming, Nicholas ;
Magalhaes, Eric ;
Morizot-Koutlidis, Regine ;
Kandelman, Stanislas ;
Allary, Jeremy ;
Moneger, Guy ;
Polito, Andrea ;
Maxime, Virginie ;
Annane, Djillali ;
Lofaso, Frederic ;
Chretien, Fabrice ;
Mantz, Jean ;
Porcher, Raphael ;
Sharshar, Tarek .
ANNALS OF INTENSIVE CARE, 2017, 7
[3]   Reliability, repeatability and reproducibility: analysis of measurement errors in continuous variables [J].
Bartlett, J. W. ;
Frost, C. .
ULTRASOUND IN OBSTETRICS & GYNECOLOGY, 2008, 31 (04) :466-475
[4]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[5]   Part 4: Advanced Life Support 2015 International Consensus on Cardiopulmonary Resuscitation and Emergency Cardiovascular Care Science With Treatment Recommendations [J].
Callaway, Clifton W. ;
Soar, Jasmeet ;
Aibiki, Mayuki ;
Bottiger, Bernd W. ;
Brooks, Steven C. ;
Deakin, Charles D. ;
Donnino, Michael W. ;
Drajer, Saul ;
Kloeck, Walter ;
Morley, Peter T. ;
Morrison, Laurie J. ;
Neumar, Robert W. ;
Nicholson, Tonia C. ;
Nolan, Jerry P. ;
Okada, Kazuo ;
O'Neil, Brian J. ;
Paiva, Edison F. ;
Parr, Michael J. ;
Wang, Tzong-Luen ;
Witt, Jonathan .
CIRCULATION, 2015, 132 (16) :S84-S145
[6]  
Chen Jeff W, 2011, Surg Neurol Int, V2, P82, DOI 10.4103/2152-7806.82248
[7]   Reliability of standard pupillometry practice in neurocritical care: an observational, double-blinded study [J].
Couret, David ;
Boumaza, Delphine ;
Grisotto, Coline ;
Triglia, Thibaut ;
Pellegrini, Lionel ;
Ocquidant, Philippe ;
Bruder, Nicolas J. ;
Velly, Lionel J. .
CRITICAL CARE, 2016, 20
[8]   When to use agreement versus reliability measures [J].
de Vet, Henrica C. W. ;
Terwee, Caroline B. ;
Knol, Dirk L. ;
Bouter, Lex M. .
JOURNAL OF CLINICAL EPIDEMIOLOGY, 2006, 59 (10) :1033-1039
[9]   Evaluation of the portable infrared pupillometer [J].
Du, R ;
Meeker, M ;
Bacchetti, P ;
Larson, MD ;
Holland, MC ;
Manley, GT .
NEUROSURGERY, 2005, 57 (01) :198-202
[10]   Neurological disorders in liver transplant candidates: Pathophysiology and clinical assessment [J].
Feltracco, Paolo ;
Cagnin, Annachiara ;
Carollo, Cristiana ;
Barbieri, Stefania ;
Ori, Carlo .
TRANSPLANTATION REVIEWS, 2017, 31 (03) :193-206