Developing a novel method to analyse Gafchromic EBT2 films in intensity modulated radiation therapy quality assurance

被引:6
作者
Hu, Yunfei [1 ]
Wang, Yang [1 ]
Fogarty, Gerald [2 ]
Liu, Guilin [2 ]
机构
[1] St Vincents Clin, Dept Radiat Therapy, Genesis Canc Care, Darlinghurst, NSW 2010, Australia
[2] Mater Hosp, Dept Radiat Therapy, Crows Nest, NSW 2065, Australia
关键词
Gafchromic film; Intensity modulated radiation therapy; Dosimetry; Greyscale; Quality assurance; Clinical testing; RADIOCHROMIC FILM; DOSIMETRY; BEAM;
D O I
10.1007/s13246-013-0232-y
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Recently individual intensity modulated radiation therapy quality assurances (IMRT QA) have been more and more performed with Gafchromic (TM) EBT series films processed in red-green-blue (R-G-B) channel due to their extremely high spatial resolution. However, the efficiency of this method is relatively low, as for each box of film, a calibration curve must be established prior to the film being used for measurement. In this study, the authors find a novel method to process the Gafchromic (TM) EBT series, that is, to use the 16-bit greyscale channel to process the exposed film rather than the conventional 48-bit R-G-B channel, which greatly increases the efficiency and even accuracy of the whole IMRT procedure. The main advantage is that when processed in greyscale channel, the Gafchromic (TM) EBT2 films exhibits a linear relationship between the net pixel value and the dose delivered. This linear relationship firstly reduces the error in calibration-curve fitting, and secondly saves the need of establishing a calibration curve for each box of films if it is only to be used for relative measurements. Clinical testing for this novel method was carried out in two radiation therapy centres that involved a total of 743 IMRT cases, and 740 cases passed the 3 mm 3 % gamma analysis criteria. The cases were also tested with small ionization chambers (cc-13) and the results were convincing. Consequently the authors recommend the use of this novel method to improve the accuracy and efficiency of individual IMRT QA procedure using Gafchromic EBT2 films.
引用
收藏
页码:487 / 494
页数:8
相关论文
共 47 条
[31]   Quality assurance of intensity modulated radiation therapy treatment planning using head and neck phantom [J].
Zahra ;
Rehman, Jalil ur ;
Asghar, H. M. Noor ul Huda Khan ;
Ahmad, Nisar ;
Gilani, Zaheer Abbas ;
Nasar, Gulfam ;
Akhter, Malik M. .
JOURNAL OF RADIOTHERAPY IN PRACTICE, 2019, 18 (03) :239-245
[32]   Development and Evaluation of Multipurpose Phantom for Delivery Quality Assurance of Intensity-modulated Radiation Therapy [J].
Jang, Jae-Uk ;
Han, Man-Seok ;
Kim, Ki-Hwan ;
Jeon, Min-Cheol ;
Hwang, Ui-Jung .
JOURNAL OF MAGNETICS, 2022, 27 (04) :430-439
[33]   Developing new extension of GafChromic RTQA2 film to patient quality assurance field using a plan-based calibration method [J].
Peng, Jiayuan ;
Zhang, Zhen ;
Wang, Jiazhou ;
Xie, Jiang ;
Chen, Junchao ;
Hu, Weigang .
PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (19) :7533-7542
[34]   Development of a novel quality assurance system based on rolled-up and rolled-out radiochromic films in volumetric modulated arc therapy [J].
Park, Ji-Yeon ;
Lee, Jeong-Woo ;
Choi, Kyoung-Sik ;
Lee, Jung Seok ;
Kim, You-Hyun ;
Hong, Semie ;
Suh, Tae-Suk .
MEDICAL PHYSICS, 2011, 38 (12) :6688-6696
[35]   Comprehensive Patient-Specific Intensity-Modulated Radiation Therapy Quality Assurance Comparing Mobius3D/FX to Conventional Methods of Evaluation [J].
Basavatia, Amar K. ;
Fiedler, Derek A. ;
Ritter, Johannes ;
Brodin, Patrik ;
Goddard, Lee C. ;
Jeong, Kyoungkeun ;
Hsu, Shu-Hui ;
Mynampati, Dinesh K. ;
Yaparpalvi, Ravindra ;
Tome, Wolfgang A. .
CUREUS JOURNAL OF MEDICAL SCIENCE, 2021, 13 (05)
[36]   Evaluation of a 3D-printed heterogeneous anthropomorphic head and neck phantom for patient-specific quality assurance in intensity-modulated radiation therapy [J].
Kadoya, Noriyuki ;
Abe, Kota ;
Nemoto, Hikaru ;
Sato, Kiyokazu ;
Ieko, Yoshiro ;
Ito, Kengo ;
Dobashi, Suguru ;
Takeda, Ken ;
Jingu, Keiichi .
RADIOLOGICAL PHYSICS AND TECHNOLOGY, 2019, 12 (03) :351-356
[37]   Effect of dose grid resolution on the results of patient-specific quality assurance for intensity-modulated radiation therapy and volumetric modulated arc therapy [J].
Chun, M. ;
Kim, J., I ;
Oh, G. H. ;
Wu, H. G. ;
Park, J. M. .
INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 2020, 18 (03) :521-530
[38]   Radiobiological model-based bio-anatomical quality assurance in intensity-modulated radiation therapy for prostate cancer [J].
Park, Ji-Yeon ;
Lee, Jeong-Woo ;
Chung, Jin-Beom ;
Choi, Kyoung-Sik ;
Kim, Yon-Lae ;
Park, Byung-Moon ;
Kim, Youhyun ;
Kim, Jungmin ;
Choi, Jonghak ;
Kim, Jae-Sung ;
Hong, Semie ;
Suh, Tae-Suk .
JOURNAL OF RADIATION RESEARCH, 2012, 53 (06) :978-988
[39]   Patient-Specific Quality Assurance for the Delivery of 60Co Intensity Modulated Radiation Therapy Subject to a 0.35-T Lateral Magnetic Field [J].
Li, H. Harold ;
Rodriguez, Vivian L. ;
Green, Olga L. ;
Hu, Yanle ;
Kashani, Rojano ;
Wooten, H. Omar ;
Yang, Deshan ;
Mutic, Sasa .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2015, 91 (01) :65-72
[40]   A non-invasive immobilization system and related quality assurance for dynamic intensity modulated radiation therapy of intracranial and head and neck disease [J].
Tsai, JS ;
Engler, MJ ;
Ling, MN ;
Wu, JK ;
Kramer, B ;
Dipetrillo, T ;
Wazer, DE .