Developing a novel method to analyse Gafchromic EBT2 films in intensity modulated radiation therapy quality assurance

被引:6
作者
Hu, Yunfei [1 ]
Wang, Yang [1 ]
Fogarty, Gerald [2 ]
Liu, Guilin [2 ]
机构
[1] St Vincents Clin, Dept Radiat Therapy, Genesis Canc Care, Darlinghurst, NSW 2010, Australia
[2] Mater Hosp, Dept Radiat Therapy, Crows Nest, NSW 2065, Australia
关键词
Gafchromic film; Intensity modulated radiation therapy; Dosimetry; Greyscale; Quality assurance; Clinical testing; RADIOCHROMIC FILM; DOSIMETRY; BEAM;
D O I
10.1007/s13246-013-0232-y
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Recently individual intensity modulated radiation therapy quality assurances (IMRT QA) have been more and more performed with Gafchromic (TM) EBT series films processed in red-green-blue (R-G-B) channel due to their extremely high spatial resolution. However, the efficiency of this method is relatively low, as for each box of film, a calibration curve must be established prior to the film being used for measurement. In this study, the authors find a novel method to process the Gafchromic (TM) EBT series, that is, to use the 16-bit greyscale channel to process the exposed film rather than the conventional 48-bit R-G-B channel, which greatly increases the efficiency and even accuracy of the whole IMRT procedure. The main advantage is that when processed in greyscale channel, the Gafchromic (TM) EBT2 films exhibits a linear relationship between the net pixel value and the dose delivered. This linear relationship firstly reduces the error in calibration-curve fitting, and secondly saves the need of establishing a calibration curve for each box of films if it is only to be used for relative measurements. Clinical testing for this novel method was carried out in two radiation therapy centres that involved a total of 743 IMRT cases, and 740 cases passed the 3 mm 3 % gamma analysis criteria. The cases were also tested with small ionization chambers (cc-13) and the results were convincing. Consequently the authors recommend the use of this novel method to improve the accuracy and efficiency of individual IMRT QA procedure using Gafchromic EBT2 films.
引用
收藏
页码:487 / 494
页数:8
相关论文
共 47 条
  • [21] Comparison of film dosimetry techniques used for quality assurance of intensity modulated radiation therapy
    Ju, Sang Gyu
    Han, Youngyih
    Kum, Oyeon
    Cheong, Kwang-Ho
    Shin, Eun Hyuk
    Shin, Jung Suk
    Kim, Jin Sung
    Ahn, Yong Chan
    MEDICAL PHYSICS, 2010, 37 (06) : 2925 - 2933
  • [22] Design and implementation of an anthropomorphic quality assurance phantom for intensity-modulated radiation therapy for the radiation therapy oncology group
    Molineu, A
    Followill, DS
    Balter, PA
    Hanson, WF
    Gillin, MT
    Huq, MS
    Eisbruch, A
    Ibbott, GS
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 63 (02): : 577 - 583
  • [23] Towards automated quality assurance for intensity modulated radiation therapy: Film densitometry
    Low, DA
    Markman, J
    Dempsey, JF
    Mutic, S
    Klein, EE
    Grigereit, TE
    Purdy, JA
    PROCEEDINGS OF THE 22ND ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-4, 2000, 22 : 184 - 187
  • [24] Improving intensity-modulated radiation therapy quality assurance by adopting statistical process control
    Cui, T.
    Peng, J-X
    Jin, X-L
    Chu, X-G
    Zhang, J.
    Cui, L-Z
    Xing, X-F
    CANCER RADIOTHERAPIE, 2022, 26 (03): : 427 - 432
  • [25] Quality control of intensity modulated radiation therapy with optically stimulated luminescent films
    Idri, K
    Santoro, L
    Charpiot, E
    Herault, J
    Costa, A
    Aillères, N
    Delard, R
    Vaillé, JR
    Fesquet, J
    Dusseau, L
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2004, 51 (06) : 3638 - 3641
  • [26] Patient Specification Quality Assurance for Glioblastoma Multiforme Brain Tumors Treated with Intensity Modulated Radiation Therapy
    Al-Mohammed, H. I.
    INTERNATIONAL JOURNAL OF MEDICAL SCIENCES, 2011, 8 (06): : 461 - 466
  • [27] Quality Assurance of Intensity Modulated Radiation Therapy: Site-Specific Results of Eulji University Hospital
    Kim, Sung Jin
    Lee, Mi Jo
    Youn, Seon Min
    RADIATION ONCOLOGY JOURNAL, 2011, 29 (02): : 99 - 106
  • [28] A novel method for dose distribution registration using fiducial marks made by a megavoltage beam in film dosimetry for intensity-modulated radiation therapy quality assurance
    Nakayama, Shinichi
    Monzen, Hajime
    Oonishi, Yuuichi
    Mizote, Rika
    Iramina, Hiraku
    Kaneshige, Souichirou
    Mizowaki, Takashi
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2015, 31 (04): : 414 - 419
  • [29] Comparison of Error Range of Delivery Quality Assurance by Type of Intensity Modulated Radiation Therapy
    Kim, Jeong Ho
    Yoo, Se Jong
    JOURNAL OF MAGNETICS, 2018, 23 (04) : 587 - 595
  • [30] Dosimetric performance of an enhanced dose range radiographic film for intensity-modulated radiation therapy quality assurance
    Olch, AJ
    MEDICAL PHYSICS, 2002, 29 (09) : 2159 - 2168