Developing a novel method to analyse Gafchromic EBT2 films in intensity modulated radiation therapy quality assurance

被引:6
|
作者
Hu, Yunfei [1 ]
Wang, Yang [1 ]
Fogarty, Gerald [2 ]
Liu, Guilin [2 ]
机构
[1] St Vincents Clin, Dept Radiat Therapy, Genesis Canc Care, Darlinghurst, NSW 2010, Australia
[2] Mater Hosp, Dept Radiat Therapy, Crows Nest, NSW 2065, Australia
关键词
Gafchromic film; Intensity modulated radiation therapy; Dosimetry; Greyscale; Quality assurance; Clinical testing; RADIOCHROMIC FILM; DOSIMETRY; BEAM;
D O I
10.1007/s13246-013-0232-y
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Recently individual intensity modulated radiation therapy quality assurances (IMRT QA) have been more and more performed with Gafchromic (TM) EBT series films processed in red-green-blue (R-G-B) channel due to their extremely high spatial resolution. However, the efficiency of this method is relatively low, as for each box of film, a calibration curve must be established prior to the film being used for measurement. In this study, the authors find a novel method to process the Gafchromic (TM) EBT series, that is, to use the 16-bit greyscale channel to process the exposed film rather than the conventional 48-bit R-G-B channel, which greatly increases the efficiency and even accuracy of the whole IMRT procedure. The main advantage is that when processed in greyscale channel, the Gafchromic (TM) EBT2 films exhibits a linear relationship between the net pixel value and the dose delivered. This linear relationship firstly reduces the error in calibration-curve fitting, and secondly saves the need of establishing a calibration curve for each box of films if it is only to be used for relative measurements. Clinical testing for this novel method was carried out in two radiation therapy centres that involved a total of 743 IMRT cases, and 740 cases passed the 3 mm 3 % gamma analysis criteria. The cases were also tested with small ionization chambers (cc-13) and the results were convincing. Consequently the authors recommend the use of this novel method to improve the accuracy and efficiency of individual IMRT QA procedure using Gafchromic EBT2 films.
引用
收藏
页码:487 / 494
页数:8
相关论文
共 48 条
  • [1] Developing a novel method to analyse Gafchromic EBT2 films in intensity modulated radiation therapy quality assurance
    Yunfei Hu
    Yang Wang
    Gerald Fogarty
    Guilin Liu
    Australasian Physical & Engineering Sciences in Medicine, 2013, 36 : 487 - 494
  • [2] Evaluation of a Gafchromic EBT2 film dosimetry system for radiotherapy quality assurance
    Aland, T.
    Kairn, T.
    Kenny, J.
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2011, 34 (02) : 251 - 260
  • [3] Application of Gafchromic EBT2 film for intraoperative radiation therapy quality assurance
    Robatjazi, Mostafa
    Mahdavi, Seied Rabi
    Takavr, Abbas
    Baghani, Hamid Reza
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2015, 31 (03): : 314 - 319
  • [4] Evaluation of a Gafchromic EBT2 film dosimetry system for radiotherapy quality assurance
    T. Aland
    T. Kairn
    J. Kenny
    Australasian Physical & Engineering Sciences in Medicine, 2011, 34 : 251 - 260
  • [5] Dosimetric verification and quality assurance for intensity-modulated radiation therapy using Gafchromic (R) EBT3 film
    Iqbal, Khalid
    Iqbal, Muhammad Mazhar
    Akram, Muhammad
    Altaf, Saima
    Buzdar, Saeed Ahmad
    JOURNAL OF RADIOTHERAPY IN PRACTICE, 2018, 17 (01) : 85 - 95
  • [6] EBT GAFCHROMIC™ film dosimetry in compensator-based intensity modulated radiation therapy
    Vaezzadeh, Seyedali
    Allahverdi, Mahmoud
    Nedaie, Hasan A.
    Ay, Mohammadreza
    Shirazi, Alireza
    Yarahmadi, Mehran
    MEDICAL DOSIMETRY, 2013, 38 (02) : 176 - 183
  • [7] Comparison of Gafchromic EBT2 and EBT3 for patient-specific quality assurance: Cranial stereotactic radiosurgery using volumetric modulated arc therapy with multiple noncoplanar arcs
    Fiandra, Christian
    Fusella, Marco
    Giglioli, Francesca Romana
    Filippi, Andrea Riccardo
    Mantovani, Cristina
    Ricardi, Umberto
    Ragona, Riccardo
    MEDICAL PHYSICS, 2013, 40 (08)
  • [8] Quality assurance of intensity-modulated radiation therapy
    Palta, Jatinder R.
    Li, Chihray
    Li, Jonathan G.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 71 (01): : S108 - S112
  • [9] Evaluation of gafchromic EBT film for intensity modulated radiation therapy dose distribution verification
    Sankar, A.
    Kurup, P. G. Goplakrishna
    Murali, V.
    Ayyangar, Komanduri M.
    Nehru, R. Mothilal
    Velmurugan, J.
    JOURNAL OF MEDICAL PHYSICS, 2006, 31 (02) : 78 - 82
  • [10] Gafchromic EBT2 film dosimetry in reflection mode with a novel plan-based calibration method
    Mendez, I.
    Hartman, V.
    Hudej, R.
    Strojnik, A.
    Casar, B.
    MEDICAL PHYSICS, 2013, 40 (01)