A randomized controlled trial of efficacy and ST change following use of the Welch-Allyn MRL PIC biphasic waveform versus damped sine monophasic waveform for external DC cardioversion

被引:21
作者
Ambler, Jonathan J. S. [1 ]
Deakin, Charles D. [1 ]
机构
[1] Southampton Univ Hosp, NHS Trust, Shackleton Dept Anaesthet, Southampton SO16 6YD, Hants, England
关键词
arrhythmia; defibrillation; electric countershock; electrocardiography;
D O I
10.1016/j.resuscitation.2006.03.017
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective: Biphasic waveforms have similar or greater efficacy at cardioverting atrial. and ventricular arrhythmias at tower energy levels than monophasic waveforms, and cause less ST depression following defibrillation of ventricular fibrillation. No studies have investigated this effect on ST change with atria[ arrhythmias. We studied the efficacy of the Welch Allyn-MRL PIC biphasic defibrillator. Methods: One hundred and thirty-nine patients undergoing elective DC cardioversion for atrial. arrhythmias were randomised to cardioversion by monophasic (Hewlett Packard Codemaster XL; 100, 200, 300, 360 and 360 J) or biphasic (Welch Allyn-MRL PIC; 70, 100, 150, 200 and 300 J) defibrillator. We analysed success of cardioversion after 0 and 30 min, cumulative energy, number of shocks and energy at successful cardioversion. The ST change in the recorded electrocardiogram was measured at 15 s after all shocks using electronic callipers. Results: Immediately after cardioversion 59/68 (86.8%) of the monophasic group versus 56/60 (93.3%) of the biphasic group were in sinus rhythm. Of the monophasic group, 55/67 (82.1%) remained in sinus rhythm at 30min versus 53/58 (91.4%) of the biphasic group. These differences were not significant at 0 min (P=0.35) or 30min (P=0.21). The biphasic group required significantly fewer shocks (P=0.006), less cumulative energy (P < 0.0001) and required tower total energy for successful cardioversion (P < 0.0001). Of the 102 patients with electrocardiogram recordings suitable for analysis, ST segment change was greater in the monophasic group (P=0.037). Conclusions: The Welch Allyn-MRL biphasic waveform for DC cardioversion results in fewer shocks, with Less cumulative energy delivered and less post shock ST change than with a Hewlett Packard Codemaster XL damped sine wave monophasic waveform. (c) 2006 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:146 / 151
页数:6
相关论文
共 12 条
[1]   Multicenter comparison of truncated biphasic shocks and standard damped sine wave monophasic shocks for transthoracic ventricular defibrillation [J].
Bardy, GH ;
Marchlinski, FE ;
Sharma, AD ;
Worley, SJ ;
Luceri, RM ;
Yee, R ;
Halperin, BD ;
Fellows, CL ;
Ahern, TS ;
Chilson, DA ;
Packer, DL ;
Wilber, DJ ;
Mattioni, TA ;
Reddy, R ;
Kronmal, RA ;
Lazzara, R .
CIRCULATION, 1996, 94 (10) :2507-2514
[2]   Determining the optimal paddle fore for external defibrillation [J].
Deakin, CD ;
Sado, DM ;
Petley, GW ;
Clewlow, F .
AMERICAN JOURNAL OF CARDIOLOGY, 2002, 90 (07) :812-+
[3]  
DORIAN P, 2001, ACAD EMERG MED, V8, P543
[4]  
Higgins S L, 2000, Prehosp Emerg Care, V4, P305, DOI 10.1080/10903120090941001
[5]  
*ILCOR, 2000, RESUSCITATION, V46, P111
[6]   A randomized trial comparing monophasic and biphasic waveform shocks for external cardioversion of atrial fibrillation [J].
Koster, RW ;
Dorian, P ;
Chapman, FW ;
Schmitt, PW ;
O'Grady, SG ;
Walker, RG .
AMERICAN HEART JOURNAL, 2004, 147 (05) :e1-e7
[7]   Comparison of a novel rectilinear biphasic waveform with a damped sine wave monophasic waveform for transthoracic ventricular defibrillation [J].
Mittal, S ;
Ayati, S ;
Stein, KM ;
Knight, BP ;
Morady, F ;
Schwartzman, D ;
Cavlovich, D ;
Platia, EV ;
Calkins, H ;
Tchou, PJ ;
Miller, JM ;
Wharton, JM ;
Sung, RJ ;
Slotwiner, DJ ;
Markowitz, SM ;
Lerman, BB .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1999, 34 (05) :1595-1601
[8]   Transthoracic cardioversion of atrial fibrillation - Comparison of rectilinear biphasic versus damped sine wave monophasic shocks [J].
Mittal, S ;
Ayati, S ;
Stein, KM ;
Schwartzman, D ;
Cavlovich, D ;
Tchou, PJ ;
Markowitz, SM ;
Slotwiner, DJ ;
Scheiner, MA ;
Lerman, BB .
CIRCULATION, 2000, 101 (11) :1282-1287
[9]   Biphasic transthoracic defibrillation causes fewer ECG ST-segmnent changes after shock [J].
Reddy, RK ;
Gleva, MJ ;
Gliner, BE ;
Dolack, GL ;
Kudenchuk, PJ ;
Poole, JE ;
Bardy, GH .
ANNALS OF EMERGENCY MEDICINE, 1997, 30 (02) :127-134
[10]   External cardioversion of atrial fibrillation:: comparison of biphasic vs monophasic waveform shocks [J].
Ricard, P ;
Lévy, S ;
Boccara, G ;
Lakhal, E ;
Bardy, G .
EUROPACE, 2001, 3 (02) :96-99