Relationship between the image quality and noise-equivalent count in time-of-flight positron emission tomography

被引:4
作者
Maebatake, Akira [1 ]
Akamatsu, Go [1 ,2 ]
Miwa, Kenta [1 ]
Tsutsui, Yuji [3 ]
Himuro, Kazuhiko [3 ]
Baba, Shingo [4 ]
Sasaki, Masayuki [1 ]
机构
[1] Kyushu Univ, Grad Sch Med Sci, Dept Hlth Sci, Div Med Quantum Sci,Higashi Ku, 3-1-1 Maidashi, Fukuoka 8128582, Japan
[2] Inst Biomed Res & Innovat, Div Mol Imaging, Kobe, Hyogo, Japan
[3] Kyushu Univ Hosp, Dept Med Technol, Div Radiol, Fukuoka 812, Japan
[4] Kyushu Univ, Grad Sch Med Sci, Dept Clin Radiol, Fukuoka 812, Japan
关键词
PET/CT; Noise-equivalent count; Time-of-flight; Point spread function; Image quality; LUNG-CANCER; PET; RECONSTRUCTION; IMPROVEMENT; DEPENDENCE; SCANNER; BENEFIT; IMPACT; RATIO; PSF;
D O I
10.1007/s12149-015-1032-0
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective The aim of this study was to investigate the relationship between the NEC and TOF-PET image quality. Methods The National Electrical Manufactures Association and International Electrical Commission (NEMA IEC) body phantom with a 10-mm diameter sphere was filled with an F-18-FDG solution with a 4:1 radioactivity ratio. The PET data were acquired in the three-dimensional list mode for 20 min. We created frame data ranging from 1 to 5 min acquisition time, which were then reconstructed using the baseline ordered-subsets expectation maximization (OSEM), the OSEM ? point spread function (PSF) algorithm, OSEM ? time-of-flight (TOF) algorithm and OSEM ? PSF ? TOF algorithm. The PET images were analyzed according to the noise-equivalent count (NEC), the coefficients of variance of the background (CVBG), the maximum count (CVmax) and the contrast (CVCONT). The results were compared with the recommended value according to the guidelines for the oncology FDG-PET/CT protocol. Results The NEC was higher than the recommended value at 3 min or longer acquisition time. The CVBG lower than 15 % were obtained at 3 min acquisition time without TOF and at 2 min acquisition time with TOF. The CVBG of 10 % or lower were obtained at 5 min or longer acquisition time without TOF and at 4 min or longer acquisition time with TOF. Both the CVmax and CVCONT lower than 10 % were obtained at 3 min or longer acquisition time without TOF and at 1 min acquisition or longer with TOF. No particular relationships were observed between the frame number and degree of the variation in the image quality. The CVCONT significantly correlated with the NEC for the data reconstructed without TOF information, while there were no significant correlations between these useful metrics for the data reconstructed with TOF. Conclusion This study demonstrated that the NEC is not a useful metric for the evaluation of the image quality on TOF-PET images.
引用
收藏
页码:68 / 74
页数:7
相关论文
共 28 条
[1]   Improvement in PET/CT Image Quality with a Combination of Point-Spread Function and Time-of-Flight in Relation to Reconstruction Parameters [J].
Akamatsu, Go ;
Ishikawa, Kaori ;
Mitsumoto, Katsuhiko ;
Taniguchi, Takafumi ;
Ohya, Nobuyoshi ;
Baba, Shingo ;
Abe, Koichiro ;
Sasaki, Masayuki .
JOURNAL OF NUCLEAR MEDICINE, 2012, 53 (11) :1716-1722
[2]  
Beyer T, 2000, J NUCL MED, V41, P1369
[3]  
Boellaard R, 2004, J NUCL MED, V45, P1519
[4]   Investigation of 18F-FDG 3D mode PET image quality versus acquisition time [J].
Brown, Colin ;
Dempsey, Mary-Frances ;
Gillen, Gerry ;
Elliott, Alex T. .
NUCLEAR MEDICINE COMMUNICATIONS, 2010, 31 (03) :254-259
[5]  
BUDINGER TF, 1983, J NUCL MED, V24, P73
[6]   Effect of randoms on signal-to-noise ratio in TOF PET [J].
Conti, Maurizio .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2006, 53 (03) :1188-1193
[7]   Reconstruction of scattered and unscattered PET coincidences using TOF and energy information [J].
Conti, Maurizio ;
Hong, Inki ;
Michel, Christian .
PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (15) :N307-N317
[8]   Comparison of noise equivalent count rates and image noise [J].
Dahlbom, M ;
Schiepers, C ;
Czernin, J .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2005, 52 (05) :1386-1390
[9]   Preoperative Staging of Lung Cancer with Combined PET-CT. [J].
Fischer, Barbara ;
Lassen, Ulrik ;
Mortensen, Jann ;
Larsen, Soren ;
Loft, Annika ;
Bertelsen, Anne ;
Ravn, Jesper ;
Clementsen, Paul ;
Hogholm, Asbjorn ;
Larsen, Klaus ;
Rasmussen, Torben ;
Keiding, Susanne ;
Dirksen, Asger ;
Gerke, Oke ;
Skov, Birgit ;
Steffensen, Ida ;
Hansen, Hanne ;
Vilmann, Peter ;
Jacobsen, Grete ;
Backer, Vibeke ;
Maltbaek, Niels ;
Pedersen, Jesper ;
Madsen, Henrik ;
Nielsen, Henrik ;
Hojgaard, Liselotte .
NEW ENGLAND JOURNAL OF MEDICINE, 2009, 361 (01) :32-39
[10]   Japanese guideline for the oncology FDG-PET/CT data acquisition protocol: synopsis of Version 1.0 [J].
Fukukita, Hiroyoshi ;
Senda, Michio ;
Terauchi, Takashi ;
Suzuki, Kazufumi ;
Daisaki, Hiromitsu ;
Matsumoto, Keiichi ;
Ikari, Yasuhiko ;
Hayashi, Masuo .
ANNALS OF NUCLEAR MEDICINE, 2010, 24 (04) :325-334