Assuring precision and calibration linearity in ultrasensitive suppressed ion chromatography by using deliberately contaminated hydroxide eluent

被引:1
作者
Eom, Jiwon [1 ]
Lee, Gahyeon [1 ]
Lee, Dong Soo [1 ]
机构
[1] Yonsei Univ, Dept Chem, Seoul 120749, South Korea
关键词
Nonlinear response; Suppressed ion chromatography; Eluent impurity analysis; Eluent contamination; Calibration sensitivity; CONDUCTIVITY DETECTION; NATURAL-WATERS; AMINES;
D O I
10.1016/j.chroma.2018.05.009
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Experimental: and theoretical studies were conducted to investigate low and non-linear responses in sub-micro molar-level suppressed ion chromatography with a hydroxide eluent. A calculated response was derived using experimentally determined detector effluent ion composition data and compared with measured experimental responses. The calibration curve was non-linear, and its slope varied considerably with different instrumental setups and operating conditions. The non-linearity of the solution conductivity response was determined by two acid-base equilibria of water and carbonic acid, and fluoride ion. By using eluent contaminated with para-toluene sulfonate at micro molar level, the non-linear response was greatly alleviated. (C) 2018 Published by Elsevier B.V.
引用
收藏
页码:90 / 98
页数:9
相关论文
共 24 条
[2]   Two-dimensional detection in ion chromatography: Sequential conductometry after suppression and passive hydroxide introduction [J].
Al-Horr, R ;
Dasgupta, PK ;
Adams, RL .
ANALYTICAL CHEMISTRY, 2001, 73 (19) :4694-4703
[3]  
[Anonymous], CURR SEP
[4]   Non-linear calibration functions in ion chromatography with suppressed conductivity detection using hydroxide eluents [J].
Brinkmann, T ;
Specht, CH ;
Frimmel, FH .
JOURNAL OF CHROMATOGRAPHY A, 2002, 957 (02) :99-109
[5]   Development of an automated method for simultaneous determination of low molecular weight aliphatic amines and ammonia in ambient air by diffusion scrubber coupled to ion chromatography [J].
Chang, IH ;
Lee, CG ;
Lee, DS .
ANALYTICAL CHEMISTRY, 2003, 75 (22) :6141-6146
[6]   THEORETICAL APPROACH OF DUAL-COLUMN ION CHROMATOGRAPHY [J].
DOURYBERTHOD, M ;
GIAMPAOLI, P ;
PITSCH, H ;
SELLA, C ;
POITRENAUD, C .
ANALYTICAL CHEMISTRY, 1985, 57 (12) :2257-2263
[7]   Low-temperature Conductivity Detection for Ultrasensitive Ion Chromatography [J].
Eom, Jiwon ;
Lee, Dong Soo .
BULLETIN OF THE KOREAN CHEMICAL SOCIETY, 2016, 37 (03) :321-329
[8]   Lower Limits and Nonlinearity of Calibration in Suppressed Ion Chromatography with Strong Acid Eluent [J].
Eom, Jiwon ;
Kim, Mi-Young ;
Lee, Bokyoung ;
Lee, Dong Soo .
BULLETIN OF THE KOREAN CHEMICAL SOCIETY, 2015, 36 (08) :2073-2080
[9]  
Ishhimoto T., 2010, MEMBRANES, V2, P395
[10]  
Lide D.R., 2018, CRC Handbook of Chemistry and Physics, V99th