Evaluation of the XE-5000 for the automated analysis of blood cells in cerebrospinal fluid

被引:51
作者
Boer, Klas [1 ]
Deufel, Thomas [1 ]
Reinhoefer, Mike [1 ]
机构
[1] Inst Klin Chem & Lab Diagnost, D-07740 Jena, Germany
关键词
CSF; Leukocyte; Cell counting; XE-5000; Evaluation; Counting chamber; Body fluid; Cerebrospinal fluid; FLOW-CYTOMETRIC ANALYSIS; LEUKOCYTES; SAMPLES;
D O I
10.1016/j.clinbiochem.2009.01.025
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Objectives: Counting of cells in cerebrospinal fluid is an important clinical laboratory test and elevated white blood cell counts in cerebrospinal fluid are frequently seen in CNS disorders. Quantification of red blood cell concentrations in CSF may help to interpret certain diagnostic constellations and may result from subarachnoid haemorrhage, surgical procedures or contamination due to traumatic puncture. Table top analyser XE-5000 (Sysmex, Norderstedt, Germany) offers, beside its use as a haematology analyser, a protocol for the quantification of red and white blood cells in body fluids such as CSF including the differentiation between polymorphonuclear and mononuclear cells. A detection limit of 1 cell/mm(3) Would render this device suitable for automated CSF analysis. Design and methods: White blood cell counting was compared between Fuchs-Rosenthal counting chamber and XE-5000 in 273 routinely collected lumbar and ventricular CSF samples. Red blood cell Counting was compared between UF-100 and XE-5000. Differentiation was performed on a slide stained after Pappenheim and compared to the differential count of the XE-5000. Results: Linearity was established between 1 and 10,000 cells/mm(3) for white blood cells and between 1000 and 1 * 10(3) particles/mm(3) for red blood cells. Functional sensitivity was established at 20 cells/mm(3) for white blood cell counting and at 1000 particles/mm(3) (lowest reported concentration) for red blood cell counting. When comparing between microscopic and automatic white blood cell counts no statistically significant slope and offset were detected in lumbar CSF samples while a significant slope and offset were detected when comparing ventricular CSF samples. Most patients were classified correctly according to their WBC count (non-pathologic, mildly, moderately, and highly elevated) by both methods although more patients had pathologic white blood cell counts on XE-5000. A significant slope and offset were detected when comparing red blood cell counts between UF-100 and XE-5000. Conclusions: In summary despite its high imprecision at low white blood cell counts (<20 particles/mm(3)) most patients were classified correctly and therefore XE-5000 is suitable for automated quantification of white blood cells in cerebrospinal fluid in a defined diagnostic setting. This could significantly improve automation in the relatively time- and manual work-intensive field of cerebrospinal fluid diagnostics. However, careful review of plausibility of the results continues to be compulsory. (C) 2009 The Canadian Society of Clinical Chemists. Published by Elsevier Inc. All rights reserved.
引用
收藏
页码:684 / 691
页数:8
相关论文
共 17 条
[1]  
Andrews J, 2005, Lab Hematol, V11, P98, DOI 10.1532/LH96.04071
[2]  
Aulesa C, 2003, Lab Hematol, V9, P214
[3]   Automated flow cytometric analysis of blood cells in cerebrospinal fluid - Analytic performance [J].
Aune, MW ;
Becker, JL ;
Brugnara, C ;
Canfield, W ;
Dorfman, DM ;
Fiehn, W ;
Fischer, G ;
Fitzpatrick, P ;
Flaming, TH ;
Henriksen, HK ;
Kunicka, JE ;
Lackner, KJ ;
Minchello, E ;
Mullenix, PA ;
Myers, M ;
Petersen, A ;
Ternstrom, W ;
Wilson, S .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2004, 121 (05) :690-700
[4]  
BARNES PW, 2004, LAB HEMATOL, V10, P27
[5]   Correction of ventricular cerebrospinal fluid (CSF) samples for blood content does not increase sensitivity and specificity for the detection of CSF infection [J].
Boeer, Klas ;
Siegmund, Robert ;
Pfister, Wolfgang ;
Isenmann, Stefan ;
Deufel, Thomas .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2008, 46 (06) :842-848
[6]   Corrections for leukocytes and percent of neutrophils do not match observations in blood-contaminated cerebrospinal fluid and have no value over uncorrected cells for diagnosis [J].
Bonsu, BK ;
Harper, MB .
PEDIATRIC INFECTIOUS DISEASE JOURNAL, 2006, 25 (01) :8-11
[7]   A multicenter evaluation of the Iris iQ200 automated urine microscopy analyzer Body Fluids Module and comparison with hemacytometer cell counts [J].
Butch, Anthony W. ;
Wises, Porntip K. ;
Wah, David T. ;
Gornet, Terrie G. ;
Fritsche, Herbert A. .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2008, 129 (03) :445-450
[8]   Characterization of chemical meningitis after neurological surgery [J].
Forgacs, P ;
Geyer, CA ;
Freidberg, SR .
CLINICAL INFECTIOUS DISEASES, 2001, 32 (02) :179-185
[9]   The Clinical Reliability of Automated Cerebrospinal Fluid Cell Counts on the Beckman-Coulter LH750 and Iris iQ200 [J].
Glasser, Lewis ;
Murphy, Carole A. ;
Machan, Jason T. .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2009, 131 (01) :58-63
[10]  
Heller T, 2008, ANAL QUANT CYTOL, V30, P139