Technical Tricks When Using the Reamer Irrigator Aspirator Technique for Autologous Bone Graft Harvesting

被引:39
作者
Quintero, Andres J. [2 ]
Tarkin, Ivan S. [2 ]
Pape, Hans-Christoph [1 ]
机构
[1] Univ Aachen, Med Ctr, Dept Orthopaed Trauma Surg, D-52074 Aachen, Germany
[2] Univ Pittsburgh, Med Ctr, Pittsburgh, PA USA
关键词
autologous bone grafting; reamer irrigator aspirator; complications; non-union; intraoperative hypotension; anterior cortical defect; ORTHOPEDIC TRAUMA; NON-UNIONS; FEMUR; OSTEOMYELITIS; SUBSTITUTES; PRESSURE; SYSTEM;
D O I
10.1097/BOT.0b013e3181b2b4eb
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
This report describes technical tricks for using the reamer irrigator aspirator to harvest autologous bone graft from the femur. This device is a focus of interest in orthopaedics because it can be used to harvest bone graft from the femoral canal and medial condyle in voluminous quantities. Moreover, according to some authors, the osteogenic potential of this graft is at least as effective as that of autogenous bone obtained from the iliac crest. The reamer irrigator aspirator device has substantially different design characteristics and technicalities compared with those of a standard reamer. First, a guidewire must be redirected into multiple areas, including the center of the distal femur and into both condyles, to harvest ample bone graft. This is accomplished by prebending the guidewire in a stronger fashion than required for regular reaming in the case of femoral nailing procedures. This bend can increase the risk for eccentric reaming as well as lodging of the suctioning device within the femoral canal. Second, the front and lateral drilling surfaces of this device are very sharp and further cleaned and maintained sharp by the irrigation process to pen-nit the surgeon to obtain significant volumes of graft with a single passage of this device. At the same time, however, this sharp front-end cutting design can increase the risk of iatrogenic fracture if reaming is performed without caution. Third, a powerful suctioning device is connected to the reamer such that the blood loss that can occur during continuous reaming, irrigation, and aspiration must be considered with this technique. We hereby discuss these potential dangers and describe the correct use of this device with technical tricks to minimize the risk of unexpected intraoperative events.
引用
收藏
页码:42 / 45
页数:4
相关论文
共 19 条
[1]   Use of the reamer irrigator aspirator for the treatment of a 20-year recurrent osteomyelitis of a healed femur fracture [J].
Bellapianta, Joseph ;
Gerdeman, Andrew ;
Sharan, Alok ;
Lozman, Jeffrey .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2007, 21 (05) :343-346
[2]   Bone Graft Harvest Using a New Intramedullary System [J].
Belthur, Mohan V. ;
Conway, Janet D. ;
Jindal, Gaurav ;
Ranade, Ashish ;
Herzenberg, John E. .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2008, 466 (12) :2973-2980
[3]   Bone, grafts and bone graft substitutes in orthopedic trauma surgery - A critical analysis [J].
De Long, William G., Jr. ;
Einhorn, Thomas A. ;
Koval, Kenneth ;
McKee, Michael ;
Smith, Wade ;
Sanders, Roy ;
Watson, Tracy .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2007, 89A (03) :649-658
[4]   Bone substitutes: An update [J].
Giannoudis, PV ;
Dinopoulos, H ;
Tsiridis, E .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2005, 36 :20-27
[5]   Effect of re-implanted particles from intramedullary reaming on mechanical properties and callus formation [J].
Hammer, T. O. ;
Wieling, R. ;
Green, J. M. ;
Suedkamp, N. P. ;
Schneider, E. ;
Mueller, C. A. .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2007, 89B (11) :1534-1538
[6]   The influence of a one-step reamer-irrigator-aspirator technique on the intramedullary pressure in the pig femur [J].
Husebye, Elisabeth Ellingsen ;
Lyberg, Torstein ;
Madsen, Jan Erik ;
Eriksen, Morten ;
Roise, Olav .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2006, 37 (10) :935-940
[7]   Biological enhancement of tibial diaphyseal aseptic non-unions: the efficacy of autologous bone grafting, BMPs and reaming by-products [J].
Kanakaris, Nikolaos K. ;
Paliobeis, Christos ;
Manidakis, Nikolaos ;
Giannoudis, Peter V. .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2007, 38 :S65-S75
[8]   Substitutes for autologous bone graft in orthopaedic trauma [J].
Keating, JF ;
McQueen, MM .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2001, 83B (01) :3-8
[9]  
Kobbe P, 2008, UNFALLCHIRURG, V111, P469, DOI 10.1007/s00113-007-1359-7
[10]   Use of the 'reamer irrigator aspirator' system for non-infected tibial non-union after failed iliac crest grafting [J].
Kobbe, Philipp ;
Tarkin, Ivan S. ;
Pape, Hans Christoph .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2008, 39 (07) :796-800