Evaluation of the point-of-care devices KetoSureTM and StatStrip Express(R) blood ketone tests using β-hydroxybutyrate spiked samples

被引:4
作者
Bjerg, Lise Norkjaer [1 ]
Thomsen, Henrik Holm [2 ,3 ]
Madsen, Jeppe Buur [4 ]
Sandfeld-Paulsen, Birgitte [1 ]
机构
[1] Reg Hosp Viborg, Dept Clin Biochem, Viborg, Denmark
[2] Reg Hosp Viborg, Dept Internal Med, Viborg, Denmark
[3] Reg Hosp Viborg, Res Unit Multimorbid, Viborg, Denmark
[4] Univ Hosp Southern Denmark, Lillebaelt Hosp, Dept Biochem & Immunol, Vejle, Denmark
关键词
Diabetic ketoacidosis; point-of-care systems; blood; ketone bodies; 3-hydroxybutyric acid; metabolic diseases; DIABETIC-KETOACIDOSIS; ACETOACETATE; MANAGEMENT; DIAGNOSIS; UTILITY;
D O I
10.1080/00365513.2022.2054027
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Measurement of beta-hydroxybutyrate (BHB) in blood is used clinically as a measure of ketosis when diabetic ketoacidosis is expected. With the introduction of point-of-care testing (POCT) for blood-BHB, nonproductive time is reduced to a minimum in a potential critical situation; however, studies have observed inferior quality of POCT-BHB. Recently, the POCT device KetoSure (Roche) has been introduced to the clinic. In this study, we evaluated the imprecision and linearity of KetoSure and compared this to the established StatStrip Express (Nova Biomedical) based on spiked full blood samples. We found comparable imprecision for KetoSure and StatStrip Express. However, linearity was only observed in the lower part of the measuring range for both devices. In a method comparison, higher values of BHB were measured by KetoSure than by an enzymatic endpoint spectrophotometric reference method (mean bias: 21% (95% confidence interval (CI): 4%-37%)). Conversely, StatStrip Express returned lower values of BHB (mean bias: -16% (95% CI: -38%-7%), while the widely applied POCT device FreeStyle Precision Neo (Abbott) returned values equivalent with the reference method (mean bias: 5% (95% CI: -14%-24%). In samples with concentrations of BHB above the measuring range, the POCT devices could be provoked to return falsely low results. In conclusion, the quality of KetoSure is in line with other established POCT devices; however, the KetoSure measures higher concentrations than other POCT devices. As, linearity only was observed in the lower part of the measuring range and as falsely low measures could be provoked, we advise users to interpret results with precaution.
引用
收藏
页码:226 / 231
页数:6
相关论文
共 23 条
[1]   Limitations to using point of care blood ketone testing to monitor DKA treatment [J].
Armer, Jane ;
Hunt, Natalie ;
Kaushal, Kalpana ;
Myers, Martin .
PRACTICAL DIABETES, 2013, 30 (09) :380-382
[2]  
Arora Sanjay, 2012, Hosp Pract (1995), V40, P73, DOI 10.3810/hp.2012.04.972
[3]   Diagnostic Accuracy of Point-of-Care Testing for Diabetic Ketoacidosis at Emergency-Department Triage β-Hydroxybutyrate versus the urine dipstick [J].
Arora, Sanjay ;
Henderson, Sean O. ;
Long, Theodore ;
Menchine, Michael .
DIABETES CARE, 2011, 34 (04) :852-854
[4]  
Ceriotti Ferruccio, 2015, J Diabetes Sci Technol, V9, P268, DOI 10.1177/1932296814563351
[5]  
CSLI, 2013, EP09A3 CSLI
[6]  
Dhatariya Ketan, 2016, Rev Diabet Stud, V13, P217, DOI 10.1900/RDS.2016.13.217
[7]   Multicentric evaluation of eight glucose and four ketone blood meters [J].
Guimont, Marie-Christine ;
Desjobert, Helene ;
Fonfrede, Michele ;
Vitoux, Dominique ;
Benoist, Jean-Francois ;
Launay, Jean-Marie ;
Peoc'h, Katell ;
Lefevre, Guillaume .
CLINICAL BIOCHEMISTRY, 2015, 48 (18) :1310-1316
[8]   Near patient blood ketone measurements and their utility in predicting diabetic ketoacidosis [J].
Harris, S ;
Ng, R ;
Syed, H ;
Hillson, R .
DIABETIC MEDICINE, 2005, 22 (02) :221-224
[9]   Evaluation of Nova StatStrip and FreeStyle Precision Pro blood ketone tests using 3-hydroxybutyrate doped samples [J].
Helmersson-Karlqvist, Johanna ;
Hammarstrom, Kristina Hoog ;
Palmberg, Kicki ;
Backman-Johansson, Carolina .
JOURNAL OF CLINICAL LABORATORY ANALYSIS, 2019, 33 (04)
[10]   Accuracy of the Precision® point-of-care ketone test examined by liquid chromatography tandem-mass spectrometry (LC-MS/MS) in the same fingerstick sample [J].
Janssen, Marcel J. W. ;
Hendrickx, Ben H. E. ;
Habets-van der Poel, Carin D. ;
van den Bergh, Joop P. W. ;
Haagen, Anton A. M. ;
Bakker, Jaap A. .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2010, 48 (12) :1781-1784