3-D flow velocity vector estimation with a triple-beam lens transducer - Experimental results

被引:23
作者
Hein, IA
机构
[1] Bio-Imaging Research, Inc., 425 Barclay Blvd, Lincolnshire
基金
美国国家卫生研究院;
关键词
D O I
10.1109/58.585197
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Current commercial ultrasound blood flow measurement systems only measure the axial component of the true blood flow velocity vector. In order to overcome this limitation, a technique which tracks blood cell scatterers as they move between three ultrasound beams has been developed. With this technique, the entire 3-D blood flow velocity vector can be estimated. Previous work has presented the theory behind the technique, lens transducer design and construction, as well as results of computer simulations and preliminary experimental results. This work presents the first experimental results obtained with a prototype system for continuous, fully developed flow in a flow phantom under a wide range of flow rates and flow directions. The results indicate that the accurate measurement of the 3-D flow velocity vector using this technique is possible.
引用
收藏
页码:85 / 95
页数:11
相关论文
共 41 条
[1]  
Akiyama I., 1988, ACOUST IMAG, V17, P499
[2]   EXPERIMENTAL VELOCITY PROFILES AND VOLUMETRIC PLOW VIA 2-DIMENSIONAL SPECKLE TRACKING [J].
BOHS, LN ;
FRIEMEL, BH ;
TRAHEY, GE .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1995, 21 (07) :885-898
[3]  
BONNEFOUS O, 1988 IEEE ULTR S P, P795
[4]  
CARSON P, 1993 IEEE ULTR S P, P1023
[5]   VECTORIAL MEASUREMENT OF BLOOD VELOCITY BY MEANS OF ULTRASOUND [J].
DOTTI, D ;
LOMBARDI, R ;
PIAZZI, P .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1992, 30 (02) :219-225
[6]   VOLUMETRIC BLOOD-FLOW VIA TIME-DOMAIN CORRELATION - EXPERIMENTAL-VERIFICATION [J].
EMBREE, PM ;
OBRIEN, WD .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1990, 37 (03) :176-189
[7]   ANGLE INDEPENDENT DOPPLER COLOR IMAGING - DETERMINATION OF ACCURACY AND A METHOD OF DISPLAY [J].
FEI, DY ;
FU, CT ;
BREWER, WH ;
KRAFT, KA .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1994, 20 (02) :147-155
[8]  
FERRARA KW, 1995 IEEE ULTR S P, P1467
[9]   FLOW VELOCITY PROFILE VIA TIME-DOMAIN CORRELATION ERROR ANALYSIS AND COMPUTER-SIMULATION [J].
FOSTER, SG ;
EMBREE, PM ;
OBRIEN, WD .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1990, 37 (03) :164-175
[10]   3-DIMENSIONAL DOPPLER VELOCIMETRY OF FLOW JETS [J].
FOX, MD ;
GARDINER, WM .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1988, 35 (10) :834-841